| Literature DB >> 29341427 |
Tianhua Wang1, Kong Meng Hoi1, Henning Stöckmann2, Corrine Wan1, Lyn Chiin Sim1, Noor Hayati Bte Kamari Shi Jie Tay1, Ce Huang Poo1, Susanto Woen1, Yuangsheng Yang1, Peiqing Zhang1, Pauline M Rudd1,2.
Abstract
Robust plate based antibody glycan analysis platforms are urgently needed for biopharmaceutical development and manufacturing as well as for clinical biomarker research. A 96-well plate based workflow has been developed to analyze both intact IgG antibodies and released N-glycans using an Orbitrap Fusion Mass Spectrometer and an LC/MS method on the Waters UNIFI platform. Here, such a workflow including protein A purification, PNGaseF digestion, 2-AB labeling, and SPE clean-up is described. The measured IgG glycan profile is consistent with that obtained from non-plate based method and commercial kit and has the advantage of less hands-on time. Also the application of the workflow in cell culture monitoring and clonal selection work is demonstrated. Apart from checking the major glycan structure changes among clones, post translational modifications (PTMs) such as C-terminal lysine residue clipping and N-terminal pyroglutamic acid formation can also be deduced from the workflow.Entities:
Keywords: cell culture; clonal selection; intact IgG; liquid chromatography mass spectrometry (LCMS); paired glycoforms
Mesh:
Substances:
Year: 2018 PMID: 29341427 DOI: 10.1002/biot.201700185
Source DB: PubMed Journal: Biotechnol J ISSN: 1860-6768 Impact factor: 4.677