Towards time-of-flight PET with a semiconductor detector.

Gerard Ariño-Estrada1, Gregory S Mitchell, Sun Il Kwon, Junwei Du, Hadong Kim, Leonard J Cirignano, Kanai S Shah, Simon R Cherry.   

Abstract

The feasibility of using Cerenkov light, generated by energetic electrons following 511 keV photon interactions in the semiconductor TlBr, to obtain fast timing information for positron emission tomography (PET) was evaluated. Due to its high refractive index, TlBr is a relatively good Cerenkov radiator and with its wide bandgap, has good optical transparency across most of the visible spectrum. Coupling an SiPM photodetector to a slab of TlBr (TlBr-SiPM) yielded a coincidence timing resolution of 620 ps FWHM between the TlBr-SiPM detector and a LFS reference detector. This value improved to 430 ps FWHM by applying a high pulse amplitude cut based on the TlBr-SiPM and reference detector signal amplitudes. These results are the best ever achieved with a semiconductor PET detector and already approach the performance required for time-of-flight. As TlBr has higher stopping power and better energy resolution than the conventional scintillation detectors currently used in PET scanners, a hybrid TlBr-SiPM detector with fast timing capability becomes an interesting option for further development.

Entities:  

Mesh:

Year:  2018        PMID: 29364135      PMCID: PMC5854393          DOI: 10.1088/1361-6560/aaaa4e

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  11 in total

Review 1.  The challenge of detector designs for PET.

Authors:  Thomas K Lewellen
Journal:  AJR Am J Roentgenol       Date:  2010-08       Impact factor: 3.959

2.  LaBr(3):Ce and SiPMs for time-of-flight PET: achieving 100 ps coincidence resolving time.

Authors:  Dennis R Schaart; Stefan Seifert; Ruud Vinke; Herman T van Dam; Peter Dendooven; Herbert Löhner; Freek J Beekman
Journal:  Phys Med Biol       Date:  2010-03-19       Impact factor: 3.609

3.  CHERENCUBE: concept definition and implementation challenges of a Cherenkov-based detector block for PET.

Authors:  I Somlai-Schweiger; S I Ziegler
Journal:  Med Phys       Date:  2015-04       Impact factor: 4.071

4.  Bismuth germanate coupled to near ultraviolet silicon photomultipliers for time-of-flight PET.

Authors:  Sun Il Kwon; Alberto Gola; Alessandro Ferri; Claudio Piemonte; Simon R Cherry
Journal:  Phys Med Biol       Date:  2016-09-02       Impact factor: 3.609

Review 5.  Update on time-of-flight PET imaging.

Authors:  Suleman Surti
Journal:  J Nucl Med       Date:  2014-12-18       Impact factor: 10.057

6.  Bright and ultra-fast scintillation from a semiconductor?

Authors:  Stephen E Derenzo; Edith Bourret-Courshesne; Gregory Bizarri; Andrew Canning
Journal:  Nucl Instrum Methods Phys Res A       Date:  2016-01-01       Impact factor: 1.455

7.  Hybrid Pixel-Waveform (HPWF) Enabled CdTe Detectors for Small Animal Gamma-Ray Imaging Applications.

Authors:  A Groll; K Kim; H Bhatia; J C Zhang; J H Wang; Z M Shen; L Cai; J Dutta; Q Li; L J Meng
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2016-11-01

8.  Parallax error in long-axial field-of-view PET scanners-a simulation study.

Authors:  Jeffrey P Schmall; Joel S Karp; Matt Werner; Suleman Surti
Journal:  Phys Med Biol       Date:  2016-07-01       Impact factor: 3.609

9.  Energy and coincidence time resolution measurements of CdTe detectors for PET.

Authors:  G Ariño; M Chmeissani; G De Lorenzo; C Puigdengoles; E Cabruja; Y Calderón; M Kolstein; J G Macias-Montero; R Martinez; E Mikhaylova; D Uzun
Journal:  J Instrum       Date:  2013-02       Impact factor: 1.415

10.  New-generation small animal positron emission tomography system for molecular imaging.

Authors:  Shiva Abbaszadeh; Craig S Levin
Journal:  J Med Imaging (Bellingham)       Date:  2017-01-12
View more
  10 in total

Review 1.  Innovations in Instrumentation for Positron Emission Tomography.

Authors:  Eric Berg; Simon R Cherry
Journal:  Semin Nucl Med       Date:  2018-03-12       Impact factor: 4.446

2.  Two-crossed-polarizers based optical property modulation method for ionizing radiation detection for positron emission tomography.

Authors:  Yuli Wang; Yingjie Li; Fei Yi; Junyu Li; Siwei Xie; Qiyu Peng; Jianfeng Xu
Journal:  Phys Med Biol       Date:  2019-07-05       Impact factor: 3.609

3.  First Cerenkov charge-induction (CCI) TlBr detector for TOF-PET and proton range verification.

Authors:  Gerard Ariño-Estrada; Gregory S Mitchell; Hadong Kim; Junwei Du; Sun Il Kwon; Leonard J Cirignano; Kanai S Shah; Simon R Cherry
Journal:  Phys Med Biol       Date:  2019-08-28       Impact factor: 3.609

4.  Roadmap toward the 10 ps time-of-flight PET challenge.

Authors:  Paul Lecoq; Christian Morel; John O Prior; Dimitris Visvikis; Stefan Gundacker; Etiennette Auffray; Peter Križan; Rosana Martinez Turtos; Dominique Thers; Edoardo Charbon; Joao Varela; Christophe de La Taille; Angelo Rivetti; Dominique Breton; Jean-François Pratte; Johan Nuyts; Suleman Surti; Stefaan Vandenberghe; Paul Marsden; Katia Parodi; Jose Maria Benlloch; Mathieu Benoit
Journal:  Phys Med Biol       Date:  2020-10-22       Impact factor: 3.609

Review 5.  Advances in Detector Instrumentation for PET.

Authors:  Andrea Gonzalez-Montoro; Muhammad Nasir Ullah; Craig S Levin
Journal:  J Nucl Med       Date:  2022-08       Impact factor: 11.082

6.  Study of Čerenkov Light Emission in the Semiconductors TlBr and TlCl for TOF-PET.

Authors:  Gerard Ariño-Estrada; Emilie Roncali; Aaron R Selfridge; Junwei Du; Jaroslaw Glodo; Kanai S Shah; Simon R Cherry
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2020-09-17

7.  Integration of polarization in the LUTDavis model for optical Monte Carlo simulation in radiation detectors.

Authors:  Carlotta Trigila; Emilie Roncali
Journal:  Phys Med Biol       Date:  2021-10-22       Impact factor: 4.174

8.  Development of TlBr detectors for PET imaging.

Authors:  Gerard Ariño-Estrada; Junwei Du; Hadong Kim; Leonard J Cirignano; Kanai S Shah; Simon R Cherry; Gregory S Mitchell
Journal:  Phys Med Biol       Date:  2018-07-03       Impact factor: 3.609

9.  Energy and electron drift time measurements in a pixel CCI TlBr detector with 1.3 MeV prompt-gammas.

Authors:  Gerard Ariño-Estrada; Hadong Kim; Junwei Du; Leonard J Cirignano; Kanai S Shah; Simon R Cherry
Journal:  Phys Med Biol       Date:  2021-02-02       Impact factor: 4.174

10.  High resolution detectors for whole-body PET scanners by using dual-ended readout.

Authors:  Zheng Liu; Ming Niu; Zhonghua Kuang; Ning Ren; San Wu; Longhan Cong; Xiaohui Wang; Ziru Sang; Crispin Williams; Yongfeng Yang
Journal:  EJNMMI Phys       Date:  2022-04-21
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.