Abstract

OBJECTIVES: Intraoperative hairline longitudinal fractures were recently reported in association with distal radius volar plating. Our aim was to further analyze this newly described complication.
METHODS: A retrospective radiographic and chart review was performed on 225 patients who underwent distal radius plating between June 2013 and June 2015. The Acu-Loc/Acu-Loc2© plating system (Acumed, Hillsboro, OR, USA) was used in 208 cases, and the VariAx© plating system (Stryker, Kalamazoo, MI, USA) was used in 17 cases. Three independent reviewers performed a blind evaluation of all relevant radiographs for the occurrence of longitudinal fractures around the plate, and validity was considered only when there was agreement among all three of them.
RESULTS: Hairline longitudinal fractures were identified in 57 cases (25%), 55 with the Acu-Loc/Acu-Loc2© system and 2 with the VariAx© system. All fractures occurred with volar plating. Fracture occurrence was associated with age over 59 years, female gender, extra-articular fractures, and the use of Hexalobe screws (Acu-Loc/Acu-Loc2© system).
CONCLUSIONS: We believe that the source of fracture occurrence lies within the screw design and that better screw design and possibly tapping in patients at risk may reduce the occurrence of intraoperative hairline longitudinal fractures. Further clinical and biomechanical research is needed to better understand this newly reported complication.

Entities:  

Keywords:  Distal radius fractures; Hairline longitudinal fractures; Hardware-related complication; Volar plating

Mesh:

Year:  2018        PMID: 29356843     DOI: 10.1007/s00256-018-2877-9

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  11 in total

Review 1.  Complications of Distal Radius Fixation.

Authors:  Dennis S Lee; Douglas R Weikert
Journal:  Orthop Clin North Am       Date:  2016-04       Impact factor: 2.472

Review 2.  The role of cortical bone and its microstructure in bone strength.

Authors:  Peter Augat; Sandra Schorlemmer
Journal:  Age Ageing       Date:  2006-09       Impact factor: 10.668

3.  Complications after volar locking plate fixation of distal radius fractures.

Authors:  N A Johnson; L Cutler; J J Dias; A S Ullah; C J Wildin; B Bhowal
Journal:  Injury       Date:  2013-10-14       Impact factor: 2.586

4.  A new complication in volar locking plating of the distal radius: longitudinal fractures of the near cortex.

Authors:  Tahir Sadık Sügün; Yusuf Gürbüz; Kemal Özaksar; Tulgar Toros; Emin Bal; Murat Kayalar
Journal:  Acta Orthop Traumatol Turc       Date:  2016       Impact factor: 1.511

5.  Complications following internal fixation of unstable distal radius fracture with a palmar locking-plate.

Authors:  Rohit Arora; Martin Lutz; Alfred Hennerbichler; Dietmar Krappinger; David Espen; Markus Gabl
Journal:  J Orthop Trauma       Date:  2007-05       Impact factor: 2.512

Review 6.  Risk assessment tools to identify women with increased risk of osteoporotic fracture: complexity or simplicity? A systematic review.

Authors:  Katrine Hass Rubin; Teresa Friis-Holmberg; Anne Pernille Hermann; Bo Abrahamsen; Kim Brixen
Journal:  J Bone Miner Res       Date:  2013-08       Impact factor: 6.741

Review 7.  Complications following volar locking plate fixation for distal radial fractures: a systematic review.

Authors:  A Bentohami; K de Burlet; N de Korte; M P J van den Bekerom; J C Goslings; N W L Schep
Journal:  J Hand Surg Eur Vol       Date:  2013-11-21

8.  Functional results and complications following locking palmar plating on the distal radius: a retrospective study.

Authors:  M Hakimi; P Jungbluth; J Windolf; M Wild
Journal:  J Hand Surg Eur Vol       Date:  2009-08-17

9.  Long-term outcome of tuberoplasty for irreparable massive rotator cuff tears: is tuberoplasty really applicable?

Authors:  Jung Gwan Park; Nam Su Cho; Jong Hoon Song; Jong Hun Baek; Yong Girl Rhee
Journal:  J Shoulder Elbow Surg       Date:  2015-10-09       Impact factor: 3.019

10.  European guidance for the diagnosis and management of osteoporosis in postmenopausal women.

Authors:  J A Kanis; E V McCloskey; H Johansson; C Cooper; R Rizzoli; J-Y Reginster
Journal:  Osteoporos Int       Date:  2012-10-19       Impact factor: 4.507

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.