Literature DB >> 29319661

Observed Length Increases of Magnetically Controlled Growing Rods are Lower Than Programmed.

Sarah E Gilday1, Mark S Schwartz2, Donita I Bylski-Austrow1, David L Glos1, Lindsay Schultz1, Sara O'Hara1, Viral V Jain1, Peter F Sturm1.   

Abstract

BACKGROUND: Magnetically controlled growing rods (MCGRs) are increasingly used in the treatment of early onset scoliosis (EOS). Few studies have reported whether desired lengthening can reliably be achieved, or if prior spine instrumentation and large tissue depths affect lengthening. In this clinical study of EOS patients, it was hypothesized that increases in rod length would equal programmed increases, patients with prior spine instrumentation would lengthen less than patients without prior surgery, and larger tissue depths would decrease lengthening success.
METHODS: A retrospective chart review was conducted on EOS patients with single and dual MCGRs placed between April 2014 to September 2015 and distracted at a single institution. Rod distraction was measured at each visit using ultrasound. Differences between programmed and actual distraction for each patient, and between groups with and without prior spine instrumentation, were determined by 2-tailed t tests. Regression and correlation were used to determine the relationship between tissue depth and length increases.
RESULTS: Thirty-one patients were included, 18 males, 13 females, age 8.1 (±2.5) years, with major curves measuring 60 (±14.6) degrees at time of MCGR insertion. In the 12 patients with prior instrumentation, time from initial growing rod placement to MCGR insertion was 23.1 (±10.6) months. The number of surgical procedures before MCGR insertion was 2.8 (±2.0). Total length increase relative to the programmed distraction was 86% (±21) (P<0.001). Length increases for patients with and without prior surgery were 87% (±23) and 86% (±19), respectively (P>0.9). Total lengthening was inversely proportional to tissue depth (r=0.38, P<0.01); the decrease in lengthening achieved was 2.1%/mm of tissue depth.
CONCLUSIONS: Increases in rod length were 14% lower than the programmed distraction. Prior instrumentation did not impact the amount of rod distraction. Greater distance between the rod and the skin surface negatively affected the magnitude of distraction.

Entities:  

Mesh:

Year:  2018        PMID: 29319661     DOI: 10.1097/BPO.0000000000001119

Source DB:  PubMed          Journal:  J Pediatr Orthop        ISSN: 0271-6798            Impact factor:   2.324


  3 in total

1.  The first magnetically controlled growing rod (MCGR) in the world - lessons learned and how the identified complications helped to develop the implant in the past decade: case report.

Authors:  Jason Pui Yin Cheung; Kam Yim Sze; Kenneth Man Chee Cheung; Teng Zhang
Journal:  BMC Musculoskelet Disord       Date:  2021-04-01       Impact factor: 2.362

2.  Premature spinal fusion after insertion of magnetically controlled growing rods for treatment of early-onset scoliosis: illustrative case.

Authors:  Anna H Green; Andrzej Brzezinski; Terrence Ishmael; Stephen Adolfsen; J Andrew Bowe
Journal:  J Neurosurg Case Lessons       Date:  2021-10-25

3.  "Law of Temporary Diminishing Distraction Gains": The Phenomenon of Temporary Diminished Distraction Lengths With Magnetically Controlled Growing Rods That Is Reverted With Rod Exchange.

Authors:  Jason Pui Yin Cheung; Cora Bow; Kenneth M C Cheung
Journal:  Global Spine J       Date:  2020-08-17