Literature DB >> 29393160

Diagnostic Accuracy of a Bioluminescence System for the Assessment of Caries Activity on Occlusal Surfaces.

Anahita Jablonski-Momeni1, Johanna Moos1, Vahid Sakhaei Manesh2, Richard Stoll2.   

Abstract

OBJECTIVE: Assessment of caries lesion activity is usually performed using visual-tactile criteria. A new approach is to use a system consisting of a photoprotein, which is specific for free calcium ions, along with an integrated camera that visualizes the elevated calcium ions on the lesion as a light signal (bioluminescence). This study aimed to evaluate the reproducibility and diagnostic accuracy of a newly developed bioluminescence system to assess caries lesion activity on occlusal surfaces in vitro.
MATERIALS AND METHODS: Ninety-four extracted permanent teeth with sound surfaces and both active and inactive caries lesions were included. The extent and activity of the investigation sites were classified visually according to the International Caries Detection and Assessment System. Digital images of the teeth were produced by the bioluminescence system and the images were analyzed for the presence of caries activity (the bioluminescent areas at the investigation sites). The images were reanalyzed after 4 weeks. Teeth were hemisected and methyl red dye was applied on the sections to validate lesion activity histologically.
RESULTS: Agreement between the bioluminescence readings was shown by κ values of 0.802-0.917. A significant positive correlation (p < 0.001) was found across all methods (rs 0.618-0.811). Sensitivity and specificity for activity assessment using histology as the gold standard (cut-off: active/inactive) were, respectively, 83.6 and 85.0% for the visual assessment, and 92.5 and 90.0% for the bioluminescence method. Comparison of areas under the receiver-operating characteristic curves showed no significant difference between the visual and bioluminescence systems (p = 0.094).
CONCLUSION: The bioluminescence system demonstrated high in vitro reproducibility and good diagnostic accuracy for activity assessment of caries lesions on occlusal surfaces which were not significantly different from the values obtained in the visual assessment.
© 2018 S. Karger AG, Basel.

Entities:  

Keywords:  Activity assessment; Bioluminescence; Caries; Occlusal caries; Photoprotein

Mesh:

Year:  2018        PMID: 29393160     DOI: 10.1159/000486139

Source DB:  PubMed          Journal:  Caries Res        ISSN: 0008-6568            Impact factor:   4.056


  7 in total

1.  Use of a laser fluorescence device for the in vitro activity assessment of incipient caries lesions.

Authors:  Anahita Jablonski-Momeni; Maria Rüter; Juliane Röttker; Heike Korbmacher-Steiner
Journal:  J Orofac Orthop       Date:  2019-10-07       Impact factor: 1.938

2.  Carious lesion activity assessment in clinical practice: a systematic review.

Authors:  Noémie Drancourt; Valérie Roger-Leroi; Stefania Martignon; Anahita Jablonski-Momeni; Nigel Pitts; Sophie Doméjean
Journal:  Clin Oral Investig       Date:  2019-02-21       Impact factor: 3.573

3.  The Calcivis story - enamel caries activity assessment from technology to practice.

Authors:  Nigel B Pitts; Chris Longbottom; Adam Christie; Bruce Vernon; Graham Bailey
Journal:  Br Dent J       Date:  2021-12-17       Impact factor: 2.727

4.  Visual or visual-tactile examination to detect and inform the diagnosis of enamel caries.

Authors:  Richard Macey; Tanya Walsh; Philip Riley; Anne-Marie Glenny; Helen V Worthington; Lucy O'Malley; Janet E Clarkson; David Ricketts
Journal:  Cochrane Database Syst Rev       Date:  2021-06-14

5.  Agreement of Bioluminescence Measurements and Visual Assessment in Monitoring Occlusal Surfaces of Permanent Teeth.

Authors:  Anahita Jablonski-Momeni; Boris Jablonski; Monika Heinzel-Gutenbrunner; Heike Korbmacher-Steiner
Journal:  J Clin Med       Date:  2022-01-17       Impact factor: 4.241

6.  New Caries Diagnostic Tools in Intraoral Scanners: A Comparative In Vitro Study to Established Methods in Permanent and Primary Teeth.

Authors:  Maximiliane Amelie Schlenz; Berenike Schupp; Alexander Schmidt; Bernd Wöstmann; Ingo Baresel; Norbert Krämer; Nelly Schulz-Weidner
Journal:  Sensors (Basel)       Date:  2022-03-10       Impact factor: 3.576

7.  Clinical validation of a novel bioluminescence imaging technology for aiding the assessment of carious lesion activity status.

Authors:  Nigel Pitts; Neil Shanks; Christopher Longbottom; Marjory Willins; Bruce Vernon
Journal:  Clin Exp Dent Res       Date:  2021-03-10
  7 in total