| Literature DB >> 29367907 |
Evandro Pereira Palacio1, Gilberto José Cação Pereira1, Paulo Roberto de Almeida Silvares1, Gabriel Guimarães Di Stasi2, Caio de Andrade Staut2, Trajano Sardenberg1.
Abstract
OBJECTIVE: Synthetic adhesives are used by various medical specialties, especially in surgery; however, studies reporting their use in orthopedic practice are scarce. The aim of this study was to compare the results in using ethyl-2-cyanoacrylate or butyl-2-cyanoacrylate in the treatment of fractures in rats.Entities:
Keywords: Cyanoacrylates; Fracture healing; Fractures, bones; Tissue adhesives
Year: 2017 PMID: 29367907 PMCID: PMC5771781 DOI: 10.1016/j.rboe.2017.11.010
Source DB: PubMed Journal: Rev Bras Ortop ISSN: 2255-4971
Fig. 1Dissection and exposure of the femur prior to osteotomy.
Fig. 2Femoral osteotomy.
Fig. 3Tibia from the HA group with abscess region (arrow).
Fig. 4Tibia from the SB group with adhesive extrusion (arrow).
Fig. 5Diagram of load vs. deformation.
Mean and standard deviation of body mass (g), according to the adhesive used and the moment of euthanasia.
| Adhesive used | Moment of euthanasia | ||
|---|---|---|---|
| 60 days | 120 days | 180 days | |
| Ethyl-2-cyanoacrylate | 204 ± 15.3 | 289.1 ± 23.6 | 330.8 ± 6.4 |
| Butyl-2-cyanoacrylate | 195.4 ± 7.2 | 273.5 ± 14.8 | 312.2 ± 21.2 |
Results of femoral tests: mean and standard deviation of the maximum load (N), yield point strength (N), and stiffness coefficient (N/mm), according to the adhesive used and the moment of euthanasia.
| Studied variable | Adhesive used | Moment of euthanasia | ||
|---|---|---|---|---|
| 60 days | 120 days | 180 days | ||
| Maximum load (N) | Ethyl-2-cyanoacrylate | 138.7 ± 12.4 | 141.4 ± 13.1 | 153.7 ± 10.6 |
| Butyl-2-cyanoacrylate | 112.6 ± 8.9 | 124.2 ± 12.2 | 129.8 ± 14.8 | |
| Yield point strength (N) | Ethyl-2-cyanoacrylate | 119.8 ± 14.1 | 124.6 ± 16.8 | 129.7 ± 13 |
| Butyl-2-cyanoacrylate | 105.8 ± 8.9 | 112.6 ± 9.2 | 113 ± 11.4 | |
| Stiffness coefficient (N/mm) | Ethyl-2-cyanoacrylate | 171.7 ± 15.1 | 214.3 ± 28.2 | 236.2 ± 32.7 |
| Butyl-2-cyanoacrylate | 130.1 ± 19.3 | 158.2 ± 11 | 176.4 ± 6.7 | |
Results of the tibial tests: mean and standard deviation of the maximum load (N), yield point strength (N), and stiffness coefficient (N/mm), according to the adhesive used and the moment of euthanasia.
| Studied variable | Adhesive used | Moment of euthanasia | ||
|---|---|---|---|---|
| 60 days | 120 days | 180 days | ||
| Maximum load (N) | Ethyl-2-cyanoacrylate | 124.7 ± 10.5 | 133.8 ± 11.3 | 137.5 ± 9.8 |
| Butyl-2-cyanoacrylate | 111.4 ± 12.6 | 121.3 ± 16.6 | 125.1 ± 8.3 | |
| Yield point strength (N) | Ethyl-2-cyanoacrylate | 121.5 ± 6.3 | 129.6 ± 10.2 | 132.9 ± 11.1 |
| Butyl-2-cyanoacrylate | 108.4 ± 4.7 | 118.3 ± 9.8 | 119.8 ± 7.5 | |
| Stiffness coefficient (N/mm) | Ethyl-2-cyanoacrylate | 178.1 ± 10.7 | 222.5 ± 17.1 | 229.8 ± 17.9 |
| Butyl-2-cyanoacrylate | 144.3 ± 11.2 | 163.7 ± 12.9 | 170.4 ± 15.2 | |
Mean and standard deviation of the bone callus area (μm2), according to the adhesive used and the moment of euthanasia.
| Adhesive used | Moment of euthanasia | ||
|---|---|---|---|
| 60 days | 120 days | 180 days | |
| Ethyl-2-cyanoacrylate | 304,468 ± 72,101 | 334,785.2 ± 111,205 | 356,903.8 ± 98,003 |
| Butyl-2-cyanoacrylate | 102,448.1 ± 86,247 | 175,978.5 ± 144,632 | 195,925.4 ± 106,022 |
Mean and standard deviation of the cellularity observed (osteoblasts, osteoclasts, inflammatory cells; n−1), according to the adhesive used and the moment of euthanasia.
| Adhesive used | Cellularity | Moment of euthanasia | ||
|---|---|---|---|---|
| 60 days | 120 days | 180 days | ||
| Ethyl-2-cyanoacrylate | Osteoblasts | 251 ± 4 | 306 ± 23 | 280 ± 26 |
| Osteoclasts | 21 ± 7 | 37 ± 19 | 28 ± 11 | |
| Inflammatory cells | 115 ± 11 | 299 ± 24 | 187 ± 25 | |
| Butyl-2-cyanoacrylate | Osteoblasts | 221 ± 16 | 274 ± 31 | 255 ± 30 |
| Osteoclasts | 19 ± 8 | 25 ± 11 | 23 ± 12 | |
| Inflammatory cells | 227 ± 12 | 512 ± 28 | 461 ± 39 | |