Literature DB >> 29318353

The Barrel Vascular Reconstruction Device : A Retrospective, Observational Multicentric Study.

C Kabbasch1, A Mpotsaris2,3, V Maus2, J C Altenbernd4, C Loehr4.   

Abstract

BACKGROUND AND
PURPOSE: The Barrel device is an electrolytically detachable laser cut, closed-cell microstent that is used for neck reconstruction in wide-necked bifurcation and branching aneurysms to support coiling. The key feature is a barrel section that herniates over the aneurysmal ostium. The objective was to evaluate the safety, feasibility and the immediate and mid-term occlusion results of this new device.
MATERIAL AND METHODS: The databases of two tertiary care centers were screened for all Barrel-based stent-assisted intracranial coil embolization of wide-necked aneurysms between June 2015 and September 2016. Case files and imaging data were retrospectively analyzed for angiographic and clinical outcome parameters, including immediate and mid-term modified Raymond-Roy aneurysm occlusion classification (RROC) rates and procedural complications.
RESULTS: A total of 21 patients comprising 21 intracranial aneurysms (20 unruptured, 1 ruptured) were treated with the Barrel device and additional coiling of the aneurysm sac. All aneurysms were wide-necked, saccular bifurcation aneurysms defined by a dome/neck ratio ≤2. Immediate complete occlusion (RROC1) was observed in 19/21 (90%). An intra-interventional in-stent thrombus formation in two cases (10%) was medically resolved without neurological sequelae. A single case of symptomatic in-stent stenosis (5%) was cleared by balloon angioplasty. Follow-up (FU) was available in 20/21 cases (95%) after a median of 282 days (range: 17-591 days). At follow-up 19/20 aneurysms (95%) were completely occluded (RROC1).
CONCLUSION: The Barrel device showed a satisfactory safety profile and a promising rate of immediate and mid-term complete aneurysm occlusion for stent-assisted coil embolization of wide-necked intracranial bifurcation aneurysms, warranting further investigation of the device.

Entities:  

Keywords:  Bifurcation stent; Coiling; Device; Endovascular; Wide-necked aneurysm

Year:  2018        PMID: 29318353     DOI: 10.1007/s00062-017-0660-2

Source DB:  PubMed          Journal:  Clin Neuroradiol        ISSN: 1869-1439            Impact factor:   3.649


  4 in total

1.  Analysis of branch artery orifice angulation: Feasibility of the shelf technique for the treatment of wide-neck bifurcation aneurysms.

Authors:  Mehmet Onay; Ali Burak Binboga; Cetin Murat Altay
Journal:  Interv Neuroradiol       Date:  2020-11-22       Impact factor: 1.764

Review 2.  Treatment of unruptured middle cerebral artery aneurysms: Systematic review in an attempt to perform a network meta-analysis.

Authors:  Ignacio Arrese; Sergio García-García; Santiago Cepeda; Rosario Sarabia
Journal:  Front Surg       Date:  2022-09-28

Review 3.  Advances in endovascular aneurysm management: coiling and adjunctive devices.

Authors:  Jessica K Campos; Brian V Lien; Alice S Wang; Li-Mei Lin
Journal:  Stroke Vasc Neurol       Date:  2020-03-15

4.  "Shelf" Technique Using a Novel Braided Self-Expandable Stent for the Treatment of Wide-Necked Bifurcation Aneurysms.

Authors:  Volker Maus; Werner Weber; Sebastian Fischer
Journal:  Clin Neuroradiol       Date:  2021-07-20       Impact factor: 3.649

  4 in total