Literature DB >> 29352689

Morbidity and Safety of Iliac Crest Reference Array Pins in Navigated Total Hip Arthroplasty: A Prospective Cohort Study.

Anton P Lambers1, Xiang G Salim1, Robert Jennings1, Andrew T Bucknill1.   

Abstract

BACKGROUND: Navigated total hip arthroplasty (THA) can employ intra-osseous pins through a separate incision to secure reference arrays to the iliac crest. This study is the first to investigate the consequences of pin use in THA in vivo.
METHODS: A prospective, consecutive series of 43 patients presenting for navigated THA were included. Two temporary 125 × 4 mm Schanz screws were inserted into the iliac crest for the attachment of a reference array. Telephone follow-up occurred at 6 and 12 weeks post-operatively. Patients were asked about pain, interference with daily activities, how often the wound was noticed, and duration of discomfort. Patient body mass index was recorded.
RESULTS: The follow-up rate was 100%. Pin site pain at any time post-operatively was reported by 24 patients (56%). This improved to 30%, 9%, and 2% at 3, 6, and 12 weeks, respectively. On average, pain lasted for 16 days total. The most common complaints after pain were clothing discomfort (23%), pain when wearing a belt (12%), or pain when mobilizing (9%). For the majority (98%) of patients, all symptoms had resolved by 12 weeks. There was no nerve injury, pin site fracture, infection, or screw breakage. Patients with body mass index greater than 30 kg/m2 were up to 3 times more likely to experience pin site pain (P = .05), and had a longer duration of pain (P = .04).
CONCLUSION: Surgeons and patients should be aware that using navigational pins for array fixation carries low complication rates but often will cause pain and irritation that resolves in the short term.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  computer-assisted navigation; iliac crest; morbidity; pain; pin site; safety; total hip arthroplasty; total hip replacement

Mesh:

Year:  2017        PMID: 29352689     DOI: 10.1016/j.arth.2017.12.032

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  3 in total

1.  Can a simple iPad app improve C-arm based component position in anterior THA?

Authors:  Ulrich Bechler; Bernhard Springer; Kilian Rueckl; Tim Rolvien; Friedrich Boettner
Journal:  Arch Orthop Trauma Surg       Date:  2021-02-13       Impact factor: 3.067

2.  A simple method to minimize leg length discrepancy in hip hemiarthroplasty.

Authors:  Guodong Wang; Ai Guo; Fei Yu; Bo Yang; Haomiao Yu; Naicheng Diao; Lifeng Ma; Hua Qiang; Erhong Zhao
Journal:  Clin Interv Aging       Date:  2019-09-05       Impact factor: 4.458

3.  Retained pelvic pin site debris after navigated total hip replacement: Masquerading as an early-stage chondrosarcomatous lesion.

Authors:  A P Kurmis
Journal:  J Postgrad Med       Date:  2020 Oct-Dec       Impact factor: 1.476

  3 in total