| Literature DB >> 29377476 |
Eric Vandervoort1, Horacio Patrocinio2, Tom Chow3, Emilie Soisson2, Dominic Béliveau Nadeau4.
Abstract
The Canadian Organization of Medical Physicists (COMP), in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR) has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. This particular TQC contains detailed performance objectives and safety criteria for CyberKnife® Technology. The quality control recommendations in this document are based upon previously published guidelines and the collective experience of all Canadian sites using this technology. This TQC guideline has been field tested at the newest Canadian CyberKnife installation site and includes recommendations for quality control of the Iris™ and InCise™ MLC collimation systems.Entities:
Keywords: CyberKnife Radiosurgery System; quality control guidelines; radiation treatment therapy equipment; robotic radiosurgery; stereotactive ablative radiotherapy
Mesh:
Year: 2018 PMID: 29377476 PMCID: PMC5849826 DOI: 10.1002/acm2.12263
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Daily quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Daily | |||
| DL1 | Emergency robotic arm motion stop circuit (if present) | Functional | |
| DL2 | Robotic arm collision detection interlocks | Functional | |
| DL3 | Visual check of beam laser and a standard floor mark | n/a | 1 mm |
| DL4 | Accelerator output | 2% | 3% |
| DL5 | Automated quality assurance test (alternate daily between fixed and Iris™ collimators and the InCise™ MLC) | 0.75 mm in any direction | 1 mm radial |
| DL6 | Modified picket fence field tests for defocused MLC | Visual inspection of junctions | |
Monthly quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Monthly | |||
| ML1 | Energy constancy (change in TPR or PDD ratio) | 1% | 2% |
| ML2 | Accelerator output | 2% | 3% |
| ML3 | Intracranial and extracranial isocentric end‐to‐end test; scheduled to cycle through each clinically used tracking method, path, and collimation system (fixed, Iris™, and InCise™ MLC) | Error in any direction: 0.75 mm (static); 1 mm (Synchrony®) | Radial error: 1 mm (static); 1.5 mm (Synchrony®) |
| ML4 | Nonisocentric patient specific quality assurance; scheduled monthly to cycle through each clinically used tracking method, path, and collimation system at least quarterly (fixed, Iris™, and InCise™ MLC) | n/a | 5% / 2 mm (static); 5% / 3 mm (Synchrony®) |
| ML5 | Iris™ field size verification | ±0.3 mm | ±0.5 mm |
| ML6 | Garden fence MLC test | n/a | ±0.5 mm for 95% of leaf positions; <2 failures/leaf |
| ML7 | Low contrast details visibility and spatial resolution of amorphous silicon detectors | n/a | Reproducible |
| ML8 | Records | Complete | |
Quarterly quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Quarterly | |||
| QL1 | Beam symmetry | 2% | 3% |
| QL2 | Beam profile shape compared to beam data | 2% / 2 mm | 3% / 2 mm |
| QL3 | Imager alignment center | 0.5 mm | 1 mm |
Annual quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Annual | |||
| AL1 | Reference dosimetry | 1% | 2% |
| AL2 | TPR or PDD and output factors for each clinically used collimation system | 1% | 2% |
| AL3 | Radial profile constancy | 1%/1 mm | 2%/2 mm |
| AL4 | Dose output linearity to lowest MU/beam used |
1%/1 MU |
2%/2 MU |
| AL5 | Verify relative location of the central axis beam laser to the radiation central axis to ensure it has not changed from the baseline and is coincident | Change from baseline: 0.5 mm | Coincidence of laser and central axes: 1 mm |
| AL6 | Verification of the second order path calibration | n/a |
Each node |
| AL7 | Run Synchrony® end‐to‐end test with at least 20° phase shift; analyze penumbra spread compared to static delivery |
Radial Error: 1.0 mm |
Radial Error: 1.5 mm |
| AL8 | InCise™ MLC Leaf transmission | 0.5% | 1% |
| AL9 | InCise™ MLC Leaf leakage between leaves | 0.5% | 1% |
| AL10 | InCise™ MLC Transmission between abutting leaves | 0.5% | 1% |
| AL11 | InCise™ MLC leaf alignment with jaws | 0.5o | 1.0o |
| AL12 | Imager kVp, mA and timer accuracy, exposure linearity, exposure reproducibility | n/a | Reproducible |
| AL13 | Quantitative assessment of contrast, noise, and spatial resolution of amorphous silicon detector | n/a | Reproducible |
| AL14 | Independent review and update of quality assurance references | Complete | |