Angle-insensitive amorphous silicon optical filter for fluorescence contact imaging.

Efthymios P Papageorgiou, Hui Zhang, Bernhard E Boser, Catherine Park, Mekhail Anwar.   

Abstract

We introduce a novel amorphous silicon absorption filter that has high rejection for all angles of incident light for wavelengths below approximately 700 nm. This filter is used for microscopic cancer tissue detection in a small intraoperative contact fluorescence imaging system that requires excitation light at oblique angles. Our 15 μm thick filter presents over five orders of magnitude rejection at 633 nm, making it compatible with several clinically tested fluorophores, including IR700DX. We have demonstrated imaging of fluorescently labeled human epidermal growth factor receptor 2+ breast cancer tissue using the filter, and we can reliably detect microscopic clusters of breast cancer cells with only a 75 ms integration time.

Entities:  

Year:  2018        PMID: 29400857     DOI: 10.1364/OL.43.000354

Source DB:  PubMed          Journal:  Opt Lett        ISSN: 0146-9592            Impact factor:   3.776


  1 in total

1.  Chip-Scale Angle-Selective Imager for In Vivo Microscopic Cancer Detection.

Authors:  Efthymios P Papageorgiou; Bernhard E Boser; Mekhail Anwar
Journal:  IEEE Trans Biomed Circuits Syst       Date:  2019-12-12       Impact factor: 3.833