Acetabular All-Suture Anchor for Labral Repair: Incidence of Intraoperative Failure due to Pullout.

J W Thomas Byrd1, Kay S Jones2, Cynthia L Loring3, Stephanie L Sparks3.   

Abstract

PURPOSE: To report on the incidence and features of intraoperative anchor pullout in a consecutive series of patients undergoing arthroscopic labral repair of the hip.
METHODS: Over an 18-month period, 434 consecutive cases underwent labral repair by a single surgeon with a particular anchor system. The following data were recorded: (1) age and gender of all cases; (2) number of anchors used; (3) number of cases in which intraoperative anchor failure occurred; (4) number of anchors that failed; and (5) age and gender of those cases in which anchor failure occurred. Failures were reported for 3-month intervals. One patient underwent repair with an alternative anchor system during this time period and was excluded.
RESULTS: Mean age was 34.2 (14-71) years with 180 males and 254 females. A total of 2,007 anchors were used, averaging 4.6 per case (1-8). Thirty-three anchors pulled out among 30 patients, representing a 1.6% incidence among all anchors. Mean age among pullouts was 37.8 (17-54) years with 11 males and 19 females. There was no difference compared with patient population in which no anchor pulled: mean 33.9 (14-71) years (P = .085) with 169 males and 235 females (P = .578). Pullouts were evenly distributed over the 3-month intervals (4, 4, 6, 6, 5, 8). Pullout was mostly due to failure to securely imbed the anchor in bone. Only 2 were known to pull out in the presence of being securely seated in bone.
CONCLUSIONS: These data support that the security of this particular all-suture anchor at implantation is exceptionally reliable for a single experienced surgeon, and there is no demonstrable learning curve. LEVEL OF EVIDENCE: Level IV, retrospective review of a case series.
Copyright © 2017 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 29373296     DOI: 10.1016/j.arthro.2017.09.049

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  2 in total

1.  Arthroscopic Acetabular Labral Repair Using the Q-FIX Suture Anchor.

Authors:  J W Thomas Byrd
Journal:  Arthrosc Tech       Date:  2019-07-18

Review 2.  The Clinical and Biomechanical Performance of All-Suture Anchors: A Systematic Review.

Authors:  Selim Ergün; Umut Akgün; F Alan Barber; Mustafa Karahan
Journal:  Arthrosc Sports Med Rehabil       Date:  2020-05-28
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.