BACKGROUND: Electrical impedance tomography (EIT) provides information on global and regional ventilation during tidal breathing and mechanical ventilation. During forced expiration maneuvers, the linearity of EIT and spirometric data has been documented in healthy persons. The present study investigates the potential diagnostic use of EIT in pediatric patients with asthma.
METHODS: EIT and spirometry were performed in 58 children with asthma (average age ± SD: 11.86 ± 3.13 years), and 58 healthy controls (average age ± SD: 12.12 ± 2.9 years). The correlation between EIT data and simultaneously acquired spirometric data were tested for FEV1, FEV0.5 , MEF75 , MEF50 , and MEF25 . Binary classification tests were performed for the EIT-derived Tiffeneau index FEV1 /FVC and the bronchodilator test index ΔFEV1 . Average flow-volume (FV) loops were generated for patients with pathologic spirometry to demonstrate the feasibility of EIT for graphic diagnosis of asthma.
RESULTS: Spirometry and global EIT-based FV loops showed a strong correlation (P < 0.001, r > 0.9 in FEV1 and FEV0.5 ). In all criteria, the binary classification tests yielded high specificity (>93%), a high positive predictive value (≥75%) and a high negative predictive value (>80%), while sensitivity was higher in ΔFEV1 (86.67%) and lower in FEV1 /FVC (25% and 35.29%). A typical concave shape of the EIT-derived average FV loops was observed for asthmatic children with improvement after bronchospasmolysis.
CONCLUSIONS: Global FV loops derived from EIT correlate well with spirometry. Positive bronchospasmolysis can be observed in EIT-derived FV loops. Flow-volume loops originated from EIT have a potential to visualize pulmonary function.
© 2018 Wiley Periodicals, Inc.

Entities:  

Keywords:  asthma; electrical impedance tomography; flow volume loop; pulmonary function testing; spirometry; tiffeneau test

Mesh:

Year:  2018        PMID: 29405616     DOI: 10.1002/ppul.23962

Source DB:  PubMed          Journal:  Pediatr Pulmonol        ISSN: 1099-0496


  2 in total

1.  Reconstruction of conductivity distribution with electrical impedance tomography based on hybrid regularization method.

Authors:  Yanyan Shi; Xiaoyue He; Meng Wang; Bin Yang; Feng Fu; Xiaolong Kong
Journal:  J Med Imaging (Bellingham)       Date:  2021-06-17

Review 2.  A narrative review of electrical impedance tomography in lung diseases with flow limitation and hyperinflation: methodologies and applications.

Authors:  Ling Sang; Zhanqi Zhao; Zhimin Lin; Xiaoqing Liu; Nanshan Zhong; Yimin Li
Journal:  Ann Transl Med       Date:  2020-12
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.