A tertiary centre experience of thoracic CT and cardiac MRI scanning in the presence of a reveal LINQ insertable cardiac monitoring system: a case series review of artefact, patient safety and data preservation.

Benedict M Wiles1, Caroline A Illingworth2, Michael Lg Couzins2, Paul R Roberts1, Stephen P Harden2.   

Abstract

OBJECTIVE: At our tertiary cardiothoracic centre, cardiac MRI and thoracic CT scans are performed in patients with implanted LINQ devices. The degree of foreign body artefact associated with the LINQ device, and its clinical importance, has not previously been assessed. A case series review was therefore performed with a simultaneous review of patient safety and data loss events, secondary to the MRI environment.
METHODS: A local database search identified LINQ device patients who underwent thoracic CT or cardiac MRI scans between March 2014 and December 2016. Images were reviewed by two radiologists, recording the presence of subcutaneous and intrathoracic artefact, and its clinical significance. Furthermore a specialist in cardiac rhythm management reviewed all LINQ data downloads undertaken before and after MRI scanning, and a search of the trust incident reporting system was performed.
RESULTS: Minor subcutaneous artefact was present on all scans. Intrathoracic artefact was observed in 25.6% of thoracic CT scans and 33.3% of cardiac MRIs; however no clinically significant artefact was observed. Device downloads were only performed by 53.8% of patients prior to their MRI scan and 56.5% after their MRI scan. No adverse patient safety or data loss events were noted.
CONCLUSION: The LINQ device does not produce clinically significant artefact, even when artefact extends into the intrathoracic space, which occurs in a third of MRIs and a quarter of CTs. MRI scanning of the LINQ device is safe with no evidence of inappropriate data loss. Advances in knowledge: This is the first published case series of CT and MRI scanning in LINQ patients and the first performed quantification of artefact related to the LINQ device.

Entities:  

Mesh:

Year:  2018        PMID: 29322834      PMCID: PMC6223274          DOI: 10.1259/bjr.20170615

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  6 in total

Review 1.  Indications for the use of diagnostic implantable and external ECG loop recorders.

Authors:  Michele Brignole; Panos Vardas; Ellen Hoffman; Heikki Huikuri; Angel Moya; Renato Ricci; Neil Sulke; Wouter Wieling; Angelo Auricchio; Gregory Y H Lip; Jesus Almendral; Paulus Kirchhof; Etienne Aliot; Maurizio Gasparini; Frieder Braunschweig; Gregory Y H Lip; Jesus Almendral; Paulus Kirchhof; Gian Luca Botto
Journal:  Europace       Date:  2009-05       Impact factor: 5.214

2.  Reveal LINQ Versus Reveal XT Implantable Loop Recorders: Intra- and Post-Procedural Comparison.

Authors:  Hoang H Nguyen; Ian H Law; Michael W Rudokas; Jennifer Lampe; Tammy M Bowman; George F Van Hare; Jennifer N Avari Silva
Journal:  J Pediatr       Date:  2017-05-22       Impact factor: 4.406

3.  Complications after implantation of a new-generation insertable cardiac monitor: Results from the LOOP study.

Authors:  Søren Zöga Diederichsen; Ketil Jørgen Haugan; Søren Højberg; Anders Gaarsdal Holst; Lars Køber; Kenneth Bruun Pedersen; Claus Graff; Derk Krieger; Axel Brandes; Jesper Hastrup Svendsen
Journal:  Int J Cardiol       Date:  2017-04-01       Impact factor: 4.164

4.  Cryptogenic stroke and underlying atrial fibrillation.

Authors:  Tommaso Sanna; Hans-Christoph Diener; Rod S Passman; Vincenzo Di Lazzaro; Richard A Bernstein; Carlos A Morillo; Marilyn Mollman Rymer; Vincent Thijs; Tyson Rogers; Frank Beckers; Kate Lindborg; Johannes Brachmann
Journal:  N Engl J Med       Date:  2014-06-26       Impact factor: 91.245

5.  Safety Profile of a Miniaturized Insertable Cardiac Monitor: Results from Two Prospective Trials.

Authors:  Suneet Mittal; Prashanthan Sanders; Evgeny Pokushalov; Lukas Dekker; Dean Kereiakes; Edward J Schloss; Erika Pouliot; Noreli Franco; Yan Zhong; Marco DI Bacco; Helmut Pürerfellner
Journal:  Pacing Clin Electrophysiol       Date:  2015-10-20       Impact factor: 1.976

6.  Cardiovascular Magnetic Resonance Imaging in Patients with an Implantable Loop Recorder.

Authors:  Florian Blaschke; Philipp Lacour; Thula Walter; Alexander Wutzler; Martin Huemer; Abdul Parwani; Philipp Attanasio; Leif-Hendrik Boldt; Marcus Markowski; Timm Denecke; Wilhelm Haverkamp
Journal:  Ann Noninvasive Electrocardiol       Date:  2015-12-24       Impact factor: 1.468

  6 in total
  1 in total

1.  Placement of Reveal LINQ Device in the Left Anterior Axillary Position.

Authors:  Heather Anderson; Joseph Dearani; M Yasir Qureshi; Kimberly Holst; Patrick O'Leary; Bryan Cannon; Philip Wackel
Journal:  Pediatr Cardiol       Date:  2019-11-19       Impact factor: 1.655

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.