Dear Editor, We read the article about Transcranial Doppler (TCD) for intensivists [1]. Although not a novel ultrasound technique, in particular the “blind” o non-imaging TCD (bTCD), authors´ efforts to promote some basic applications of the Duplex technique (transcranial color-coded duplex sonography, TCCS) are remarkable. However, some technical points and assertions are dubious and/or incorrect, as noted below: In the first place, regarding the midline shift (MLS) measurement technique by TCCS, (A-B)/2 is well-studied and validated against CT [2]. While proposed by authors’ as an “internal standard” [1], as shown in Fig. 1 of the original article [1], measuring the distance to the contralateral cranial bone is not described in the original technique, it is unnecessary and adds complexity; thus, it should not be taken into account, as is the case with the mentioned “C and D” technique. To the authors´ knowledge, whether methodologically correct or not, there are no study validating either of them. Practitioners should be aware that the MLS measurement by TCCS is not reliable in the presence of bone defects (like decompressive craniectomy or skull fractures), temporal cephalohematoma, or changes in intracranial anatomy secondary to trauma [3], citing the most common examples observed in daily practice. Particularly in patients with a decompressive craniectomy, an alternative MLS measurement technique is well validated against CT [4].
Fig. 1

(corresponding to b and c of Fig. 2 [1]). Note the different and confusing nomenclature regarding “mean velocities”. As depicted from the trace of the envelope of the Doppler spectra (yellow arrows), time-averaged maximum velocity is recorded, namely, TAV (time-averaged velocity) in (a) and TAP (time-averaged peak velocity) in (b). There is also no doubt in a that is TAP, because pulsatility index (PI) is calculated using this value [peak systolic velocity (PSV)-end-diastolic velocity (EDV)/TAV]. Time-averaged mean velocity is not recorded in (a), but is shown in (b) as TAM, traced in the middle of the Doppler spectra (white arrow). In TCCS, time-averaged maximum or peak velocities are the “mean” velocities that should be considered. The correct Lindegaard Index (Middle cerebral artery TAP/internal carotid artery TAP) in this case is 123/58, equal to 2.1 (corresponding to hyperemia if considered independently). It is thus clearly incorrect to use different “mean velocities” when calculating the LI, such as TAP/TAMEAN. Note: the waveform in (b) is consistent with an external carotid artery flow, given its sharp systolic upstroke, high-resistance velocity profile, and early diastolic notch (another mistake that should be taken into account)

(corresponding to b and c of Fig. 2 [1]). Note the different and confusing nomenclature regarding “mean velocities”. As depicted from the trace of the envelope of the Doppler spectra (yellow arrows), time-averaged maximum velocity is recorded, namely, TAV (time-averaged velocity) in (a) and TAP (time-averaged peak velocity) in (b). There is also no doubt in a that is TAP, because pulsatility index (PI) is calculated using this value [peak systolic velocity (PSV)-end-diastolic velocity (EDV)/TAV]. Time-averaged mean velocity is not recorded in (a), but is shown in (b) as TAM, traced in the middle of the Doppler spectra (white arrow). In TCCS, time-averaged maximum or peak velocities are the “mean” velocities that should be considered. The correct Lindegaard Index (Middle cerebral artery TAP/internal carotid artery TAP) in this case is 123/58, equal to 2.1 (corresponding to hyperemia if considered independently). It is thus clearly incorrect to use different “mean velocities” when calculating the LI, such as TAP/TAMEAN. Note: the waveform in (b) is consistent with an external carotid artery flow, given its sharp systolic upstroke, high-resistance velocity profile, and early diastolic notch (another mistake that should be taken into account)
Fig. 2

a Transcervical window, phased-array probe. Note that the distal internal carotid artery is insonated and that angle correction is not needed in pulsed-wave Doppler. b Transforaminal window, phased- array probe. Note the inverted V configuration of the posterior circulation on color Doppler imaging (coded blue, indicating that blood is moving away from the transducer), depicted by both vertebral arteries (VA) and the basilar artery (BA), showing also the corresponding spectral Doppler on the inferior channel. F: foramen magnum; VA vertebral artery; BA basilar artery. c Transorbital window, phased-array probe. G: ocular globe

Second, when moving from a bTCD technique to the Duplex technique, practitioners must be aware of the “mean velocities” recorded by the ultrasound machine: time-averaged maximum velocity, known as TAMAX or TAP and time-averaged mean velocity, also known as TAMEAN or TAMV. While both are “mean” velocities, TAMEAN is approximately half the TAMAX [5]. Since in TCCS, the velocity considered is the TAMAX [5], using TAMEAN instead of TAP leads to underestimating velocities. This is clearly evidenced in Fig. 2 [1], where in the TCCS image, TAP is correctly used, but in the transcervical insonation, TAMV is used instead of TAP. Indeed, TAP should be compared when the Lindegaard Index (LI) is used, but comparing TAMAX/TAMEAN as is performed in Fig. 2 is an obvious mistake and readers need to be cautioned from making the same error. The actual LI in this case is 2.1, which indicates hyperemia (Fig. 1). According to this now well-performed TCD ratio, the angiographic finding of vasospasm was fortuitous, at least if this index is used independently [6]. In addition, transcervical insonation should be performed with the same phased-array probe to observe the “distal” extracranial internal carotid artery (ICA)—TAP (Fig. 2a) [7]. It should be noted that this segment is not assessed with the linear probe as shown in Fig. 2c of the original article. In addition, large correction angles (60°) result when a linear transducer is used and this must be especially avoided when comparing middle cerebral artery (MCA)/ICA TAP. Thus, the Doppler correction angle should not be used [8]. As noted, transcervical insonation should be a basic part of point-of-care ultrasound (POCUS)-TCD, at least if vasospasm evaluation is considered. a Transcervical window, phased-array probe. Note that the distal internal carotid artery is insonated and that angle correction is not needed in pulsed-wave Doppler. b Transforaminal window, phased- array probe. Note the inverted V configuration of the posterior circulation on color Doppler imaging (coded blue, indicating that blood is moving away from the transducer), depicted by both vertebral arteries (VA) and the basilar artery (BA), showing also the corresponding spectral Doppler on the inferior channel. F: foramen magnum; VA vertebral artery; BA basilar artery. c Transorbital window, phased-array probe. G: ocular globe Third, to the best of our knowledge, we are not aware of any guidelines that recommend TCD as a screening tool for further indication of an ancillary test to confirm the diagnosis of brain death. When determining the presence of cerebral circulatory arrest (CCA), many countries around the world accept this tool as an ancillary test to confirm the clinical diagnosis of brain death [9]. For example, there are formal TCD guidelines in Latin-American addressing this issue [10, 11]. For this indication, accepted TCD-CCA criteria for both “anterior” and “posterior” cerebral arterial circulation must be registered [12, 13]. Thus, intuitively, the transtemporal window is not enough for this indication. As a point-of-care application, transforaminal window should also be considered a basic window, at least if a CCA application is proposed (Fig. 2b). Transorbital (Fig. 2c) and transcervical (Fig. 2a) are also useful (although not fully accepted) in some actual patients to determine CCA, in particular when intracranial arterial flows are not detected on first examination, due to inadequate bone insonation windows, for example (observed in at least 25% of the patients) [13]. Regarding Doppler CCA criteria, the oscillating flow, although a biphasic flow, needs to be clearly differentiated from a high-resistance biphasic flow with a net forward flow (Fig. 3). In doubtful cases, always correlating with the clinical signs of brain death, modifications of the waveforms with interventions, such as osmotic therapy, may allow practitioners to discard the CCA diagnosis given the reversibility of the case on follow-up examinations.
Fig. 3

a High-resistance biphasic flow, with a net forward flow, not compatible with cerebral circulatory arrest. b Oscillating flow, with a net flow of 0, corresponding to cerebral circulatory arrest. S systole; D diastole

a High-resistance biphasic flow, with a net forward flow, not compatible with cerebral circulatory arrest. b Oscillating flow, with a net flow of 0, corresponding to cerebral circulatory arrest. S systole; D diastole Finally, velocities and indices (e.g., pulsatility index) are highly variable, resulting from physiologic (arousal, for example) to pathologic conditions (e.g., raising intracranial pressure) (Tables 1 and 2). Thus, caution should be exercised when interpreting TCD findings, which should always be considered within a multimodality monitoring, and not in isolation. The phrase “trends are your friend” is highly applicable when interpreting TCD velocities and indices.
Table 1

Physiologic and pathologic conditions that can modify TCCS flow velocities [3]

Increase
 Hyperemia
 Fever, anemia, high cardiac output, arterial hypertension
 Vasospasm
 Intracranial arterial stenosis (for example atherosclerotic plaque)
 Hypercapnia
 Bacterial meningitis
 Pre-eclampsia
Decrease
 Raised intracranial pressure
 Decreased cerebral perfusion pressure
 Cerebral circulatory arrest
 Hypocapnia
 Hypothermia
 Wrong insonation angle
Table 2

Physiologic and pathologic conditions that can modify TCCS Doppler indices [3]

Increase
 Raised intracranial pressure
 Decreased cerebral perfusion pressure
 Hypocapnia
 Hypothermia
 Cerebral circulatory arrest
 Hyperviscosity
 Intracranial artery occlusion
 Advanced age (vessel stiffness)
Decrease
 Hyperemia
 Anemia, fever, high cardiac output, arterial hypertension
 Hypercapnia
 Vasospasm
 Intracranial artery stenosis
 Arteriovenous malformation
 Rewarming following hypothermia
Physiologic and pathologic conditions that can modify TCCS flow velocities [3] Physiologic and pathologic conditions that can modify TCCS Doppler indices [3] In conclusion, POCUS TCD is not a perfect technique. Many aspects (technical and interpretative) should be considered to obtain a reliable TCD exam. In addition, for the reasons explained above, TCCS should not be limited to transtemporal windows, since transforaminal, transcervical, and transorbital windows have a defined role in basic TCD applications. The entire TCCS exam is performed with the same phased-array probe, based on the simplicity of POCUS, without the need of formal TCD examinations or dedicated machines, as happens with most (if not all) POCUS applications in the ICU. It is clear that a TCD-training curricula is mandatory to fulfill intensivists’ needs.
  13 in total

1.  Brazilian guidelines for the application of transcranial ultrasound as a diagnostic test for the confirmation of brain death.

Authors:  Marcos C Lange; Viviane H F Zétola; Maramélia Miranda-Alves; Carla H C Moro; Carlos E Silvado; Daniela L G Rodrigues; Edilene Guiotti di Gregorio; Gisele S Silva; Jamary Oliveira-Filho; Marco T A Perdatella; Octávio M Pontes-Neto; Soraia R C Fábio; Wagner M Avelar; Gabriel R de Freitas
Journal:  Arq Neuropsiquiatr       Date:  2012-05       Impact factor: 1.420

2.  Dislocation of the third ventricle due to space-occupying stroke evaluated by transcranial duplex sonography.

Authors:  G Seidel; T Gerriets; M Kaps; U Missler
Journal:  J Neuroimaging       Date:  1996-10       Impact factor: 2.486

3.  Volumetric blood flow measurement using Doppler ultrasound: concerns about the technique.

Authors:  Pablo Blanco
Journal:  J Ultrasound       Date:  2015-03-18

Review 4.  Applications of Transcranial Color-Coded Sonography in the Emergency Department.

Authors:  Pablo Blanco; Michael Blaivas
Journal:  J Ultrasound Med       Date:  2017-02-27       Impact factor: 2.153

5.  Variability in brain death determination in europe: looking for a solution.

Authors:  Giuseppe Citerio; Ilaria Alice Crippa; Alfio Bronco; Alessia Vargiolu; Martin Smith
Journal:  Neurocrit Care       Date:  2014-12       Impact factor: 3.210

6.  Usefulness of transcranial echography in patients with decompressive craniectomy: a comparison with computed tomography scan.

Authors:  Anselmo Caricato; Vittorio Mignani; Maria Grazia Bocci; Mariano Alberto Pennisi; Claudio Sandroni; Alessandra Tersali; Alessandra Antonaci; Chiara de Waure; Massimo Antonelli
Journal:  Crit Care Med       Date:  2012-06       Impact factor: 7.598

7.  Latin American consensus on the use of transcranial Doppler in the diagnosis of brain death.

Authors: 
Journal:  Rev Bras Ter Intensiva       Date:  2014 Jul-Sep

8.  Practice standards for transcranial Doppler ultrasound: part I--test performance.

Authors:  Andrei V Alexandrov; Michael A Sloan; Lawrence K S Wong; Colleen Douville; Alexander Y Razumovsky; Walter J Koroshetz; Manfred Kaps; Charles H Tegeler
Journal:  J Neuroimaging       Date:  2007-01       Impact factor: 2.486

Review 9.  Transcranial Doppler ultrasound: a review of the physical principles and major applications in critical care.

Authors:  Jawad Naqvi; Kok Hooi Yap; Gulraiz Ahmad; Jonathan Ghosh
Journal:  Int J Vasc Med       Date:  2013-12-12

Review 10.  Point-of-care transcranial Doppler by intensivists.

Authors:  Vincent Issac Lau; Robert Thomas Arntfield
Journal:  Crit Ultrasound J       Date:  2017-10-13
View more
  1 in total

1.  Personalising cardiovascular network models in pregnancy: A two-tiered parameter estimation approach.

Authors:  Jason Carson; Lynne Warrander; Edward Johnstone; Raoul van Loon
Journal:  Int J Numer Method Biomed Eng       Date:  2020-01-13       Impact factor: 2.648

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.