Literature DB >> 2930986

Late rupture of knitted Dacron double velour arterial prostheses. Report on four cases.

A Trippestad.   

Abstract

Late fibre breakage with longitudinal tears occurred in 4 patients of a series of some 300 implanted knitted Dacron double velour arterial prostheses. The four patients all received their grafts for occlusive aortoiliac disease in the year 1978, and the rupture was found 4 to 6 years later. Three of the failures occurred in axillo-bifemoral bypass grafts, whereas the fourth patient had a bifurcated aortic prosthesis. Dilatation was significant in the latter graft only. Structural weakness might appear during production, during later handling, or possibly as a result of interference from infected material or host factors. The cause of graft failures in the present series could not be ascertained. Whatever type of knitted Dacron used, a 2-3% incidence of graft deterioration might be expected. Therefore regular, life-long follow-up of patients with Dacron arterial prostheses seems warranted.

Entities:  

Mesh:

Substances:

Year:  1985        PMID: 2930986

Source DB:  PubMed          Journal:  Acta Chir Scand        ISSN: 0001-5482


  6 in total

1.  Successful repair of graft failure and abdominal aortic aneurysm following thoraco-abdominal bypass for atypical aortic coarctation.

Authors:  Tsuyoshi Taketani; Tetsuro Morota; Ko Shibata; Noboru Motomura; Yutaka Kotsuka; Shinichi Takamoto
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2004-02

2.  Efficacy of contrast-enhanced ultrasonography in detecting graft rupture sites after abdominal aortic aneurysm repair.

Authors:  Koji Nakanishi; Ryoji Watanabe; Shuji Shimizu; Mikizo Nakai
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-11-11

3.  [PTFE prostheses in infrarenal aneurysms of the abdominal aorta].

Authors:  M M Kaiser; H Wenk; J Link; G Müller; H P Bruch
Journal:  Langenbecks Arch Chir       Date:  1994

4.  Bleeding through the fiber interstices of a knitted dacron graft 12 years after its implantation: report of a case.

Authors:  K Akiyama; A Takazawa; J Hirota; H Yamanishi; T Akazawa; T Maeda
Journal:  Surg Today       Date:  1999       Impact factor: 2.549

5.  Clinical analysis of non-anastomotic aneurysms of implanted prosthetic grafts.

Authors:  Satoshi Yamamoto; Katsuyuki Hoshina; Hideo Kimura; Hiroyuki Okamoto; Kunihiro Shigematsu; Tetsuro Miyata; Toshiaki Watanabe
Journal:  Surg Today       Date:  2014-04-03       Impact factor: 2.549

Review 6.  Nonanastomotic aneurysm formation in a Dacron arterial graft: report of a case.

Authors:  I Han; H Shigematsu; M Nunokawa; H Matsuzaki; K Iwata; S Nakamura; M Ishimaru; A Sugiura; Y Kobayashi; Y Morioka
Journal:  Surg Today       Date:  1994       Impact factor: 2.549

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.