Precision of CAPILLARYS 2 for the Detection of Hemoglobin Variants Based on Their Migration Positions.

Jean Riou1, Jessica Szuberski2, Christian Godart1, Henri Wajcman3, Jennifer L Oliveira2, James D Hoyer2, Josiane Bardakdjian-Michau1.   

Abstract

OBJECTIVES: In this report, we evaluated utility of the capillary electrophoresis (CE) migration position of the CAPILLARYS 2 CE instrument.
METHODS: The precision of this x-axis number was determined on a selection of common hemoglobin (Hb) variants (Hb S, Hb C, Hb D-Punjab, Hb E, Hb Hope), and the reproducibility of this number was evaluated by comparing the results obtained by two large reference laboratories on 81 Hb variants. Additionally, the CE migration position is given for a total of 409 Hb variants.
RESULTS: The x-axis migration position showed excellent intra- and interassay precision. Comparison of Hb variants seen by both laboratories showed that 83% had a difference in migration position of 1 unit or less. Only three rare Hb variants showed a difference of more than 2 units.
CONCLUSION: In summary, the CE migration position is a reproducible value and can be used as an aid in the identification of Hb variants. © American Society for Clinical Pathology, 2018. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com

Entities:  

Keywords:  CAPILLARYS 2; Capillary electrophoresis; Hemoglobinopathies

Mesh:

Substances:

Year:  2018        PMID: 29365076     DOI: 10.1093/ajcp/aqx148

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  1 in total

1.  Silent hemoglobin variant during capillary electrophoresis: A case report.

Authors:  Yanping Yuan; Xianghai Zhou; Leili Gao; Qian Ren; Linong Ji
Journal:  J Diabetes Investig       Date:  2020-03-18       Impact factor: 4.232

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.