Literature DB >> 29362583

Percutaneous mitral balloon valvuloplasty. Difficult mitral valve crossing.

Zbigniew Chmielak1, Marcin Demkow2, Jarosław Skowroński1, Paweł Tyczyński1, Dariusz Zakrzewski3, Adam Witkowski1.   

Abstract

Entities:  

Year:  2017        PMID: 29362583      PMCID: PMC5770871          DOI: 10.5114/aic.2017.71622

Source DB:  PubMed          Journal:  Postepy Kardiol Interwencyjnej        ISSN: 1734-9338            Impact factor:   1.426


× No keyword cloud information.

Introduction

Percutaneous mitral balloon valvuloplasty (PMBV) is the treatment of choice for significant mitral stenosis (MS). The inability to cross the mitral valve accounts for a non-negligible rate of unsuccessful PMBV. This may be caused by unfavorable septal puncture or extensive enlargement of the left atrium (LA), both resulting in insufficient support. Advanced valvular and subvalvular degeneration may further impede valve crossing. Surprisingly, only a few reports focus on this problem. In a study by Feldman, failure to cross the mitral valve occurred in 1.7% of cases [1]. Cribier et al. reported the occurrence of this failure at 2.6% [2]. We present a case of a successful re-attempt of PMBV done with the support of a veno-arterial loop after an unsuccessful mitral valve crossing with an Inoue balloon in the first procedure.

Case report

A 38-year-old woman with significant MS, history of ischemic stroke, and hypertension was admitted for PMBV. In echocardiography LA enlargement to 30.0 cm2, thickening of mitral leaflets and subvalvular apparatus, fusion of posterior commissure were observed. Mitral valve area (MVA) was 0.8 cm2 by the pressure half-time (PHT) formula. Trans-mitral gradient was 21.7/10.5 mm Hg maximum and mean, respectively. The standard PMBV with the Inoue balloon was initiated via the right femoral vein. Atrial septum puncture was performed under transesophageal echocardiography (TEE) guidance. In the case of LA enlargement the right atrium is pushed to the front. So, as in the typical place of puncture the ascending aorta might be present, we usually puncture the atrial septum slightly posteriorly. Then hemodynamic measurements were made. After having placed the balloon in the LA, all of the maneuvers aiming to cross the mitral valve orifice with the balloon failed. During the second approach, a modified PMBV technique was used with a transseptal veno-arterial loop serving as a rail for the balloon’s entry into the LV (Figure 1). The TEE-guided atrial septum puncture was performed again. The dedicated 6 Fr Swan-Ganz catheter, with 0.035 mm lumen, was introduced to the LA and next to the left ventricle (LV), through which a 300 mm long, 0.035 inch wire was advanced from the LA into the LV. Using the arterial access, a Multi-snare loop (PFM Medical, Germany) was advanced in a retrograde fashion through a 6 Fr-JR guide catheter into the ascending aorta and directed into the LV. Then the wire tip was caught with it and pulled out through the guide catheter and arterial sheath, completing the veno-arterial loop. The next steps of PMBV were done using the standard approach. The Inoue balloon was advanced over a 0.035 inch wire across the mitral valve and dilated to 28 mm. The procedure was successful. Pulmonary artery pressure decreased from 105/32/65 mm Hg to 58/25/37 mm Hg, systolic/diastolic/mean, respectively. The fluoroscopy time was 39 min.
Figure 1

Procedure. A – Wire and pigtail, both in the left ventricle, B – catching the wire’s tip with the multisnare loop, C – magnification of image B, D – veno-arterial loop, E – Inoue balloon’s crossing though the mitral valve, F – balloon dilatation

Procedure. A – Wire and pigtail, both in the left ventricle, B – catching the wire’s tip with the multisnare loop, C – magnification of image B, D – veno-arterial loop, E – Inoue balloon’s crossing though the mitral valve, F – balloon dilatation Trans-mitral gradient decreased to 13/5.3 mm Hg (maximum/mean) and MVA increased to 1.25 cm2 by the PHT formula and 1.6 cm2 by planimetry.

Discussion

Various techniques have been proposed to enable the anterograde trans-septal or retrograde non-transseptal approach (Stefanadis technique) for balloon entry into the LV [3]. The veno-arterial loop technique is used in different strategies such as closure of paravalvular leaks, ventricular septal defects or the transcatheter mitral valve-in-ring (TVIR) procedure, to offer better delivery support. Ateş et al. presented probably the first use of a veno-arterial loop for a challenging mitral valve crossing [4]. Technically, the wire was directed in an anterograde fashion from the LA through the LV into the ascending and descending aorta using a multipurpose catheter. Finally the loop was completed by snaring the common iliac artery and pulling it out from the sheath [4]. Thus, no multi-snare loop was used here, unlike in our case. Recently Nanjappa et al. reported the first series of three patients with difficult mitral valve crossings, where a veno-arterial loop was created and proved to be helpful [5]. In our opinion the crossing maneuver is safe and useful, but should be done gently.

Conclusions

A veno-arterial loop may be considered a bail-out technique in cases of a problematic crossing of the mitral valve orifice during PMBV.

Conflict of interest

The authors declare no conflict of interest.
  5 in total

1.  Percutaneous mechanical mitral commissurotomy with a newly designed metallic valvulotome: immediate results of the initial experience in 153 patients.

Authors:  A Cribier; H Eltchaninoff; R Koning; P C Rath; R Arora; A Imam; M El-Sayed; S Dani; G Derumeaux; J Benichou; C Tron; S Janorkar; G Pontier; B Letac
Journal:  Circulation       Date:  1999-02-16       Impact factor: 29.690

2.  Balloon mitral valvuloplasty: our experience with a modified technique of crossing the mitral valve in difficult cases.

Authors:  C N Manjunath; K H Srinivasa; C B Patil; H V Venkatesh; T S Bhoopal; C Dhanalakshmi
Journal:  Cathet Cardiovasc Diagn       Date:  1998-05

3.  Hemodynamic results, clinical outcome, and complications of Inoue balloon mitral valvotomy.

Authors:  T Feldman
Journal:  Cathet Cardiovasc Diagn       Date:  1994

4.  New technique for challenging cases of percutaneous balloon mitral valvuloplasty: The venoarterial looping.

Authors:  İsmail Ateş; Şeref Ulucan; Zeynettin Kaya; Mehmet Doğru; Hüseyin Katlandur; Ahmet Keser
Journal:  Anatol J Cardiol       Date:  2015-05       Impact factor: 1.596

5.  Case series: Difficult PTMC using novel technique of veno-arterial looping.

Authors:  Veena Nanjappa; K S Sadanand; K Santhosh; Harsha Basappa; C N Manjunath; Mohan H Nayak
Journal:  Indian Heart J       Date:  2016-12-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.