INTRODUCTION: The LVIS Blue is an FDA-approved stent with 28% metallic coverage that is indicated for use in conjunction with coil embolization for the treatment of intracranial aneurysms. Given a porosity similar to approved flow diverters and higher than currently available intracranial stents, we sought to evaluate the effectiveness of this device for the treatment of intracranial aneurysms.
METHODS: We performed an observational single-center study to evaluate initial occlusion and occlusion at 6-month follow-up for patients treated with the LVIS Blue in conjunction with coil embolization at our institution using the modified Raymond-Roy classification (mRRC), where mRRC 1 indicates complete embolization, mRRC 2 persistent opacification of the aneurysm neck, mRRC 3a filling of the aneurysm dome within coil interstices, and mRRC 3b filling of the aneurysm dome.
RESULTS: Sixteen aneurysms were treated with the LVIS Blue device in conjunction with coil embolization with 6-month angiographic follow-up. Aneurysms were treated throughout the intracranial circulation: five proximal internal carotid artery (ICA) (ophthalmic or communicating segments), two superior cerebellar artery, two ICA terminus, two anterior communicating artery, two distal middle cerebral artery, one posterior inferior cerebellar artery, and two basilar tip aneurysms. Post-procedurally, there was one mRRC 1 closure, five mRRC 2 closures, and 10 mRRC 3a or 3b occlusion. At follow-up, all the mRRC 1 and mRRC 3a closures, 85% of the mRRC 3b closures and 75% of the mRRC 2 closures were stable or improved to an mRRC 1 or 2 at follow-up.
CONCLUSIONS: The LVIS Blue represents a safe option as a coil adjunct for endovascular embolization within both the proximal and distal anterior and posterior circulation. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2018. All rights reserved. No commercial use is permitted unless otherwise expressly granted.

Entities:  

Keywords:  Aneurysm; Flow Diverter; Stent

Mesh:

Year:  2018        PMID: 29331948     DOI: 10.1136/neurintsurg-2017-013608

Source DB:  PubMed          Journal:  J Neurointerv Surg        ISSN: 1759-8478            Impact factor:   5.836


  3 in total

1.  High-Definition Zoom Mode: A High Resolution X-ray Microscope for Neurointerventional Treatment Procedures.

Authors:  Swetadri Vasan Setlur Nagesh; Kunal Vakharia; Muhammad Waqas; Vernard S Fennell; Gursant S Atwal; Hussain Shallwani; Daniel R Bednarek; Jason M Davies; Kenneth V Snyder; Maxim Mokin; Stephen Rudin; Elad I Levy; Adnan H Siddiqui
Journal:  J Neuroimaging       Date:  2019-07-24       Impact factor: 2.486

2.  Stent-Assisted Coiling Using Leo+ Baby Stent : Immediate and Mid-Term Results.

Authors:  Hannes Luecking; Tobias Struffert; Philipp Goelitz; Tobias Engelhorn; Sebastian Brandner; Joji B Kuramatsu; Stefan Lang; Manuel Schmidt; Arnd Doerfler
Journal:  Clin Neuroradiol       Date:  2020-05-08       Impact factor: 3.649

3.  Outcomes following aneurysmal coil embolization with intentionally shortened low-profile visible intraluminal support stent deployment.

Authors:  Kenji Yatomi; Yumiko Mitome-Mishima; Takashi Fujii; Kohsuke Teranishi; Hidenori Oishi; Akihide Kondo
Journal:  Neuroradiol J       Date:  2021-06-29
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.