Accuracy and precision of flash glucose monitoring sensors inserted into the abdomen and upper thigh compared with the upper arm.

Sara Charleer1,2, Chantal Mathieu1, Frank Nobels3, Pieter Gillard1.   

Abstract

Nowadays, most Belgian patients with type 1 diabetes use flash glucose monitoring (FreeStyle Libre [FSL]; Abbott Diabetes Care, Alameda, California) to check their glucose values, but some patients find the sensor on the upper arm too visible. The aim of the present study was to compare the accuracy and precision of FSL sensors when placed on different sites. A total of 23 adults with type 1 diabetes used three FSL sensors simultaneously for 14 days on the upper arm, abdomen and upper thigh. FSL measurements were compared with capillary blood glucose (BG) measurements obtained with a built-in FSL BG meter. The aggregated mean absolute relative difference was 11.8 ± 12.0%, 18.5 ± 18.4% and 12.3 ± 13.8% for the arm, abdomen (P = .002 vs arm) and thigh (P = .5 vs arm), respectively. Results of Clarke error grid analysis for the arm and thigh were similar (zone A: 84.9% vs 84.5%; P = .6), while less accuracy was seen for the abdomen (zone A: 69.4%; P = .01). Apart from the first day, the accuracy of FSL sensors on the arm and thigh was more stable across the 14-day wear duration than accuracy of sensors on the abdomen, which deteriorated mainly during week 2 (P < .0005). The aggregated precision absolute relative difference was markedly lower for the arm/thigh (10.9 ± 11.9%) compared with the arm/abdomen (20.9 ± 22.8%; P = .002). Our results indicate that the accuracy and precision of FSL sensors placed on the upper thigh are similar to the upper arm, whereas the abdomen performed unacceptably poorly.
© 2018 John Wiley & Sons Ltd.

Entities:  

Keywords:  clinical trial; flash glucose monitoring; type 1 diabetes

Mesh:

Substances:

Year:  2018        PMID: 29381253     DOI: 10.1111/dom.13239

Source DB:  PubMed          Journal:  Diabetes Obes Metab        ISSN: 1462-8902            Impact factor:   6.577


  4 in total

1.  Comparative Accuracy Analysis of a Real-time and an Intermittent-Scanning Continuous Glucose Monitoring System.

Authors:  Manuela Link; Ulrike Kamecke; Delia Waldenmaier; Stefan Pleus; Arturo Garcia; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2019-12-17

Review 2.  Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices.

Authors:  Guido Freckmann; Stefan Pleus; Mike Grady; Steven Setford; Brian Levy
Journal:  J Diabetes Sci Technol       Date:  2018-11-19

3.  Comparison of Interstitial Fluid Glucose Levels Obtained by Continuous Glucose Monitoring and Flash Glucose Monitoring in Patients With Type 2 Diabetes Mellitus Undergoing Hemodialysis.

Authors:  Takahiro Yajima; Hiroshi Takahashi; Keigo Yasuda
Journal:  J Diabetes Sci Technol       Date:  2019-10-18

4.  Flash Glucose Monitoring System for People with Type 1 or Type 2 Diabetes: A Health Technology Assessment.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2019-12-12
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.