Accuracy and efficiency of published film dosimetry techniques using a flat-bed scanner and EBT3 film.

E Spelleken1, S B Crowe2,3, B Sutherland4, C Challens5, T Kairn3,6.   

Abstract

Gafchromic EBT3 film is widely used for patient specific quality assurance of complex treatment plans. Film dosimetry techniques commonly involve the use of transmission scanning to produce TIFF files, which are analysed using a non-linear calibration relationship between the dose and red channel net optical density (netOD). Numerous film calibration techniques featured in the literature have not been independently verified or evaluated. A range of previously published film dosimetry techniques were re-evaluated, to identify whether these methods produce better results than the commonly-used non-linear, netOD method. EBT3 film was irradiated at calibration doses between 0 and 4000 cGy and 25 pieces of film were irradiated at 200 cGy to evaluate uniformity. The film was scanned using two different scanners: The Epson Perfection V800 and the Epson Expression 10000XL. Calibration curves, uncertainty in the fit of the curve, overall uncertainty and uniformity were calculated following the methods described by the different calibration techniques. It was found that protocols based on a conventional film dosimetry technique produced results that were accurate and uniform to within 1%, while some of the unconventional techniques produced much higher uncertainties (> 25% for some techniques). Some of the uncommon methods produced reliable results when irradiated to the standard treatment doses (< 400 cGy), however none could be recommended as an efficient or accurate replacement for a common film analysis technique which uses transmission scanning, red colour channel analysis, netOD and a non-linear calibration curve for measuring doses up to 4000 cGy when using EBT3 film.

Entities:  

Keywords:  Dosimetry; EBT3; Flat-bed scanner; Gafchromic film; Radiotherapy

Mesh:

Year:  2018        PMID: 29344782     DOI: 10.1007/s13246-018-0620-4

Source DB:  PubMed          Journal:  Australas Phys Eng Sci Med        ISSN: 0158-9938            Impact factor:   1.430


  2 in total

1.  A focused very high energy electron beam for fractionated stereotactic radiotherapy.

Authors:  Kristoffer Svendsen; Diego Guénot; Jonas Björklund Svensson; Kristoffer Petersson; Anders Persson; Olle Lundh
Journal:  Sci Rep       Date:  2021-03-12       Impact factor: 4.379

2.  Probability Distribution of Pixel Intensities of EBT3 Films and its Application in the Correction of Uncertainty Budget.

Authors:  Rahul Kumar Chaudhary; Munir Pathan; Rajesh Kumar; S D Sharma; B K Sapra
Journal:  J Med Phys       Date:  2021-05-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.