The repair of a type Ia endoleak following thoracic endovascular aortic repair using a stented elephant trunk procedure.

Rui-Dong Qi1, Jun-Ming Zhu2, Yong-Min Liu1, Lei Chen1, Cheng-Nan Li1, Xiao-Yan Xing1, Li-Zhong Sun1.   

Abstract

BACKGROUND: Type Ia endoleaks are not uncommon complications that occur after thoracic endovascular aortic repair (TEVAR). Because aortic arch vessels prevent extension of the landing zone, it is very difficult to manipulate a type Ia endoleak using an extension cuff or stent-graft, especially when the aortic arch is involved. Here, we retrospectively review our experience of surgical treatment of type Ia endoleak after TEVAR using a stented elephant trunk procedure.
METHODS: From July 2010 to August 2016, we treated 17 patients diagnosed with a type Ia endoleak following TEVAR using stented elephant trunk procedure. The mean age of our patients was 52 ± 8 years. The mean interval between TEVAR and the open surgical repair was 38 ± 43 months.
RESULTS: All cases of type Ia endoleak (100%) were repaired successfully. There were no in-hospital deaths. One case required reintubation and continuous renal replacement therapy due to renal failure; this patient recovered smoothly before discharge. One other patient suffered a stroke and renal failure and did not fully recover following discharge, or follow-up. During follow-up, there were 3 deaths.
CONCLUSIONS: Acceptable results were obtained using a stented elephant trunk procedure in patients with a type Ia endoleak after TEVAR. This technique allowed us to repair the proximal aortic arch lesions, surgically correct the type Ia endoleak, and promote false lumen thrombosis in the distal aorta. Implantation of a stented elephant trunk, with or without a concomitant aortic arch procedure, is an alternative approach for this type of lesion.
Copyright © 2017 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  endoleak; stented elephant trunk; subclavian artery transposition; thoracic endovascular aortic repair

Mesh:

Year:  2017        PMID: 29338866     DOI: 10.1016/j.jtcvs.2017.11.088

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  2 in total

1.  Aortic balloon occlusion technique in total arch replacement with frozen elephant trunk after thoracic endovascular aortic repair.

Authors:  Yaojun Dun; Yi Shi; Hongwei Guo; Yanxiang Liu; Bowen Zhang; Xiaogang Sun
Journal:  J Thorac Dis       Date:  2020-05       Impact factor: 2.895

2.  The short-term safety and effectiveness of a new distal perforating stent graft in Type B aortic dissection: a retrospective study.

Authors:  Xunqiang Liu; Wenkai Ji; Min Tian; Huanjun Chen; Cuihong Li; Liqiong Zhang; Ying Yang; Jifeng Wang; Min Ji; Chunxin Yang; Enshuai Zhu; Lei Cong; Xili Zhang; Xiaona Zhou; Hailong Liu; Jiaping Wang; Jing Tan; Jinhui Zhang
Journal:  BMC Cardiovasc Disord       Date:  2021-09-21       Impact factor: 2.298

  2 in total