The primary strategy to avoid mother-to-child transmission of human immunodeficiency virus (HIV) through breastfeeding is administration of highly active antiretroviral therapy (HAART) to HIV-positive pregnant women. Because significant changes in the pharmacokinetics of antiretroviral (ARV) drugs occur during pregnancy, quantifying HAART and the viral load in breast milk in this population is essential. Here, we developed an analytical assay for the simultaneous quantification of four ARV drugs in breast milk using ultra-performance liquid chromatography coupled to tandem mass spectrometry. We validated this method following Mexican and international guidelines. ARV drugs. We extracted the ARV drugs from 200 μL samples of breast milk and detected these drugs in a triple quadrupole mass spectrometer with positive electrospray ionization. The validated concentration ranges (ng/mL) for zidovudine, lamivudine, lopinavir, and ritonavir were 12.5-750, 50-2500, 100-5000 and 5 to 250, respectively. Additionally, the absolute recovery percentages (and matrix effects) were 91.4 (8.39), 88.78 (28.75), 91.38 (11.77) and 89.78 (12.37), respectively. We determined that ARV drugs are stable for 24 h at 8°C and 24°C for 15 days at -80°C. This methodology had the capacity for simultaneous detection; separation; and accurate, precise quantification of ARV drugs in human breast milk samples according to Mexican standard laws and United States Food and Drug Administration guidelines.Entities:
Mesh:
Substances:
Year: 2018 PMID: 29351333 PMCID: PMC5774716 DOI: 10.1371/journal.pone.0191236
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Chromatographic conditions: Mobile phase gradient program.
| Time (minutes) | % phase A | % phase B | Speed of gradient change (Curve type |
|---|---|---|---|
| 0.15 | 90 | 10 | 2 |
| 0.70 | 60 | 40 | 2 |
| 2.0 | 10 | 90 | 6 |
| 2.5 | 90 | 10 | 11 |
*Curve 2: concave type. Curve 6: lineal type.
Transition ions and optimal conditions to obtain relative abundance of product ions.
| Parameter | LMV | ZDV | LPV | RTV | SMV |
|---|---|---|---|---|---|
| 230.15 → 112.08 | 268.18 → 127.10 | 629.55 → 447.35 | 721.50 → 296.20 | 419.40 → 199.20 | |
| 0.161 | 0.161 | 0.105 | 0.105 | 0.105 | |
| 15 | 10 | 20 | 23 | 20 | |
| 15 | 10 | 15 | 15 | 15 | |
| 1.02e7 | 1.18e6 | 6.11e6 | 4.55e6 | 3.41e5 |
m/z: mass/charge relationship; kV: kilovolts.
Nominal concentrations of each calibration curve and quality control standards in breast milk samples.
| POINTS | ZDV | LMV | LPV | RTV |
|---|---|---|---|---|
| 12.5 | 50 | 100 | 5 | |
| 25 | 100 | 200 | 10 | |
| 50 | 200 | 400 | 20 | |
| 125 | 500 | 1000 | 50 | |
| 250 | 1000 | 2000 | 100 | |
| 500 | 2000 | 4000 | 200 | |
| 750 | 3000 | 6000 | 300 | |
| 75 | 300 | 600 | 30 | |
| 425 | 1700 | 3400 | 170 | |
| 625 | 2500 | 5000 | 250 |
LQC: Low quality control standard. MQC: Medium quality control standard. HQC: High quality control standard
Fig 1Representative ion chromatograms of separation and retention time (minutes) of each ARV drug and IS: 0.29, 0.62, 1.61, 1.71 and 2.11 for LMV, ZDV, RTV, LPV, and SMV, respectively, corresponding only to LQC (ng/mL) of each ARV.
Total run time was 2.5 minutes, with an ACQUITY UPLC BEH C 18 column, formic acid and ACN gradient as mobile phase.
Fig 2Correlation between nominal concentration vs measured concentration (ng/mL) for intra-day (n = 3) and inter-day (n = 3) validation of each ARV.
Accuracy and precision for the quantification of antiretroviral in breast milk.
| QCs (ng/mL) | Inter-day (measured concentration) | Intra-day (measured concentration) | ||||||
|---|---|---|---|---|---|---|---|---|
| Mean | SD | Precision (%CV) | Accuracy (%) | Mean | SD | Precision (%CV) | Accuracy (%) | |
| 72.62 | 7.63 | 10.51 | 96.84 | 74.16 | 8.54 | 11.52 | 91.86 | |
| 408.20 | 56.72 | 13.89 | 89.05 | 435.47 | 30.16 | 6.91 | 94.45 | |
| 609.40 | 53.37 | 8.75 | 92.61 | 596.87 | 38.24 | 6.40 | 92.61 | |
| 307.41 | 17.75 | 5.77 | 94.86 | 309.18 | 22.76 | 7.36 | 93.99 | |
| 1761.26 | 217.89 | 12.37 | 88.02 | 1841.41 | 120.57 | 6.54 | 90.10 | |
| 2562.23 | 232.23 | 9.06 | 92.14 | 2611.72 | 109.65 | 4.19 | 93.84 | |
| 647.52 | 74.27 | 11.47 | 89.70 | 619.84 | 27.47 | 4.43 | 95.11 | |
| 3847.57 | 266.31 | 6.92 | 86.84 | 3689.62 | 212.61 | 5.76 | 91.49 | |
| 5942.90 | 603.76 | 10.15 | 5610.38 | 232.20 | 4.13 | 86.69 | ||
| 30.20 | 2.99 | 9.91 | 92.38 | 29.66 | 0.94 | 3.17 | 97.74 | |
| 194.26 | 19.55 | 10.06 | 85.22 | 193.23 | 10.13 | 5.24 | 86.33 | |
| 307.35 | 31.32 | 10.19 | 296.34 | 11.28 | 3.80 | |||
LQC: Low quality control standard. MQC: Medium quality control standard. HQC: High quality control standard
Short-term stability of antiretroviral drugs in processed samples.
| Storage condition | Percentage stability | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| QC level | ZDV | LMV | LPV | RTV | |||||
| Precision (%CV) | Accuracy (%) | Precision (%CV) | Accuracy (%) | Precision (%CV) | Accuracy (%) | Precision (%CV) | Accuracy (%) | ||
| LQC | 4.87 | 95.50 | 2.63 | 90.00 | 4.30 | 95.72 | 4.49 | 89.69 | |
| MQC | 2.75 | 95.83 | 9.88 | 91.35 | 3.43 | 92.92 | 2.90 | 91.25 | |
| HQC | 3.18 | 94.68 | 2.78 | 94.16 | 7.90 | 94.43 | 6.67 | 90.83 | |
| LQC | 9.11 | 91.10 | 3.33 | 89.86 | 8.82 | 92.27 | 9.73 | 86.07 | |
| MQC | 4.64 | 91.24 | 3.84 | 91.40 | 3.79 | 88.37 | 4.45 | 86.68 | |
| HQC | 9.19 | 85.65 | 4.98 | 89.81 | 7.24 | 87.29 | 6.56 | ||
| LQC | 2.02 | 95.96 | 1.33 | 98.00 | 4.59 | 85.64 | 25.93 | ||
| MQC | 7.91 | 1.46 | 97.08 | 4.15 | 95.34 | 4.97 | 96.12 | ||
| HQC | 1.85 | 88.56 | 1.27 | 96.69 | 3.40 | 95.54 | 1.59 | 95.07 | |
| LQC | 15.00 | 88.23 | 1.90 | 94.36 | 6.86 | 94.32 | 6.33 | 91.18 | |
| MQC | 1.39 | 94.88 | 1.19 | 96.52 | 3.55 | 91.06 | 4.42 | 97.00 | |
| HQC | 6.20 | 92.97 | 3.04 | 96.68 | 4.38 | 93.73 | 2.74 | 93.10 | |
| LQC | 7.08 | 95.37 | 2.37 | 94.26 | 0.83 | 85.69 | 1.60 | ||
| MQC | 2.11 | 97.41 | 1.90 | 95.95 | 1.27 | 89.77 | 4.55 | 96.31 | |
| HQC | 2.08 | 96.73 | 3.10 | 97.73 | 2.90 | 89.54 | 5.68 | 95.70 | |
| LQC | 8.25 | 87.04 | 2.58 | 96.81 | 5.51 | 7.82 | 94.20 | ||
| MQC | 4.87 | 90.29 | 4.11 | 96.74 | 9.44 | 9.44 | 91.99 | ||
| HQC | 4.54 | 89.79 | 1.09 | 94.98 | 5.21 | 4.02 | 88.17 | ||
| LQC | 2.35 | 1.04 | 93.56 | 2.18 | 1.10 | 89.27 | |||
| MQC | 3.89 | 90.31 | 1.23 | 98.45 | 7.00 | 89.86 | 2.19 | 91.90 | |
| HQC | 4.75 | 87.57 | 2.91 | 97.37 | 2.36 | 4.90 | |||
*After three freeze-thaw cycles at -84°C.
**After 15 days at -84°C.