Toshihiko Matsuo1, Tetsuya Uchida2, Koichiro Yamashita2, Shigiko Takei3, Daisuke Ido3, Mamoru Tanaka3, Masao Oguchi3, Toshinori Furukawa4. 1. Ophthalmology, Okayama University Medical School and Graduate School of Medicine, Dentistry, and Pharmaceutical Sciences, Okayama-shi, Okayama 700-8558, Japan. 2. Polymer Materials Science, Okayama University Faculty of Engineering and Graduate School of Natural Science and Technology, Okayama-shi, Okayama 700-8530, Japan. 3. Ina Research, Inc., Ina-shi, Nagano 399-4501, Japan. 4. Kurashiki University of Science and the Arts, Kurashiki-shi, Okayama 712-8505, Japan.
Abstract
Okayama University-type retinal prosthesis (OURePTM) is a photoelectric dye-coupled polyethylene film which generates electric potential in response to light and stimulates nearby neurons. This study aims to test surgical feasibility for subretinal film implantation and to examine functional durability of films in subretinal space. Dye-coupled films were implanted subretinally by vitrectomy in the right eye of normal white rabbits: 8 rabbits for 1 month and 8 rabbits for 6 months. The implanted films were removed by vitrectomy in 4 of these 8 rabbits in 1-month or 6-month implantation group. The films were also implanted in 4 rhodopsin-transgenic retinal dystrophic rabbits. Visual evoked potential was measured before film implantation as well as 1 or 6 months after film implantation, or 1 month after film removal. The films were successfully implanted in subretinal space of retinal detachment induced by subretinal fluid injection with a 38G polyimide tip. The retina was reattached by fluid-air exchange in vitreous cavity, retinal laser coagulation, and silicone oil injection. The ratios of P2 amplitudes of visual evoked potential in the implanted right eye over control left eye did not show significant changes between pre-implantation and post-implantation or post-removal (paired t-test). In Kelvin probe measurements, 4 pieces each of removed films which were implanted for 1 or 6 months showed proportional increase of surface electric potential in response to increasing light intensity. The film implantation was safe and implanted films were capable of responding to light.
Okayama University-type retinal prosthesis (OURePTM) is a photoelectric dye-coupled polyethylene film which generates electric potential in response to light and stimulates nearby neurons. This study aims to test surgical feasibility for subretinal film implantation and to examine functional durability of films in subretinal space. Dye-coupled films were implanted subretinally by vitrectomy in the right eye of normal white rabbits: 8 rabbits for 1 month and 8 rabbits for 6 months. The implanted films were removed by vitrectomy in 4 of these 8 rabbits in 1-month or 6-month implantation group. The films were also implanted in 4 rhodopsin-transgenic retinal dystrophic rabbits. Visual evoked potential was measured before film implantation as well as 1 or 6 months after film implantation, or 1 month after film removal. The films were successfully implanted in subretinal space of retinal detachment induced by subretinal fluid injection with a 38G polyimide tip. The retina was reattached by fluid-air exchange in vitreous cavity, retinal laser coagulation, and silicone oil injection. The ratios of P2 amplitudes of visual evoked potential in the implanted right eye over control left eye did not show significant changes between pre-implantation and post-implantation or post-removal (paired t-test). In Kelvin probe measurements, 4 pieces each of removed films which were implanted for 1 or 6 months showed proportional increase of surface electric potential in response to increasing light intensity. The film implantation was safe and implanted films were capable of responding to light.
Blind patients with hereditary retinal diseases, such as retinitis pigmentosa, have dead
photoreceptor cells, but the other retinal neurons, which send axons to the brain, remain
alive [11]. The basic concept of retinal prostheses is
to stimulate surviving retinal neurons such as ganglion cells and bipolar cells with
artificial devices and to exploit the function of these living neurons, and finally to send
messages to the brain, following artificial stimulation in response to light [14, 19].Okayama University-type retinal prosthesis (OURePTM) is a novel type of retinal
prostheses, so called photoelectric dye-coupled thin film retinal prosthesis [1,2,3, 12, 15, 20,21,22,23]. Stable photoelectric dye molecules with absorption
spectrum of visible light [10, 18] were chemically coupled to polyethylene film surface. The dye-coupled
film generates electric potential in response to light and stimulates nearby neuronal cells to
induce action potential. The dye-coupled film, implanted in subretinal space, serves as a
light-receiver and an electric potential-generator, and thus, replaces the function of dead
photoreceptor cells in retinal dystrophy [14, 19] to send signals to the brain through living retinal
bipolar cells, ganglion cells and their axons as optic nerve fibers [11].The main stream of retinal prosthesis utilizes a multielectrode array [7]. Camera-captured image is disintegrated to 60 pixels, and the electric
current, corresponding to grayscale tone in each pixel, is outputted from 60 electrodes to
stimulate retinal living neurons in Argus II Retinal Prosthesis System (Second Sight Medical
Products, Inc., Sylmar, CA, U.S.A.). Sophisticated techniques are required to implant
surgically the multielectrode arrays. In contrast, the dye-coupled film is a soft and thin
sheet which would be rolled up and inserted in subretinal space by routine vitrectomy [16, 17].In our previous study, the dye-coupled films were implanted subretinally in canine eyes and
were also removed by vitrectomy in order to examine the feasibility of surgical techniques
[17]. In this study, the dye-coupled films were
implanted in normal rabbits’ eyes to be monitored with visual evoked potential. The films were
also implanted in retinal dystrophic eyes of rhodopsin-transgenic rabbits to test whether
surgical techniques would work in the condition of dystrophic thin retina. In addition,
spectrophotometric absorbance and light-evoked surface electric potential was examined on the
dye-coupled films which were implanted for 1 or 6 months in normal eyes of rabbits and then
were removed by vitrectomy.
MATERIALS AND METHODS
Preparation of dye-coupled polyethylene film
Thin films were made from polyethylene powder and exposed to fuming nitric acid to
introduce carboxyl moieties on the film surface. Photoelectric dye molecules,
2-[2-[4-(dibutylamino) phenyl]ethenyl]-3-carboxymethylbenzothiazolium bromide (NK-5962,
Hayashibara, Inc., Okayama, Japan), were coupled to carboxyl moieties of the polyethylene
film surface via ethylenediamine, as described previously [3, 21, 23]. The fuming nitric acid-treated only polyethylene film and the photoelectric
dye-coupled polyethylene film were designated as the plain film and the dye-coupled film,
respectively. Films were manufactured in quality management system at a clean-room
facility in Okayama University Incubator.
Animals
Normal male white rabbits (Kbl: NZW) and transgenic male white rabbits with rhodopsinP347L mutation (Kbl:NZW, specific pathogen-free, Kitayama Labes Co., Ina, Japan) [5, 6, 8, 9] were used in
this study. A pilot study used 3 normal male white rabbits at the age of 31–32 weeks and 2
rhodopsin transgenic rabbits at the age of 13 weeks to test surgical feasibility of
dye-coupled film implantation as well as recording for electroretinography and visual
evoked potential. In pivotal studies, dye-coupled films were implanted for 1 month in 8
normal rabbits and for 6 months in 8 normal rabbits at the age of 14–16 weeks or 28–29
weeks. Of these, dye-coupled films were removed in 4 rabbits with the 1-month implantation
and in 4 rabbits with the 6-month implantation. In addition, dye-coupled films were
implanted for 1 month in 4 rhodopsin-transgenic rabbits at the age of 14 weeks. This study
was approved by the Animal Care and Use Committee in Okayama University and also by the
Committee at Ina Research, Inc., based on the Animal Welfare and Management Act in
Japan.All surgeries were done by T. M. only in the right eye. The dye-coupled films, removed by
vitrectomy, were analyzed for spectrophotometric absorbance and light-evoked surface
electric potential as described later. In the initial plan, electroretinography was
scheduled to be recorded immediately after removal of silicone oil in vitreous cavity
since silicone oil was a non-conducting material to inhibit electric recording with a
contact lens electrode placed on the corneal surface. In 1-month implantation study,
removal of silicone oil resulted in immediate severe inflammation with massive fibrin
deposition in the anterior chamber, and electroretinography was not recorded accurately.
Therefore, in 6-month implantation study, removal of silicone oil was not done and
electroretinography was not recorded.All animals were sacrificed with bleeding after overdose of intravenous thiopental
(Ravonal, Mitsubishi Tanabe Pharma, Osaka, Japan), and the eyes were enucleated. After the
cornea and iris were removed by circumferential incision of the eye ball, the posterior
segment was cut meridionally to view the entire retina. The posterior segment was then
fixed with phosphate-buffered 1% formaldehyde and 2.5% glutaraldehyde, stored in 10%
neutral-pH formalin, and embedded in paraffin. Paraffin sections were cut and stained with
hematoxylin and eosin for pathological examinations.
Surgical procedures
Rabbits were anesthetized with a mixture (1.5 ml/kg body weight) of
intramuscular ketamine (50 mg/kg, Ketamine 5%, Fujita Pharm, Tokyo, Japan) and xylasine
(10 mg/kg, Celactal 2%, Bayer Animal Health, Tokyo, Japan). Mydriasis in the right eye was
induced by 0.5% tropicamide and 0.5% phenylephrine eye drops (Mydrin-P, Santen
Pharmaceutical, Osaka, Japan) on the day of surgery. Subcutaneous injection of meloxicam
(0.2 mg/0.04 ml/kg of body weight, Metacam 0.5%, Boehringer Ingelheim,
Ingelheim am Rhein, Germany) was given once daily for three days after the surgery as a
non-steroidal anti-inflammatory drug. Subcutaneous enrofloxacin (5 mg/kg body weight,
Baytril 2.5%, Bayer Animal Health) was given once daily for 5 days. Postoperative
instillation of 0.5% levofloxacin (Cravit, Santen Pharmaceutical) and 0.1% betamethasone
(Rinderon, Shionogi & Co., Osaka, Japan) twice daily as well as 1% atropine (Nitten,
Nagoya, Japan) once daily was continued for postoperative one month.After disinfection with 10% povidone iodine (Negmin Solution, Pfizer Japan, Tokyo) on the
haired skin around the eye and then with 40-time saline-diluted povidone iodine on the
ocular surface, the rabbit’s head was positioned on the left side down, and covered with a
surgical drape. Topical anesthesia was further obtained with 4% lidocaine (Xylocaine
Ophthalmic Solution, AstraZeneka, London, U.K.). The surgery was done under a surgical
microscope (OPMI VISU150, Carl Zeiss Meditec, Tokyo, Japan) with a surgical machine
(Constellation Vision System, Alcon Laboratories, Inc., Fort Worth, TX, U.S.A.). Anterior
capsulectomy (Fig. 1a) was done with a 25-gauge vitreous cutter under irrigation with a 25-gauge infusion
cannula through two side ports which were made at the corneal limbus with a 20-gauge knife
(V-Lance Knife, Alcon), as done in congenital cataract surgery in human eyes [13]. Phacoemulsification and aspiration of the lens in
the capsular bag (Fig. 1b) was done through a
2.4 mm-wide corneal incision made on the superior side with a disposable knife (Safety
Knife, Kai Medical, Seki, Japan). The corneal incision was sutured with 8–0 Vicryl
(polyglactin 910) suture (Ethicon, Johnson & Johnson, New Brunswick, NJ, U.S.A.).
Three 25-gauge trocars were inserted into the vitreous through the conjunctiva and sclera
2.5 mm from the limbus on the superior to temporal side within 120 degrees of meridian
(Fig. 1c). The presence of a large
vascularized nictitating membrane on the nasal side of the conjunctiva limited the
surgical area used for placing trocars.
Fig. 1.
Surgical procedures to implant retinal prosthesis, OURePTM, in right eye
of rhodopsin-transgenic white rabbit. a) Lens anterior capsule is cut with 25G
vitreous cutter under irrigation with 25G infusion cannula in anterior chamber. b)
Lens nucleus and cortex is aspirated with phaco-tip from corneal incision. c) Three
25G trocars are inserted over conjunctiva through sclera into vitreous at 2.5 mm
from corneal limbus: a middle trocar is connected with infusion cannula, and the
other two trocars are used for vitreous cutter and light guide. Posterior capsule is
cut with vitreous cutter. d) After vitreous gel has been cut, subretinal fluid
infusion is started with 38G tip. e) Bleb retinal detachment (arrow) is made by 38G
tip infusion of BSS-Plus solution. f) Bleb (arrow) is enlarged with further
infusion. g) Retinal tear is made by retinal coagulation with 25G bipolar diathermy.
h) Scleral incision is made with 22.5° knife after conjunctival incision. i)
Rolled-up dye-coupled film (arrow) is inserted through scleral incision with 20G
subretinal forceps. j) Rolled-up film is inserted into vitreous with 20G subretinal
forceps. k) Rolled-up film is inserted into subretinal space through retinal tear
with 20G subretinal forceps. l) Film is now under detachment retina. m) Fluid-air
exchange in vitreous cavity is accomplished with 25G vitreous cutter in aspiration
mode to reattach the retina. n) Laser photocoagulation is applied around retinal
tear. o) Silicone oil is injected in vitreous cavity with 25G tip. Finally, scleral
and conjunctival incision is sutured and trocars are removed (not shown).
Surgical procedures to implant retinal prosthesis, OURePTM, in right eye
of rhodopsin-transgenic white rabbit. a) Lens anterior capsule is cut with 25G
vitreous cutter under irrigation with 25G infusion cannula in anterior chamber. b)
Lens nucleus and cortex is aspirated with phaco-tip from corneal incision. c) Three
25G trocars are inserted over conjunctiva through sclera into vitreous at 2.5 mm
from corneal limbus: a middle trocar is connected with infusion cannula, and the
other two trocars are used for vitreous cutter and light guide. Posterior capsule is
cut with vitreous cutter. d) After vitreous gel has been cut, subretinal fluid
infusion is started with 38G tip. e) Bleb retinal detachment (arrow) is made by 38G
tip infusion of BSS-Plus solution. f) Bleb (arrow) is enlarged with further
infusion. g) Retinal tear is made by retinal coagulation with 25G bipolar diathermy.
h) Scleral incision is made with 22.5° knife after conjunctival incision. i)
Rolled-up dye-coupled film (arrow) is inserted through scleral incision with 20G
subretinal forceps. j) Rolled-up film is inserted into vitreous with 20G subretinal
forceps. k) Rolled-up film is inserted into subretinal space through retinal tear
with 20G subretinal forceps. l) Film is now under detachment retina. m) Fluid-air
exchange in vitreous cavity is accomplished with 25G vitreous cutter in aspiration
mode to reattach the retina. n) Laser photocoagulation is applied around retinal
tear. o) Silicone oil is injected in vitreous cavity with 25G tip. Finally, scleral
and conjunctival incision is sutured and trocars are removed (not shown).The wide-field fundus was viewed with a +128-diopter front lens by Resight 500 fundus
viewing system (Carl Zeiss Meditec). Posterior capsulectomy (Fig. 1c) and core vitrectomy was done under irrigation with a
25-gauge cannula placed at the middle trocar on the superior side. Retinal detachment
(Fig. 1d, 1e and 1f) was then made by infusing
irrigation solution (BSS-Plus Intraocular Irrigating Solution, Alcon) into the subretinal
space with a 38-gauge polyimide tip (PolyTip Cannula 25G/38G, MedOne Surgical, Inc.,
Sarasota, FL, U.S.A.) attached to a 10-ml syringe for the viscous fluid
control (VFC) system at the setting of low intraocular pressure. A retinotomy (Fig. 1g) was made by 25-gauge diathermy (Grieshaber
Diathermy Probe DSP 25Ga, Alcon) at the edge of retinal detachment. After conjunctival
incision, a 3 mm-wide scleral incision (Fig. 1h)
was placed with a microsurgery knife (Straight/Stab 22.5°, Kai Medical) 2 mm posteriorly
in parallel with the corneal limbus, and wound hemostasis was done with a wet-field
hemostatic eraser bipolar instrument (Beaver-Visitec International, Inc., Waltham, MA,
U.S.A.). A rolled-up sheet of the dye-coupled film in 5 × 5 mm square size (Fig. 1i) was grasped with a 20-gauge subretinal
forceps (Synergetics 39.21S, Bausch+Lomb Retina, St. Louis, MO, U.S.A.), and inserted into
the vitreous (Fig. 1j) and then under the
detached retina through a retinotomy (Fig. 1k
and 1l). The scleral incision for film insertion was sutured with 8-0 Vicryl. The
subretinal fluid was aspirated with a vitreous cutter, and then fluid in vitreous cavity
was exchanged with air to reattach the retina (Fig.
1m). Laser photocoagulation (Fig. 1n)
was applied around the retinal tear caused by retinotomy, and silicone oil
(polydimethylsiloxane, Silikon 1000, Alcon) was infused into vitreous cavity (Fig. 1o) by the VFC syringe. Trocars were removed,
and the conjunctiva was sutured with 8-0 Vicryl.At surgery to remove dye-coupled films implanted in the subretinal space, three 25-gauge
trocars were placed and retinal detachment was induced by infusing irrigation fluid into
the subretinal space by a 38-gauge tip. A retinotomy was made by diathermy and a
subretinal dye-coupled film was grasped with a 25-gauge forceps (Grieshaber Revolution DSP
25Ga ILM Forceps, Alcon) and brought to the vitreous cavity. The film was brought out of
the eye ball with a 25-gauge forceps through a newly made 3-mm-wide scleral incision 2 mm
posterior to the corneal limbus. The retina was reattached by fluid-air exchange, laser
photocoagulation, and silicone oil infusion. The films were immersed in distilled water
for functional analyses. Microscopic surgical view was recorded with a digital processor
miniature 3CCD color camera (THD-311, Ikegami Tsushinki Co., Tokyo, Japan).
Electroretinography and visual evoked potential
Rabbits were sedated with intramuscular injection of the mixture of medetomidine (0.1
mg/0.1 ml to 0.2 mg/0.2 ml/kg of body weight, Domitor,
Nippon Zenyaku Kogyo Co., Koriyama, Japan) and midazolam (0.75 mg/0.15 ml
to 1.5 mg/0.3 ml/kg of body weight, Dormicum, Astellas Pharma Inc.,
Tokyo, Japan). Mydriasis was obtained with 0.5% tropicamide and 0.5% phenylephrine eye
drops (Mydrin-P), and dark adaptation was achieved for at least 40 min. After ocular
surface anesthesia with 0.4% bupivacaine eye drops (Benoxil, Santen Pharmaceutical), a
contact lens electrode with white light-emitting diode (LED) was placed on the corneal
surface with hydroxyethylcellulose gel (Scopisol, Senju Pharmaceutical, Osaka, Japan), a
reference electrode was put at cranial front, and a ground clip was placed along the tail.
Maximal response to standard flash (5,623 cd/m2, 3 msec) was recorded (White
LED Visual Stimulator LS-W and White LED electrode, Mayo Corp., Aichi, Japan, Dual BioAmp
ML135 and A/D Converter ML820 PowerLab 2/25, ADInstruments Japan, Nagoya). After the
measurements, sedation was reversed with intramuscular injection of atipamezole (1 mg/kg
of body weight, Antisedan, Nippon Zenyaku Kogyo Co.).Visual evoked potential was measured with subcutaneous needle electrodes placed on the
parietal center for recording, on the dorsum of nose for reference, and along the right
auricle for ground by LE-4000 (Tomey Co., Nagoya, Japan). After dark adaptation and
mydriasis for at least 15 min, flash light (LE-4000) stimuli in the intensity of 30 cd ×
s/m2 (100,000 cd/m2 × 0.3 msec) at the frequency of 2 Hz were
given to the right eye and then to the left eye. The amplitude and latency of
P2 and the latency of N2 in the 128 summation of recordings were
used for statistical analysis, based on the International Society for Clinical
Electrophysiology of Vision (ISCEV) standards for clinical visual evoked potentials in
humans. P2 amplitude and latency as well as N2 latency were
automatically calculated after the measurement by LE-4000.
Spectrophotometry
Absorbance spectra of dye-coupled films were measured by an ultraviolet and visible light
spectrophotometer with an integrating sphere unit (V-750 and PIV-756, JASCO Corp., Tokyo,
Japan). Plain polyethylene films were used to obtain baseline absorbance. Absorbance was
measured in the wavelength ranging from 300 to 800 nm at 1 nm spectral bandwidth. Maximum
absorbance values around 500 nm were used for comparison between the implanted film and
the non-implanted same lot.
Light-evoked surface electric potential
Light-evoked surface electric potential on dye-coupled films in the dry condition was
measured by the scanning Kelvin Probe system (SKP5050, KP Technology, Ltd., Highlands and
Islands, U.K.) in the surface photovoltage mode. The entire measuring system was placed in
a humidity-controlling box and kept at low humidity. The dye-coupled film was fixated on a
sample device. The capacitance between the probe and the sample was changed by oscillating
the probe. The surface potential of the sample was measured by adjusting the bias to set
electrostatic attractive force at zero. The distance between the probe and the sample was
kept constant by setting the gradient at 200. To confirm the measuring system to be
stable, work function at a single point on the sample was measured repeatedly 100 times
until standard deviation of work function became 10 or less. Only after a waiting time to
obtain the stability, the surface potential was measured at changing light intensity with
a light source (Surface Photovoltage Spectroscopy SPS040, KP Technology). Surface
potential changes in response to increase of light intensity as well as the surface
potential in light intensity of dial setting at 2,500 (300 lux equivalent) was used as the
outcome.
RESULTS
Surgical feasibility
The lens of rabbits at younger ages was soft enough to be aspirated by an ultrasound tip
without phacoemulsification (Fig.
1b). The vitreous cavity in rabbits was smaller in space, compared
with human eyes. Bleb retinal detachment could be made successfully in dystrophic thin
retina as done in the normal retina by puncture with an irrigating 38-gauge tip (Fig. 1d, 1e and 1f) in the lower part of the retina
to avoid the medullary rays. A major complication was dislodgment of 25G trocars which
were placed over the conjunctiva through the sclera because the rabbits’ sclera is thinner
compared with human sclera. A surgical assistant monitored the position of the trocars and
pushed the trocars with a forceps to avoid the dislodgment during the surgery.Electroretinogrpahy was recorded in both eyes of all rabbits before film implantation in
the study. The amplitudes of a-wave and b-wave before surgery in 4 normal rabbits for
1-month implantation were 172.6 ± 44.8 µV of a-wave and 460.6 ± 122.1
µV of b-wave in the right eyes with surgery and 178.7 ± 48.2
µV of a-wave and 488.4 ± 136.0 µV of b-wave in the
left eyes with no intervention, described as mean ± standard deviation (n=4). In contrast,
the amplitudes of a-wave and b-wave were reduced in 4 rhodopsin-transgenic rabbits before
surgery: 65.9 ± 6.5µV of a-wave and 234.0 ± 39.9 µV of
b-wave in the right eyes with surgery and 67.4 ± 11.8 µV of a-wave and
246.0 ± 78.3 µV of b-wave in the left eyes with no intervention, as mean
± standard deviation.P2 amplitude, P2 and N2 latency of visual evoked
potential in the right eyes with dye-coupled film implantation did not change
significantly between pre-implantation and 1-month or 6-month implantation in normal
rabbits (n=4 for each group, paired t-test, Table 1). P2 amplitude, P2 and N2 latency of visual
evoked potential in the left eyes of these rabbits as control also did not show
significant change (n=4 for each group, paired t-test, Table 1). The ratio of P2 amplitude of
the right eye over the left eye did not show significant change in 1-month or 6-month
implantation (n=4 for each group, paired t-test, Table 1).
Table 1.
P2 amplitude, P2 latency and N2 latency of
visual evoked potential in normal rabbits with 1-month or 6-month dye-coupled film
implantation and removal, and in rhodopsin-transgenic rabbits with 1-month
dye-coupled film implantation
P2 amplitude in right eye (surgery)Mean ± SD
(µV)
P2 amplitude in left eye (control)
Mean ± SD (µV)
Right eye/Left eye ratio of P2
amplitude
P2 latency in right eye (surgery)
Mean ± SD (msec)
P2 latency in left eye (control)
Mean ± SD (msec)
N2 latency in right eye (surgery)
Mean ± SD (msec)
N2 latency in left eye (control) Mean ± SD
(msec)
Normal rabbits with 1-month
implantation
Pre-implantation
2.51 ± 0.97 (n=4)
2.18 ± 0.93 (n=4)
1.21 ± 0.29 (n=4)
90.2 ± 13.8 (n=4)
84.5 ± 2.0 (n=4)
75.5 ± 12.8 (n=4)
69.5 ± 3.6 (n=4)
1-month implantation
1.55 ± 1.39 (n=4)
2.18 ± 1.09 (n=4)
1.20 ± 0.80 (n=4)
89.2 ± 10.8 (n=4)
92.5 ± 26.0 (n=4)
79.2 ± 10.0 (n=4)
65.2 ± 1.2 (n=4)
Paired t-test P
value
P=0.0736
P>0.9999
P=0.9853
P=0.8571
P=0.5742
P=0.3622
P=0.1203
Normal rabbits with 1-month
implantation and removal
Pre-implantation
4.67 ± 1.77 (n=4)
4.30 ± 1.46 (n=4)
1.16 ± 0.45 (n=4)
90.0 ± 10.4 (n=4)
86.0 ± 9.8 (n=4)
63.7 ± 1.5 (n=4)
64.7 ± 5.1 (n=4)
1 month after removal
1.55 ± 1.86 (n=3)
3.63 ± 1.72 (n=3)
0.79 ± 0.59 (n=3)
83.3 ± 8.0 (n=3)
97.6 ± 7.0 (n=3)
74.3 ± 6.0 (n=3)
74.0 ± 4.0 (n=3)
Paired t-test P
value
P=0.1282
P=0.3472
P=0.4136
P=0.8171
P=0.1567
P=0.1009
P=0.0128
Normal rabbits with 6-month
implantation
Pre-implantation
2.23 ± 0.20 (n=4)
1.75 ± 0.80 (n=4)
1.41 ± 0.42 (n=4)
82.5 ± 1.9 (n=4)
83.7 ± 2.9 (n=4)
68.0 ± 5.7 (n=4)
68.0 ± 6.1 (n=4)
6-month implantation
1.67 ± 0.88 (n=4)
2.06 ± 0.39 (n=4)
0.79 ± 0.38 (n=4)
100.5 ± 22.4 (n=4)
87.0 ± 11.4 (n=4)
72.5 ± 12.7 (n=4)
69.7 ± 12.5 (n=4)
Paired t-test P
value
P=0.2805
P=0.4908
P=0.2119
P=0.2110
P=0.5222
P=0.5934
P=0.8310
Normal rabbits with 6-month
implantation and removal
Pre-implantation
4.63 ± 2.83 (n=4)
3.26 ± 1.24 (n=4)
1.31 ± 0.63 (n=4)
83.2 ± 4.9 (n=4)
83.2 ± 2.7 (n=4)
65.2 ± 3.8 (n=4)
68.2 ± 3.5 (n=4)
6 month implantation
0.81 ± 0.33 (n=4)
2.63 ± 0.53 (n=4)
0.32 ± 0.16 (n=4)
85.5 ± 4.5 (n=4)
78.5 ± 4.7 (n=4)
74.2 ± 5.8 (n=4)
65.7 ± 4.5 (n=4)
1 month after removal
1.36 ± 0.38 (n=3)
1.95 ± 0.48 (n=3)
0.78 ± 0.48 (n=3)
108.2 ± 30.5 (n=3)
78.5 ± 2.6 (n=3)
76.2 ± 9.2 (n=3)
64.0 ± 2.1 (n=3)
Paired t-test P
value (pre vs 6 month)
P=0.0668
P=0.4486
P=0.0595
P=0.6289
P=0.1169
P=0.1432
P=0.5696
Paired t-test P
value (pre vs removal)
P=0.2523
P=0.2105
P=0.4376
P=0.2189
P=0.0192
P=0.1269
P=0.1828
Rhodopsin-transgenic rabbits with
1-month implantation
Pre-implantation
1.46 ± 0.61 (n=4)
1.86 ± 0.85 (n=4)
0.86 ± 0.31 (n=4)
78.7 ± 2.3 (n=4)
83.0 ± 7.4 (n=4)
69.0 ± 1.8 (n=4)
72.7 ± 9.1 (n=4)
1-month implantation
0.86 ± 0.56 (n=4)
1.67 ± 1.15 (n=4)
0.82 ± 0.63 (n=4)
96.5 ± 18.6 (n=4)
95.7 ± 11.0 (n=4)
73.2 ± 6.3 (n=4)
80.5 ± 9.0 (n=4)
Paired t-test P
value
P=0.0469
P=0.8633
P=0.9358
P=0.1196
P=0.1704
P=0.3000
P=0.1134
SD, standard deviation; µV, microvolt.
SD, standard deviation; µV, microvolt.In 4 normal rabbits with 1-month film implantation and film removal, P2
amplitude, P2 and N2 latency of visual evoked potential in the right
eyes with surgeries did not change significantly between pre-implantation and 1 month
after film removal (n=3 due to death of 1 rabbit, paired t-test, Table 1). P2 amplitude, P2
and N2 latency of visual evoked potential in the left eyes of these rabbits as
control also did not show significant change (n=3, paired t-test, Table 1). In 4 normal rabbits with 6-month film
implantation and film removal, visual evoked potential was measured at 3 time points:
pre-implantation, 6-month implantation, and 1 month after surgical removal of 6-month
implanted films. In statistical analysis by repeat-measure analysis of variance (ANOVA),
P2 amplitude showed significant change in the time course
(P=0.0270), but did not show significant change between the right eyes
with surgeries and the left eyes with no intervention (P=0.2152).
P2 amplitude, P2 and N2 latency in the right eyes with
surgery and in the left eyes with no intervention did not show significant change between
pre-implantation and 6-month implantation or between pre-implantation and 1-month after
film removal (n=4 for 6-month implantation, n=3 for film removal due to loss of 1 rabbit,
paired t-test, Table 1).In rhodopsin-transgenic rabbits with retinal dystrophy, P2 amplitude in the
right eyes with 1-month film implantation show significant decrease compared with
pre-implantation (n=4, P=0.0469, paired t-test, Table 1) while P2 amplitude in the left
eyes with no intervention showed no significant change (n=4, P=0.8633).
However, the ratio of P2 amplitude of the right eye over the left eye did not
show significant change between pre-implantation and 1-month implantation (n=4,
P=0.9358, paired t-test, Table 1). P2 and N2 latency in the right
eyes with surgery and in the left eyes with no intervention did not show significant
change between pre-implantation and 1-month after implantation (Table 1).
Pathology
Gross anatomy of enucleated eyes with 1-month or 6-month film implantation showed no
marked proliferation (Fig. 2a). The enucleated eyes were cut meridionally to take photographs before fixation.
This process resulted in flowing-out of dye-coupled film under the retina, together with
silicone oil in vitreous cavity. Microscopic observation disclosed no infiltration with
inflammatory cells or no proliferation. In the eyes of normal rabbits with 1-month film
implantation, photoreceptor outer segments in the retina were present but shortened in
length (Fig. 2b). In the eyes of
rhodopsin-transgenic rabbits with 1-month film implantation, photoreceptor outer segments
were almost lost (Fig. 2c). In the eyes of
normal rabbits with 6-month film implantation, retinal outer nuclear layer with
photoreceptor outer segments was lost (Fig.
2d).
Fig. 2.
Macroscopic view of unfixed and dissected eye with subretinal square dye-coupled
film (5 × 5 mm, arrows) in 1-month implantation (a). Light microscopic sections of
the retina of the posterior pole near film implantation in normal rabbit with
1-month film implantation (b), rhodopsin-transgenic rabbit with 1-month film
implantation (c), and normal rabbit with 1 month observation after removal of
6-month implanted film (d). The photoreceptors are at bottom of photographs.
Separation between the choroid and sclera is artifact (b). Photoreceptor outer
segments are relatively maintained in normal rabbit (b) while outer segments are
shortened with eosin-stained serous fluid in rhodopsin-transgenic rabbit (c). Loss
of retinal outer nuclear layer is noted 1 month after removal of 6-month implanted
film in normal rabbit (d). Hematoxylin-eosin stain. Scale in ruler is 1 mm in a.
Scale bar=100 µm in b, c and d.
Macroscopic view of unfixed and dissected eye with subretinal square dye-coupled
film (5 × 5 mm, arrows) in 1-month implantation (a). Light microscopic sections of
the retina of the posterior pole near film implantation in normal rabbit with
1-month film implantation (b), rhodopsin-transgenic rabbit with 1-month film
implantation (c), and normal rabbit with 1 month observation after removal of
6-month implanted film (d). The photoreceptors are at bottom of photographs.
Separation between the choroid and sclera is artifact (b). Photoreceptor outer
segments are relatively maintained in normal rabbit (b) while outer segments are
shortened with eosin-stained serous fluid in rhodopsin-transgenic rabbit (c). Loss
of retinal outer nuclear layer is noted 1 month after removal of 6-month implanted
film in normal rabbit (d). Hematoxylin-eosin stain. Scale in ruler is 1 mm in a.
Scale bar=100 µm in b, c and d.
Absorbance and light-evoked surface electric potential
Figures 3 and 4 show spectrophotometric absorbance and light-evoked surface electric potential,
respectively, of dye-coupled films which were implanted for 1 month and removed by
vitrectomy. The mean of maximum absorbance around 500 nm of 4 pieces of 1-month implanted
films was 0.02363, with the range from 0.02257 to 0.02455. The percentage of the mean
absorbance, relative to the non-implanted same lot film (0.04586), was 51.5%. The mean of
surface potential at light intensity of 2500, equivalent to 300 lux, of 4 pieces of
1-month implanted films was 65 mV, with the range from 58 to 69.3 mV. The percentage of
the mean surface potential, relative to the non-implanted same lot film (67.6 mV), was
96.2%.
Fig. 3.
Spectrophotometric absorbance spectra (4 bottom panels) of 4 pieces of dye-coupled
films which have been implanted for 1 month in subretinal space of rabbits’ eyes and
removed by vitrectomy. Top panel shows absorbance spectrum of the non-implanted same
lot. Insets are photographs of films. Values in each panel represent maximum
absorbance around the wavelength of 500 nm.
Fig. 4.
Surface electric potential in response to increasing light intensity on 4 pieces of
dye-coupled films which have been implanted for 1 month in subretinal space of
rabbits’ eyes and removed by vitrectomy. Top panel shows surface electric potential
on the non-implanted same lot. Red dots on each panel represent electric potential
at 2,500 arbitrary unit (AU) of light intensity which corresponds to 300 lux.
Spectrophotometric absorbance spectra (4 bottom panels) of 4 pieces of dye-coupled
films which have been implanted for 1 month in subretinal space of rabbits’ eyes and
removed by vitrectomy. Top panel shows absorbance spectrum of the non-implanted same
lot. Insets are photographs of films. Values in each panel represent maximum
absorbance around the wavelength of 500 nm.Surface electric potential in response to increasing light intensity on 4 pieces of
dye-coupled films which have been implanted for 1 month in subretinal space of
rabbits’ eyes and removed by vitrectomy. Top panel shows surface electric potential
on the non-implanted same lot. Red dots on each panel represent electric potential
at 2,500 arbitrary unit (AU) of light intensity which corresponds to 300 lux.Figures 5 and 6 show spectrophotometric absorbance and light-evoked surface electric potential,
respectively, of dye-coupled films which were implanted for 6 months and removed by
vitrectomy. The mean of maximum absorbance around 500 nm of 4 pieces of 6-month implanted
films was 0.01465, with the range from 0.01253 to 0.01705. The percentage of the mean
absorbance, relative to the non-implanted same lot film (0.09523), was 15.4%. The mean of
surface potential at light intensity of 2,500, equivalent to 300 lux, of 4 pieces of
6-month implanted films was 76 mV, with the range from 43 to 112.5 mV. The percentage of
the mean surface potential, relative to the non-implanted same lot film (129.8 mV), was
59.1%.
Fig. 5.
Spectrophotometric absorbance spectra (4 bottom panels) of 4 pieces of dye-coupled
films which have been implanted for 6 months in subretinal space of rabbits’ eyes
and removed by vitrectomy. Top panel shows absorbance spectrum of the non-implanted
same lot. Insets are photographs of films. Values in each panel represent maximum
absorbance around the wavelength of 500 nm.
Fig. 6.
Surface electric potential in response to increasing light intensity on 4 pieces of
dye-coupled films which have been implanted for 6 months in subretinal space of
rabbits’ eyes and removed by vitrectomy. Top panel shows surface electric potential
on the non-implanted same lot. Red dots on each panel represent electric potential
at 2,500 arbitrary unit (AU) of light intensity which corresponds to 300 lux.
Spectrophotometric absorbance spectra (4 bottom panels) of 4 pieces of dye-coupled
films which have been implanted for 6 months in subretinal space of rabbits’ eyes
and removed by vitrectomy. Top panel shows absorbance spectrum of the non-implanted
same lot. Insets are photographs of films. Values in each panel represent maximum
absorbance around the wavelength of 500 nm.Surface electric potential in response to increasing light intensity on 4 pieces of
dye-coupled films which have been implanted for 6 months in subretinal space of
rabbits’ eyes and removed by vitrectomy. Top panel shows surface electric potential
on the non-implanted same lot. Red dots on each panel represent electric potential
at 2,500 arbitrary unit (AU) of light intensity which corresponds to 300 lux.
DISCUSSION
The goals of this study are two folds: the first to test surgical feasibility of subretinal
film implantation by vitrectomy in rabbits, and the second to test functional durability of
the implanted films. In the previous study [17],
beagle dogs were used for dye-coupled film implantation. The canine study demonstrated
technical feasibility of film implantation and 5-month functional durability of the
implanted films. In the present study, rabbits were used for film implantation since rabbits
have been more frequently used as an animal model for ophthalmic surgeries [4]. In addition, visual evoked potential can be easily
recorded in rabbits with thin bony skull, in contrast with difficulty in the recording in
dogs with thick bony skull.We could successfully implant dye-coupled films in square size of 5 × 5 mm in subretinal
space of rabbits’ eyes by vitrectomy even though the film size was rather larger compared
with smaller size of rabbits’ eyes, relative to human eyes. The dye-coupled films which had
been implanted in subretinal space of the eyes were scheduled to be removed 1 or 6 months
after implantation, and to be submitted to functional analyses. Base on a technological
limitation at that time, the films in this square size were required for fixation on the
sample device to measure light-evoked surface electric potential by Kelvin probe.We also tested to implant dye-coupled films in rhodopsin-transgenic rabbits which developed
retinal dystrophy. Retinal prosthesis is scheduled to be applied in patients with retinitis
pigmentosa who have thin degenerative retina. Therefore, we thought that technical
feasibility of implantation surgery should be tested in an animal model with thin
degenerative retina. We could successfully induce retinal detachment by infusing solution
with a 38-gauge tip under the thin degenerative retina in rhodopsin-transgenic rabbits.In our previous study using dogs [17], light-evoked
surface electric potential could be measured by Kelvin probe only on one piece of
dye-coupled films which were implanted in subretinal space for 5 months and then removed by
vitrectomy. Surface electric potential could not be measured on the other 4 pieces of
dye-coupled films which were implanted either for 3 or for 5 months [17]. The Kelvin probe system has been renovated since then to control
humidity in the environment to measure surface electric potential of dye-coupled films
repeatedly at a constant level. In the present study, light-evoked surface electric
potential could be measured on all pieces of 1-month or 6-month implanted dye-coupled
films.In addition, we measured absorbance of dye-coupled films with a new type of
spectrophotometry with an integrating sphere in the present study. In the previous study
with dogs [17], absorbance of one piece of implanted
films with surface wrinkling could not be measured with a preceding type of
spectrophotometry. With spectrophotometry with an integrating sphere unit, absorbance of all
the implanted films was measured successfully in the present study.It should be noted that surface electric potential on 4 pieces of 1-month implanted
dye-coupled films was basically at the same level as that of the non-implanted same lot. In
contrast, spectrophotometric absorbance of 1-month implanted films was reduced to about the
half of absorbance of the non-implanted same lot. This discrepancy between the absorbance
and surface electric potential would be explained by the presence of non-covalently bound
dye molecules which were attached to the film surface by ionic binding. Non-covalently bound
dye molecules would contribute to absorption but would not contribute to the generation of
surface electric potential.The range of surface electric potential on 4 pieces of 6-month implanted films was large
while surface electric potential generated on the non-implanted same lot was rather high.
Therefore, mean percentage of electric potential of 6-month implanted films was reduced to
60%, relative to the potential of the non-implanted same lot, but actual values of potential
were comparable to values for 1-month implanted films. The absorbance of 6-month implanted
films was further reduced to 15% of the non-implanted same lot, but was not in parallel with
light-evoked surface electric potential. In other words, the discrepancy between the
absorbance and surface electric potential was further widened in the 6-month implanted
films. It should be emphasized that absorbance and surface electric potential did not change
in parallel with each other among 4 pieces of dye-coupled films which were implanted not
only for 6 months but also for 1 month. A method for measuring absorbance of films has to be
further refined in future to assess accurately the dye-coupled films.We initially planned to assess safety of dye-coupled film implantation by
electroretinography and visual evoked potential. Visual evoked potential is recorded from
the skull in response to repeat flashing light to the eyes. The presence of silicone oil in
the vitreous did not influence the recording of visual evoked potential. In contrast,
electroretinography records retinal electric activity in response to light via the vitreous
with a contact lens electrode placed on the corneal surface. The presence of silicone oil, a
non-conductive material, in the vitreous prevents electroretinographic recording. We,
therefore, planned to remove silicone oil just before electroretinographic recording in the
studies of 1-month and 6-month film implantation. However, the procedure of silicone oil
removal by vitrectomy led to so early and so severe intraocular inflammation as massive
fibrin deposition in the anterior chamber, and resulted in no recording of
electroretinography in 1-month implantation study.We, therefore, used visual evoked potential as an indicator to assess safety of dye-coupled
film implantation in rabbits. The amplitude and latency of visual evoked potential did not
show significant change in the right eyes with film implantation and in the left eyes with
no intervention for 1 and 6 months in normal rabbits. In addition, the ratio of amplitudes
of the right eye over the left eye did not show significant change in the time course of 1
and 6 months in normal rabbits. Furthermore, no significant change in the right eye/left eye
ratio of amplitudes was noted at 1 month after surgical removal of 1- or 6-month implanted
films. These facts suggest the safety of film implantation and film removal by vitrectomy in
rabbits. No significant change in the right eye/left eye ratio of amplitudes of visual
evoked potential was noted also in rhodopsin-transgenic rabbits with retinal dystrophy. The
presence of visual evoked potential in rhodopsin-transgenic rabbits suggests that these
rabbits at this stage of the age would not be suitable for a model of retinal dystrophy with
no vision.In conclusion, this study proved surgical feasibility of subretinal implantation and
removal as well as 6-month functional durability of retinal prosthesis, OURePTM,
in rabbits. The filing of a first-in-human clinical trial for OURePTM is now
negotiated at Pharmaceuticals and Medical Devices Agency (PMDA) in Japan.