Validation of radiostereometric analysis in six degrees of freedom for use with reverse total shoulder arthroplasty.

Madeleine L Van de Kleut1, Xunhua Yuan2, George S Athwal3, Matthew G Teeter4.   

Abstract

A phantom study was conducted to determine bias in motion and bias at zero motion of radiostereometric analysis (RSA) for evaluating implant relative displacement in reverse total shoulder arthroplasty (RTSA). A Sawbones shoulder phantom was fitted with a RTSA implant set and 13 tantalum markers. The model was fixed to a manual micrometer, providing controlled movements though fifteen known increments in translation and twelve increments in rotation (0.02-5.00 mm and 0.1-6.0°), along each translation and rotation axis. Movement between the glenoid and humerus was assessed using beads vs. beads (B/B), model vs. beads (M/B), and model vs. model (M/M) measurement methods in a model-based RSA environment. Bias in motion and bias at zero motion were defined as the difference between measured and accepted reference values, and the difference between double examinations with a theoretical displacement of zero, respectively. Bias in motion ranged from 0.054 ± 0.010 to 0.129 ± 0.014 mm and 0.076 ± 0.025 to 0.126 ± 0.025° (B/B), 0.023 ± 0.009 to 0.126 ± 0.016 mm and 0.111 ± 0.033 to 0.794 ± 0.251° (M/B), and 0.029 ± 0.010 to 0.135 ± 0.030 mm and 0.243 ± 0.088 to 0.384 ± 0.153° (M/M). Bias at zero motion ranged from 0.120 to 0.156 mm and 0.075 to 0.206° (B/B), 0.074 to 0.149 mm and 0.067 to 1.953° (M/B), and 0.069 to 0.259 mm and 0.284 to 1.273° (M/M). This is the first RSA for RTSA study, with results comparable to those validating the use of RSA for hip and knee arthroplasties (accepted as 0.05-0.50 mm and 0.15-1.15°), justifying the potential use of RSA as a tool for measuring implant displacement in the shoulder.
Copyright © 2017 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Implant relative displacement; Radiostereometric analysis; Reverse total shoulder arthroplasty; Shoulder replacement; Validation

Mesh:

Year:  2017        PMID: 29336821     DOI: 10.1016/j.jbiomech.2017.12.027

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  1 in total

1.  Stable glenoid component of reverse total shoulder arthroplasty at 2 years as measured with model-based radiostereometric analysis (RSA).

Authors:  Alexander Nilsskog Fraser; Berte Bøe; Tore Fjalestad; Jan Erik Madsen; Stephan M Röhrl
Journal:  Acta Orthop       Date:  2021-07-01       Impact factor: 3.717

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.