Kirsty Symons1, Colm Morrison1,2, Jason Parry1, Simon Woodings1,3, Yvonne Zissiadis1,4.
Abstract
A study was undertaken to explore the use of volumetric modulated arc therapy (VMAT) for total body irradiation (TBI). Five patient plans were created in Pinnacle3 using nine 6 MV photon dynamic arcs. A dose of 12 Gy in six fractions was prescribed. The planning target volume (PTV) was split into four subsections for the head, chest, abdomen, and pelvis. The head and chest beams were optimized together, followed by the abdomen and pelvis beams. The last stage of the planning process involved turning all beams on and performing a final optimization to achieve a clinically acceptable plan. Beam isocenters were shifted by 3 or 5 mm in the left-right, anterior-posterior, and superior-inferior directions to simulate the effect of setup errors on the dose distribution. Treatment plan verification consisted of ArcCheck measurements compared to calculated doses using a global 3%/3 mm gamma analysis. All five patient plans achieved the planning aim of delivering 12 Gy to at least 90% of the target. The mean dose in the PTV was 12.7 Gy. Mean lung dose was restricted to 8 Gy, and a dose reduction of up to 40% for organs such as the liver and kidneys proved feasible. The VMAT technique was found to be sensitive to patient setup errors particularly in the superior-inferior direction. The dose predicted by the planning system agreed with measured doses and had an average pass rate of 99.2% for all arcs. VMAT was found to be a viable treatment technique for total body irradiation.Entities:
Keywords: total body irradiation (TBI); total marrow irradiation (TMI); volumetric arc therapy (VMAT)
Mesh:
Year: 2018 PMID: 29368389 PMCID: PMC5849856 DOI: 10.1002/acm2.12257
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1The beam arrangement and stages for TBI planning in Pinnacle. Nine VMAT arcs are arranged longitudinally along the patient and assigned to four isocenters. Each isocenter has a superior (red) and inferiorly (yellow) offset arc. The chest isocenter has an extra arc with no field size restrictions (orange). (a) Step 1: of the planning process is to optimize and convert the head and chest beams together. (b) Step 2: head and chest beams are set to “None” in IMRT parameters and the abdomen and pelvis beams are optimized together. (c) Step 3: All beams are turned on and a final optimization process performed to smooth out junction regions.
Pinnacle3 optimization objectives used for VMAT TBI
| ROI | Optimization type | Target (cGy) | % Volume | Weight | a |
|---|---|---|---|---|---|
| ptv_VMAT | Min DVH | 1200 | 98 | 98 | – |
| ptv_VMAT | Max DVH | 1320 | 5 | 50 | – |
| ptv_VMAT | Uniform dose | 1220 | – | 85 | – |
| Liver_cont | Max EUD | 750 | – | 20 | 2 |
| Kidneys_prv | Max EUD | 750 | – | 20 | 2 |
| Lungs_cont | Max EUD | 680 | – | 40 | 3 |
| ptv_skeleton | Min DVH | 1200 | 98 | 100 | – |
Figure 2A “beams eye view” of a chest arc on the ArcCheck phantom. Both a left and right 8 cm shift was applied to the ArcCheck during measurement to capture the open segments at the field edges. The blue ROI represents the internal diode volume.
Figure 3Isodose distribution of patient 3 for the chest and pelvis in the axial view, and a sagittal and coronal view.
Planning outcomes for the PTV
| Patient | PTV Volume (cc) | Mean (Gy) | V100% | V95% | D2 cc (Gy) | D5 cc (Gy) | D90% (Gy) | D10% (Gy) | Homogeneity (D90/D10) |
|---|---|---|---|---|---|---|---|---|---|
| 1 | 39722 | 12.6 | 95.5 | 98.3 | 14.9 | 14.7 | 12.2 | 13.2 | 0.93 |
| 2 | 41082 | 12.8 | 96.1 | 98.2 | 14.5 | 14.4 | 12.3 | 13.2 | 0.93 |
| 3 | 44847 | 12.6 | 93.4 | 97.6 | 15.0 | 14.8 | 12.2 | 13.2 | 0.92 |
| 4 | 45458 | 12.6 | 92.1 | 97.2 | 14.9 | 14.7 | 12.1 | 13.1 | 0.92 |
| 5 | 61900 | 12.7 | 94.4 | 98.2 | 15.4 | 15.6 | 12.2 | 13.5 | 0.90 |
Percentage volume in the PTV receiving 90%, 100%, 105%, 110%, and 120% of the prescription dose
| Patient | Volume (%) | ||||
|---|---|---|---|---|---|
| 90% | 100% | 105% | 110% | 120% | |
| 1 | 99.1 | 95.5 | 50.5 | 6.6 | 0.03 |
| 2 | 99.0 | 96.1 | 71.6 | 10.5 | 0.01 |
| 3 | 98.7 | 93.4 | 57.9 | 8.6 | 0.08 |
| 4 | 98.3 | 92.1 | 48.7 | 7.5 | 0.06 |
| 5 | 98.7 | 94.4 | 70.2 | 20.4 | 0.46 |
Planning outcomes for the organs at risk
| Patient | Lungs | Liver | Kidneys | ||
|---|---|---|---|---|---|
| Lung Volume (cc) | Mean (Gy) | Mean (Gy) | Lt Mean (Gy) | Rt Mean (Gy) | |
| 1 | 2275 | 7.8 | 8.6 | 6.0 | 6.9 |
| 2 | 3396 | 7.7 | 8.1 | 7.4 | 6.9 |
| 3 | 2671 | 7.8 | 8.0 | 7.7 | 6.7 |
| 4 | 2236 | 7.4 | 7.5 | 6.2 | 6.2 |
| 5 | 3534 | 7.1 | 7.6 | 7.3 | 7.4 |
DVH analysis for the head and pelvis PTVs relative to the original treatment plan
| Head & Pelvis | 5, 5, 0 mm | 3, 3, 3 mm | ||||||
|---|---|---|---|---|---|---|---|---|
| L‐R 3 mm | L‐R 5 mm | A‐P 3 mm | A‐P 5 mm | S‐I 3 mm | S‐I 5 mm | |||
| PTV | ||||||||
| Mean | 100.0 | 99.9 | 100.0 | 99.8 | 99.9 | 99.1 | 100.0 | 99.8 |
| V100% | 99.8 | 99.1 | 99.8 | 98.9 | 95.3 | 88.5 | 99.1 | 95.1 |
| V95% | 100.0 | 99.8 | 100.0 | 99.9 | 99.8 | 97.5 | 99.9 | 99.7 |
| Max dose | 100.9 | 98.8 | 99.2 | 101.4 | 100.0 | 99.8 | 103.1 | 101.7 |
DVH analysis for the chest and abdomen PTVs relative to the original treatment plan
| Chest & Abdomen | 3, 3, 0 mm | 3, 3, 3 mm | ||||||
|---|---|---|---|---|---|---|---|---|
| L‐R 3 mm | L‐R 5 mm | A‐P 3 mm | A‐P 5 mm | S‐I 3 mm | S‐I 5 mm | |||
| PTV | ||||||||
| Mean | 99.9 | 99.8 | 100.0 | 99.9 | 100.0 | 100.1 | 100.2 | 99.5 |
| V100% | 99.6 | 99.0 | 99.9 | 99.4 | 96.2 | 92.3 | 100.4 | 94.0 |
| V95% | 99.8 | 99.3 | 100.0 | 99.7 | 100.2 | 99.1 | 100.1 | 98.4 |
| Max dose | 99.9 | 100.0 | 99.5 | 99.7 | 100.8 | 102.2 | 100.1 | 100.5 |
| OAR (mean dose) | ||||||||
| Lungs | 100.3 | 100.7 | 100.2 | 100.6 | 101.7 | 104.0 | 101.3 | 99.9 |
| Liver | 100.2 | 100.5 | 99.8 | 99.7 | 103.1 | 105.3 | 99.7 | 97.5 |
| Kidney L | 100.5 | 102.2 | 100.5 | 101.8 | 102.4 | 104.9 | 102.4 | 99.1 |
| Kidney R | 98.3 | 98.4 | 100.2 | 101.3 | 101.9 | 103.8 | 103.6 | 97.6 |
| Skeleton + 3 mm | ||||||||
| Mean | 99.8 | 99.4 | 99.9 | 99.8 | 99.9 | 98.9 | 100.5 | 99.6 |
| V100% | 99.7 | 98.5 | 101.6 | 101.5 | 91.8 | 86.2 | 101.5 | 93.7 |
| V95% | 99.4 | 98.1 | 100.7 | 100.4 | 99.6 | 91.7 | 100.5 | 96.5 |
ArcCheck Measurement results for the five patient test plans
| Patient | 0 cm lateral offset – All Beams | ±8 cm lateral offset – Chest beams | ||
|---|---|---|---|---|
| Average | Std Dev | Average | Std Dev | |
| 1 | 99.5 | 0.5 | 99.2 | 0.8 |
| 2 | 99.7 | 0.5 | 98.4 | 1.1 |
| 3 | 99.0 | 0.9 | 99.2 | 0.3 |
| 4 | 99.0 | 1.4 | 99.0 | 0.8 |
| 5 | 98.6 | 0.7 | 97.9 | 1.1 |