Santosh K Padala1, Kenneth A Ellenbogen1, Jayanthi N Koneru1.
Abstract
Entities:
Keywords: Capture management; Cardiac resynchronization therapy; His bundle pacing; Loss of capture; Pacemaker
Year: 2017 PMID: 29387549 PMCID: PMC5778145 DOI: 10.1016/j.hrcr.2017.07.018
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Device parameters on postoperative day 1
| Mode = DDDR | AP = 1%, VP = 99.9% |
|---|---|
| Lower rate limit = 60 bpm | RA lead |
| Maximum tracking rate = 150 beats/min | |
| Maximum sensor rate = 140 beats/min | RV lead |
| Paced AV delay = 150 ms | |
| His bundle lead | |
| Capture management |
AP = atrial pacing; AV = atrioventricular; CRT = cardiac resynchronization therapy; LV = left ventricular; RA = right atrial; RV = right ventricular; VP = ventricular pacing.
Figure 1Telemetry strips showing intermittent third-degree atrioventricular block.
Figure 2Atrial Capture Management algorithm.
Figure 3A: Subthreshold right ventricular support paces during Atrial Capture Management algorithm. B: Enough device-determined “premature ventricular contractions” occur in each episode that the Atrial Capture Management test is aborted. The device returns to “biventricular” pacing, which in this case is His-bundle pacing. ACM = atrial capture management; AP = atrial pacing; AS = atrial sensing; LVP = left ventricular pacing; PVC = premature ventricular contraction; RVP = right ventricular pacing; VS = ventricular sensing.