Nihar Ranjan Pani1, Lilakant Nath2, Akhilesh Vikram Singh2, Santosh Kumar Mahapatra1.
Abstract
Nateglinide has been widely used in the treatment of type-2 diabetics as an insulin secretogoga. A reliable, rapid, simple and sensitive reversed-phase high performance liquid chromatography (RP-HPLC) method was developed and validated for determination of nateglinide in rabbit plasma. The method was developed on Hypersil BDSC-18 column (250 mm×4.6 mm, 5 mm) using a mobile phase of 10 mM phosphate buffer (pH 2.5) and acetonitrile (35:65, v/v). The elute was monitored with the UV-vis detector at 210 nm with a flow rate of 1 mL/min. Calibration curve was linear over the concentration range of 25-2000 ng/mL. The retention times of nateglinide and internal standard (gliclazide) were 9.608 min and 11.821 min respectively. The developed RP-HPLC method can be successfully applied to the quantitative pharmacokinetic parameters determination of nateglinide in rabbit model.Entities:
Keywords: HPLC; Nateglinide; Pharmacokinetics; Rabbit plasma
Year: 2012 PMID: 29403789 PMCID: PMC5760937 DOI: 10.1016/j.jpha.2012.05.002
Source DB: PubMed Journal: J Pharm Anal ISSN: 2214-0883
Figure 1Chemical structure of nateglinide.
Figure 2Chromatogram of the mixture of nateglinide and Internal Standard in mobile phase. (1) Nateglinide-9.261; (2) internal standard-11.663.
Figure 3Chromatogram of blank plasma of rabbit.
Figure 4Overlapping of chromatograms of NTG-IS mixture with blank plasma.
Figure 5Chromatogram of medium level quality control sample of nateglinide (489.48 ng/mL). (1) Nateglinide-9.261; (2) internal standard-11.663.
Peak area of the chromatogram of NTG and IS at different concentrations of NTG.
| Concentration of nateglinide (ng/mL) | Peak area of the chromatogram | Ratio of peak area | |
|---|---|---|---|
| Nateglinide | IS (Gliclazide) | ||
| 100.74 | 13,175 | 118,875 | 0.11 |
| 167.89 | 24,406 | 104,728 | 0.23 |
| 239.85 | 36,044 | 104,448 | 0.35 |
| 342.64 | 47,183 | 94,784 | 0.50 |
| 489.48 | 72,401 | 111,815 | 0.65 |
| 699.26 | 104,507 | 100,611 | 1.04 |
| 998.95 | 160,217 | 103,005 | 1.56 |
| 1664.91 | 275,030 | 117,616 | 2.34 |
| 3329.83 | 550,296 | 113,730 | 4.84 |
Precision and accuracy of the method for the estimation of nateglinide in rabbit plasma.
| Concentration (ng/mL) | Intra-day precision | Inter-day precision | ||||
|---|---|---|---|---|---|---|
| Mean (ng/mL) | RSD (%) | Accuracy (%) | Mean (ng/mL) | RSD (%) | Accuracy (%) | |
| 239.85 | 236.81 | 4.95 | 98.73 | 235.51 | 10.17 | 98.19 |
| 489.48 | 485.32 | 10.58 | 99.15 | 486.05 | 10.72 | 99.30 |
| 998.95 | 996.22 | 7.56 | 99.73 | 997.24 | 7.91 | 99.83 |
Stability of nateglinide in rabbit plasma.
| Stability | Concentration of NTG (ng/mL) | Initial concentration(mean; ng/mL) | Final concentration(mean; ng/mL) | Deviation (%) | RSD (%) | Accuracy (%) |
|---|---|---|---|---|---|---|
| Freeze–thaw stability | 239.85 | 236.67 | 236.32 | −0.15 | 8.72 | 98.53 |
| 489.48 | 486.60 | 493.80 | 1.48 | 13.78 | 100.88 | |
| 998.95 | 999.83 | 1003.00 | 0.32 | 6.32 | 100.41 | |
| 24 h stability | 239.85 | 243.73 | 236.04 | 2.90 | 3.81 | 98.67 |
| 489.48 | 484.92 | 493.16 | −0.57 | 11.56 | 99.63 | |
| 998.95 | 996.55 | 997.31 | 0.92 | 8.71 | 98.84 | |
| 1 month stability | 239.85 | 236.04 | 239.08 | −1.29 | 7.29 | 99.68 |
| 489.48 | 493.16 | 482.68 | 2.13 | 2.42 | 98.61 | |
| 998.95 | 997.31 | 1007.42 | −1.01 | 8.90 | 100.85 | |
Figure 6Mean plasma drug concentration versus time profile after single oral administration of nateglinide (60 mg) in rabbits.
Pharmacokinetic parameters of single dose administration of 60 mg nateglinide to rabbits (n=3).
| Pharmacokinetic parameters | Observed value |
|---|---|
| Maximum plasma concentration, | 13.407±0.476 |
| Time required to reach maximum plasma concentration, | 1.167±0.373 |
| Area under the curve, AUC0−∞ (h μg/mL) | 45.353±1.219 |
| Area under momentum curve, AUMC0−∞ (h2 μg/mL) | 115.939±6.618 |
| Volume of distribution, | 2.714±0.116 |
| Elimination half life, | 0.317±0.068 |
| Elimination rate constant, | 0.489±0.023 |
| Absorption rate constant, | 2.225±0.038 |
| Mean residence time, MRT (h) | 2.556±0.146 |
| Clearance, Cl (L/h) | 1.326±0.003 |