Ananda Giri Babu Alagar1,2, Kadirampatti Mani Ganesh, Karunakaran Kaviarasu.
Abstract
Dose calculation for small field radiotherapy with heterogeneity often involves discrepancies, so that algorithms used by treatment planning systems (TPS) should be evaluated with reference to achieving optimal treatment results. Accuracy of two model based algorithms, AcurosXB (AcXB) and the analytical anisotropic algorithm (AAA) from Eclipse TPS, were here tested. Measurements are made using Gafchromic EBT3 films with indigenously generated lung phantoms irradiated with 6 MV photons. Lung phantoms contained two types of tumor plugs, one kept at an interface attached to the chest wall in right lung (RIT) and the other at the centre of the left lung (LCT). RIT and LCT were studied with two different tumor diameters, 1.5 cm and 2.5 cm. Scanned images were planned in TPS with 3D-CRT, IMRT and VMAT and individual plans for each tumor were irradiated keeping the Gafchromic film at the centre of the tumor to evaluate the dose distribution in the central plane. Both algorithms, irrespective of delivery techniques, showed more deviation with smaller than larger diameter tumors. Also, both demonstrated maximum deviation at the junction of tumor and lung in both RIT and LCT cases. However, the deviation observed was higher with AAA and a minimal acceptable deviation of within 4 % was achieved with AcurosXB. Creative Commons Attribution LicenseEntities:
Keywords: TPS algorithms; lung heterogeneity; dose plane
Year: 2018 PMID: 29374410 PMCID: PMC5844627 DOI: 10.22034/APJCP.2018.19.1.253
Source DB: PubMed Journal: Asian Pac J Cancer Prev ISSN: 1513-7368
Figure 1Central Slice of CT Image of Phantom Showing Normal Lung, RIT, LCT and Hounsfield Unit of Structures. Horizontal Lines Cutting Tumors are the Respective Film Planes
Results for Left Lung Central Tumor (LCT) diameter of 2.5 cm
| Evaluation Parameters | Percentage Dose Difference w.r.t. Measured | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| 3D-CRT | IMRT-Static | IMRT-Dynamic | VMAT | ||||||
| AAA | AcXB | AAA | AcXB | AAA | AcXB | AAA | AcXB | ||
| CAX Absolute Dose deviation (in %) | 5.1 | 3 | 4 | 1.5 | 3.8 | 1 | 2.8 | 0.8 | |
| 2D-Gamma (Pass%) at Isocentre | 2%& 2mm | 92.73 | 94.88 | 92.8 | 95.7 | 2.54 | 95.33 | 90.65 | 94.72 |
| 3%&3mm | 93.65 | 96.24 | 95.3 | 97.2 | 97.56 | 98.82 | 93.93 | 97.47 | |
| 4%&4mm | 95.57 | 98.33 | 97.1 | 98.8 | 99.12 | 99.34 | 96.56 | 99.04 | |
| Correlation Coefficient | 0.951 | 0.965 | 0.932 | 0.973 | 0.9476 | 0.9772 | 0.9357 | 0.9781 | |
| 2D Gamma (Pass% for 3%&3mm) at 4 cm from isocentre in normal lung | 95.2 | 98 | 95 | 98.3 | 95.3 | 98.3 | 96.5 | 98.7 | |
Results for Left Lung Central Tumor (LCT) Diameter of 1.5 cm
| Evaluation Parameters | Percentage Dose Difference w.r.t. Measured | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| 3D-CRT | IMRT-Static | IMRT-Dynamic | VMAT | ||||||
| AAA | AcXB | AAA | AcXB | AAA | AcXB | AAA | AcXB | ||
| CAX Absolute Dose deviation (in %) | 5.2 | 3.3 | 5.5 | 1.8 | 5.2 | 1.3 | 3.2 | 1.3 | |
| 2D-Gamma (Pass%) at Isocentre | 2%&2mm | 91.8 | 94.3 | 91.9 | 95.8 | 92 | 95.1 | 90.4 | 94.8 |
| 3%&3mm | 93.2 | 95.8 | 95.2 | 97.4 | 96.7 | 98.2 | 93.2 | 97.1 | |
| 4%&4mm | 96.1 | 97.9 | 96.7 | 98.2 | 98.1 | 98.2 | 96.3 | 98.9 | |
| Correlation Coefficient | 0.94 | 0.959 | 0.9278 | 0.968 | 0.9423 | 0.979 | 0.924 | 0.975 | |
| 2D Gamma (Pass% for 3%&3mm) at 4 cm from isocentre in normal lung | 96.1 | 97.8 | 95.3 | 98.6 | 96.2 | 98.4 | 96.2 | 98.1 | |
Figure 2Dose Profile Difference between Algorithms Calculated 3D-CRT Plans and Film Measured Along Right to Left Direction at Central Axis for Left Lung Central Tumor (LCT).
Results for Right Lung Interface Tumor (RIT) Diameter of 2.5 cm
| Evaluation Parameters | Percentage Dose Difference w.r.t. Measured | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| 3D-CRT | IMRT-Static | IMRT-Dynamic | VMAT | ||||||
| AAA | AcXB | AAA | AcXB | AAA | AcXB | AAA | AcXB | ||
| CAX Absolute Dose deviation (in %) | 4 | 3 | 3 | 1.2 | 3.2 | 0.8 | 3.3 | 0.7 | |
| 2D-Gamma (Pass%) at Isocentre | 2%&2mm | 92.3 | 94.1 | 91.7 | 95.2 | 91.9 | 95.58 | 90.1 | 94.1 |
| 3%&3mm | 93.5 | 96.1 | 95.1 | 96.8 | 96.32 | 97.32 | 95.1 | 96.37 | |
| 4%&4mm | 95.2 | 96.93 | 96.1 | 97.8 | 97.35 | 98.14 | 96.3 | 98.04 | |
| Correlation Coefficient | 0.9459 | 0.963 | 0.922 | 0.965 | 0.932 | 0.9689 | 0.946 | 0.988 | |
| 2D Gamma (Pass% for 3%&3mm) at 4 cm from isocentre in normal lung | 95.2 | 97 | 96.3 | 97.3 | 96.2 | 98.1 | 97.2 | 98.2 | |
Results for Right Lung Interface Tumor (RIT) Diameter of 2.5 cm
| Evaluation Parameters | Percentage Dose Difference w.r.t. Measured | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| 3D-CRT | IMRT-Static | IMRT-Dynamic | VMAT | ||||||
| AAA | AcXB | AAA | AcXB | AAA | AcXB | AAA | AcXB | ||
| CAX Absolute Dose deviation (in %) | 6.4 | 3.3 | 5.1 | 1.5 | 5.2 | 1.5 | 3.5 | 1.6 | |
| 2D-Gamma (Pass%) at Isocentre | 2%&2mm | 95.3 | 95.3 | 92.7 | 95.7 | 92.1 | 95.5 | 90.2 | 94.5 |
| 3%&3mm | 93.6 | 96.8 | 95.4 | 97.1 | 96.3 | 97.7 | 94.3 | 96.4 | |
| 4%&4mm | 96.9 | 98.9 | 96.9 | 98.7 | 97.4 | 98.7 | 95.9 | 98.1 | |
| Correlation Coefficient | 0.9585 | 0.951 | 0.919 | 0.9698 | 0.94 | 0.971 | 0.9202 | 0.9703 | |
| 2D Gamma (Pass% for 3%&3mm) at 4 cm from isocentre in normal lung | 95.5 | 97.3 | 95.2 | 96.4 | 97.3 | 98.5 | 97.6 | 98.1 | |
Figure 6Dose Profile Difference between Algorithms Calculated 3D-CRT Plans and Film Measured along Right to Left direction at Central Axis for Right Lung Interface Tumor (RIT)