Dislodgement of circular mapping catheter electrode in the left atrium: A near miss.

Fengxiang Zhang1,2, Lijun Tang3, Bing Yang1, Gang Yang1, Yinsu Zhu3, Minglong Chen1.   

Abstract

Entities:  

Keywords:  Atrial fibrillation; Circular mapping catheter; Complication; Cryoablation balloon ablation; Foreign matter

Year:  2017        PMID: 29387542      PMCID: PMC5778103          DOI: 10.1016/j.hrcr.2017.06.001

Source DB:  PubMed          Journal:  HeartRhythm Case Rep        ISSN: 2214-0271


× No keyword cloud information.

Introduction

Use of a circular mapping catheter (CMC) in pulmonary vein (PV) isolation procedures is standard practice and generally considered to be safe in patients with atrial fibrillation (AF). However, we report a case where a metallic electrode displaced from a distal CMC was seen free-floating in the left atrium (LA) on fluoroscopy. Unexpectedly, the electrode traveled retrograde and became lodged in a terminal branch of the right inferior pulmonary vein (RIPV).

Case report

A 66-year-old Chinese man with paroxysmal AF underwent isolation of all 4 PVs using a cryoablation balloon. All 4 PVs were successfully isolated. A CMC (AFocus II; 7F, 15 mm diameter; St. Jude Medical, St. Paul, MN) was placed through a steerable sheath (FlexCath Advance; 12F inner diameter; Medtronic, Minneapolis, MN) to confirm PV isolation at the end of the procedure. Upon removal of the CMC from the LA with significant exertion owing to resistance, a free-floating metallic electrode that had displaced from the CMC was seen on fluoroscopy free-floating within the LA (Figure 1A–D, Supplemental video 1). This catheter was removed and the absence of the distal pole was observed. Potential for cerebral embolization was of utmost concern and the patient was counseled on options and risks. The patient agreed to proceed with urgent cardiac surgery to retrieve the electrode. While we were waiting for the operating theatre to be ready, the electrode had surprisingly traveled retrograde and become lodged in a terminal branch of the RIPV, as seen on fluoroscopy (Figure 1E, F, Supplemental video 2) and computed tomography angiography (Figure 1G, H). Repeat imaging during 12-month follow-up confirmed that the electrode had not migrated any further, and the patient remained asymptomatic, with no significant mitral regurgitation on echocardiography.
Figure 1

Pulmonary vein (PV) angiography and thoracic computed tomography angiography (CTA) of displaced electrode. A: Right PV angiography at right anterior oblique 30-degree projection. B: Left PV angiography at left anterior oblique 45-degree projection. C: Foreign matter (arrow) was revealed at the right posterosuperior left atrium (LA). D: Mobile foreign matter (arrow) now seen at the left posterosuperior LA. E: Foreign matter (arrow) seen in a vessel outside of cardiac shadow. F: Foreign matter (arrow) was revealed at the left posteroinferior border of fluorescopic image. G: The foreign matter (arrow) was confirmed in a small terminal branch of the right inferior PV (RIPV): transverse view of CTA. H: The foreign matter (arrow) was confirmed in a small terminal branch of the RIPV: sagittal view.

Pulmonary vein (PV) angiography and thoracic computed tomography angiography (CTA) of displaced electrode. A: Right PV angiography at right anterior oblique 30-degree projection. B: Left PV angiography at left anterior oblique 45-degree projection. C: Foreign matter (arrow) was revealed at the right posterosuperior left atrium (LA). D: Mobile foreign matter (arrow) now seen at the left posterosuperior LA. E: Foreign matter (arrow) seen in a vessel outside of cardiac shadow. F: Foreign matter (arrow) was revealed at the left posteroinferior border of fluorescopic image. G: The foreign matter (arrow) was confirmed in a small terminal branch of the right inferior PV (RIPV): transverse view of CTA. H: The foreign matter (arrow) was confirmed in a small terminal branch of the RIPV: sagittal view.

Discussion

A variety of complications have been described during AF ablation. Wu and colleagues reported CMC entrapment in the mitral valve apparatus in a patient during AF ablation. The case of catheter fragments being displaced and retained in patients is not a commonly reported occurrence. The case reported by Calvo and colleagues is similar, except that it involved a proximal electrode from a different CMC (Reflexion Spiral, St. Jude Medical). The floating metallic electrodes did not induce serious complications because they lodged in a terminal branch of the RIPV. Catheter fragment displacement involves the important sheath–electrode relationship during catheter removal. Some combinations of CMC and sheath design are more prone to this complication and should warrant careful removal of the CMC. Particular care should be given when the proximal and distal portions of the CMC interface with the sheath tip. The mechanism of dislodged electrode ending up in the RIPV is unclear but may be owing to the weight of the catheter fragments and the posterior anatomy of the RIPV. This may allow for retrograde travel into the RIPV as opposed to the more deleterious route into the left ventricle or aorta. We considered this incident a near miss with potential for systemic embolization, and the appropriate regulatory authorities have been notified. It is important to be aware of possible complications involving electrode dislodgement occurring during circular mapping catheter removal. Serial imaging should be performed to monitor location of retained fragments while preparing for possible intervention. Given the risk of distal embolism, careful consideration should be given to urgent intervention to remove a mobile, retained foreign object.
  4 in total

1.  Circular mapping catheter entrapment in the mitral valve apparatus: a previously unrecognized complication of focal atrial fibrillation ablation.

Authors:  Richard C Wu; Jeffrey A Brinker; David D Yuh; Ronald D Berger; Hugh G Calkins
Journal:  J Cardiovasc Electrophysiol       Date:  2002-08

2.  A Missing Pole from a Circular Mapping Catheter Ablation.

Authors:  Naiara Calvo; Hugo Arguedas-Jimenez; Ignacio García-Bolao
Journal:  J Cardiovasc Electrophysiol       Date:  2015-06-16

Review 3.  Complication rates of atrial fibrillation ablations: comparison of safety outcomes from real world to contemporary randomized control trials.

Authors:  Naga Venkata K Pothineni; Abhishek Deshmukh; Sadip Pant; Nileshkumar J Patel; Apurva Badheka; Ankit Chothani; Neeraj Shah; Kathan Mehta; Ghanshyambhai T Savani; Vikas Singh; Peeyush Grover; Vipulkumar Bhalara; Nilay Patel; Shilpkumar Arora; Ankit Rathod; Juan Viles-Gonzalez; Hakan Paydak
Journal:  Int J Cardiol       Date:  2014-05-02       Impact factor: 4.164

4.  Electrophysiological findings during ablation of persistent atrial fibrillation with electroanatomic mapping and double Lasso catheter technique.

Authors:  Feifan Ouyang; Sabine Ernst; Julian Chun; Dietmar Bänsch; Yigang Li; Anselm Schaumann; Hercules Mavrakis; Xingpeng Liu; Florian T Deger; Boris Schmidt; Yumei Xue; Jiang Cao; Detlef Hennig; He Huang; Karl-Heinz Kuck; Matthias Antz
Journal:  Circulation       Date:  2005-11-07       Impact factor: 29.690

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.