Literature DB >> 29302755

Heat generation by ultrasonic bone curette comparing with high-speed drill.

Kengo Suzuki1,2, Masahiko Wanibuchi3, Yoshihiro Minamida1, Yukinori Akiyama1, Takeshi Mikami1, Masahito Fujishige2, Akinori Yamamura2, Toshio Nakagawa2, Nobuhiro Mikuni1.   

Abstract

BACKGROUND: Ultrasonic bone curettes have been used as with high-speed drills. However, the amount of heat generated by the ultrasonic bone curette is not well known. This study quantitatively assessed the heat generated by an ultrasonic bone curette and compared it to that by a high-speed drill.
METHODS: The thermal change in a swine skull during bone curetting using an ultrasonic device and a high-speed drill were assessed. The investigation focused on the type of surgical manipulation (brush-like strokes vs. pushing motion) and irrigation (room temperature vs. cold water; low-volume irrigation vs. high-volume irrigation).
RESULTS: The thermal elevation during drill use was suppressed when using brush-like strokes compared to pushing motion (brush-like strokes, 44.7 °C; pushing motion, 69.2 °C; p < 0.01). Cold-water irrigation while drilling had a small effect compared to room temperature (RT) water (RT, 44.7 °C; cold, 35.2 °C; p = 0.12). The temperature generated by the curette was higher than that generated by the drill (curette, 72.5 °C; drill, 44.7 °C; p < 0.01). High-volume irrigation was required to reduce the heat generated by the curette (no irrigation, 88.6 °C; low-volume, 72.5 °C; high-volume, 60.5 °C; p < 0.01).
CONCLUSIONS: The ultrasonic bone curate generated more heat than the high-speed drill. During surgical manipulation, the use of brush-like strokes by both the high-speed drill and the ultrasonic bone curette is necessary to avoid excess thermal elevation. Irrigation with RT water is sufficient to avoid heat generation. High-volume irrigation is required to reduce the heat generated by the curette.

Entities:  

Keywords:  CUSA; Heat generation; High-speed drill; Thermography; Ultrasonic bone curette

Mesh:

Year:  2018        PMID: 29302755     DOI: 10.1007/s00701-017-3445-0

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  4 in total

Review 1.  [Piezoelectric ear surgery: a systematic review. German version].

Authors:  Stefan Lyutenski; Anja Lieder; Marc Bloching
Journal:  HNO       Date:  2022-08-12       Impact factor: 1.330

Review 2.  Piezoelectric ear surgery: a systematic review.

Authors:  Stefan Lyutenski; Anja Lieder; Marc Bloching
Journal:  HNO       Date:  2022-10-07       Impact factor: 1.330

3.  Visual Outcomes after Surgery for Paraclinoid Aneurysms: A Fujita Experience.

Authors:  Raja K Kutty; Ambuj Kumar; Yasuhiro Yamada; Tsukosa Kawase; Riki Tanaka; Kyosuke Miyatani; Saeko Higashiguchi; Vigneswar Ravishankar; Katsumi Takizawa; Yoko Kato
Journal:  Asian J Neurosurg       Date:  2020-05-29

4.  Efficacy and Safety Analysis of Ultrasonic Bone Curette in the Treatment of Thoracic Spinal Stenosis.

Authors:  Xiang-Dong Lu; Yi-Bo Zhao; Xiao-Feng Zhao; De-Tai Qi; Xu Yang; Xiao-Nan Wang; Run-Tian Zhou; Yuan-Zhang Jin; Bin Zhao
Journal:  Orthop Surg       Date:  2019-12       Impact factor: 2.071

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.