Performance evaluation of the new measurement channels on the automated Sysmex XN-9000 hematology analyzer.

E Schapkaitz1, S Raburabu2.   

Abstract

BACKGROUND: The automated Sysmex XN-9000 hematology system has been designed to meet the throughput and efficiency requirements of high volume laboratories with predominantly abnormal samples. New measurement channels have been introduced namely the white cell nucleated (WNR), white cell differential (WDF), white cell precursor (WPC) and fluorescent platelet (PLT-F) channels.
METHODS: The performance of the new measurement channels was evaluated with regards to precision, accuracy, linearity, carryover, throughput and stability. 275 slides were assessed for morphology flagging. Adult and pediatric samples with normal and abnormal hematology profiles were included.
RESULTS: The XN-9000 demonstrated acceptable imprecision, good linearity for high and low ranges and no carryover. The full blood count and reticulocyte on the XN-9000 correlated well with the reference ADVIA(2)120. The PLT-F (127±84×109/l) compared with the optical platelet count (131±76×109/l) (r=0.97) and the imprecision was <4% on thrombocytopenic samples. The low white blood cell (WBC) mode reported accurate differentials for samples with a WBC count<0.5×109/l (r=0.93). The nucleated red blood cell count from the WNR (1.22±3.96%) showed an excellent correlation with the manual method (1.12±4.79%) (r=0.99). The WPC channel showed 100% sensitivity for the detection of blasts and abnormal lymphocytes. Further, the WPC correctly suppressed the initial blast/abnormal lymphocyte flag in 34% of the reflexed samples.
CONCLUSION: The XN-9000 showed enhanced analytical performance and workflow efficiency for a wide range of patient samples which can be attributed to the incorporation of new measurement channels.
Copyright © 2018 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Full blood count; Hematology analyzer; Morphology flags; Sysmex XN-9000; Verification

Mesh:

Year:  2018        PMID: 29374555     DOI: 10.1016/j.clinbiochem.2018.01.014

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  4 in total

1.  Concurrent Immune Suppression and Hyperinflammation in Patients With Community-Acquired Pneumonia.

Authors:  Xanthe Brands; Bastiaan W Haak; Augustijn M Klarenbeek; Natasja A Otto; Daniël R Faber; René Lutter; Brendon P Scicluna; W Joost Wiersinga; Tom van der Poll
Journal:  Front Immunol       Date:  2020-05-06       Impact factor: 7.561

2.  Rectal microbiota are coupled with altered cytokine production capacity following community-acquired pneumonia hospitalization.

Authors:  Robert F J Kullberg; Xanthe Brands; Augustijn M Klarenbeek; Joe M Butler; Natasja A Otto; Daniël R Faber; Brendon P Scicluna; Tom van der Poll; W Joost Wiersinga; Bastiaan W Haak
Journal:  iScience       Date:  2022-07-14

3.  Use of Middleware Data to Dissect and Optimize Hematology Autoverification.

Authors:  Rachel D Starks; Anna E Merrill; Scott R Davis; Dena R Voss; Pamela J Goldsmith; Bonnie S Brown; Jeff Kulhavy; Matthew D Krasowski
Journal:  J Pathol Inform       Date:  2021-04-07

4.  Performance evaluation of the XPEN60 CRP&SAA, a novel automated hematology analyzer, in detecting complete blood count, C-reactive protein, and serum amyloid A.

Authors:  Zengyan Zong; Lijuan Kan; Yanping Luo; Cheng Wang; Xibu Wang; Mengmeng Wang; Dan Xiong; Xiuming Zhang
Journal:  J Clin Lab Anal       Date:  2020-06-09       Impact factor: 2.352

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.