Radiographic Position of the Electrode as a Predictor of the Outcome of InterStim Therapy.

Abdullah Ahmed Gahzi1, Mai Ahmed Banakhar2, Dean S Elterman1, Magdy Hassouna1.   

Abstract

PURPOSE: Sacral neuromodulation (SNM) therapy is indicated for some refractory urological conditions. The electrode lead position in sacral x-rays during routine follow-up may predict the outcome of SNM therapy. To determine whether the radiographic position of the electrode in the sacral foramen predicted the long-term outcome of SNM therapy.
METHODS: This was a retrospective study of patients who underwent InterStim SNM at Toronto Western Hospital by 2 surgeons from July 2013 to March 2014. The position of electrodes in relation to the sacral bone was assessed on follow-up sacral x-rays. In the lateral view, we determined the location of the radio-opaque marker of the electrode relative to the inner surface of the sacrum (P3, D3, P2, D2, P1, D1, P0, and D0). In the anteroposterior view, the angle between a line through the spinous process shadow and the electrode was measured (0°-30°, 30°-60°, 60°-90°, >90°, or medial). Dissatisfied patients were defined as those who did not improve based on a voiding diary or those who needed salvage treatment after SNM. The primary endpoint was to determine whether the electrode lead position on sacral x-rays predicted the outcome of SNM therapy.
RESULTS: A total of 69 patients (61 female and 8 male patients) were included, with a median age of 55 years. Forty-two of the patients (60.9%) had refractory overactive bladder, 21 (30.4%) suffered from chronic urinary retention, and 6 (8.7%) had lower urinary tract symptoms and chronic pelvic pain syndrome. The univariate analysis did not show any correlation between SNM response and the electrode position or angle. Dummy regression analysis using response to implantation as the dependent outcome variable did not show any significance for any of the predictors.
CONCLUSIONS: Our study did not show a correlation between the long-term response to SNM and the electrode position on follow-up sacral x-rays. In this study, electrode lead position in sacral x-ray at follow-up was not correlated with the outcome of SNM therapy.

Entities:  

Keywords:  Sacral Neuromodulation; Urinary Bladder, Overactive; Urinary Incontinence; Urinary Retention

Year:  2017        PMID: 29298468      PMCID: PMC5756817          DOI: 10.5213/inj.1734942.471

Source DB:  PubMed          Journal:  Int Neurourol J        ISSN: 2093-4777            Impact factor:   2.835


INTRODUCTION

Sacral neuromodulation (SNM) was introduced in the 1980s as a treatment for chronic voiding dysfunction [1]. SNM is a useful alternative therapy for a variety of refractory voiding conditions after failure of conservative and medical therapy [2]. In 1997, the U.S. Food and Drug Administration approved SNM for urgency incontinence, and in 1999 for refractory urgency-frequency syndrome and nonobstructive chronic urinary retention [3,4]. SNM has also shown benefits to patients suffering from fecal incontinence, chronic constipation, and chronic pelvic pain (CPP) syndrome [2,5]. The overall prevalence of overactive bladder (OAB) in the Canadian population is estimated to be 18.1%, with the prevalence lower in men (14.8%) than in women (21.2%) [6]. Various treatment options are available for OAB, including lifestyle or behavioral changes, such as pelvic floor strengthening, bladder retraining, and management of daily fluid intake. Oral medications include anticholinergic (antimuscarinic) agents and a β3 agonist [7-10]. For those individuals who are refractory to medications, intravesical botulinum toxin A injections and SNM have shown significant positive effects in controlling OAB symptoms, including urgency incontinence [11]. SNM is safe and effective in OAB patients, with reported rates of success ranging between 50% and 90% [12-16]. SNM improved quality of life metrics in 86% of patients [12]. SNM therapy has a modestly high rate of complications, reaching up to 18.2%. Most complications are related to lead migration (11.8%) [7]. A quadripolar permanent electrode is usually placed under fluoroscopic guidance. Once it is inserted into the S3 foramen using a specialized foramen needle, followed by a guide wire and foramen dilator, it is tested in all 4 electrode positions (0, 1, 2, and 3) for a response [17]. The intraoperative insertion of the electrode into the sacral foramen using a curved stylet has been shown to achieve motor responses at lower amplitudes. This could maximize the battery life and facilitate improved programming [18]. Intraoperative electrode positioning has previously been described, although to our knowledge this is the first study to examine follow-up radiographic positioning of the electrode with respect to clinical outcomes. Clinically, in some instances we have noted changes in electrode positioning over time between the intraoperative and follow-up x-rays (Fig. 1).
Fig. 1.

Change in electrode position. (A) At the time of insertion. (B) Six months postoperatively.

This study aimed to determine whether the radiographic location of the tined leads on follow-up sacral x-rays predicted the long-term outcomes of SNM therapy.

MATERIALS AND METHODS

This was a retrospective study of patients who underwent SNM from July 2013 to March 2014 at a single institution by 2 surgeons. We excluded patients with bilateral SNM and those who did not have a sacral x-ray done during the follow-up period. A total of 117 patients underwent SNM implantation at our institution during that period. Postoperative sacral x-rays were not performed in 48 patients, so they were excluded. Sixty-nine patients were included in the study. The main indications for SNM therapy were refractory OAB in 42 patients (60.9%), chronic urinary retention (CUR) in 21 patients (30.4%), and CPP syndrome in 6 patients (8.7%) (Table 1).
Table 1.

Patient demographics (n=69)

VariableValue
Median age (yr)55
Sex
 Male8
 Female61
Indications
 Refractory overactive bladder42
 Retention21
 Pelvic pain6
Position
 D00
 P04
 D15
 P17
 D27
 P211
 D314
 P321
Angle
 0°–30°24
 30°–60°27
 60°–90°9
 > 90°4
 Medial5
Effect of insertion
 Working55
 Not working14
Site
 Right43
 Left26

Values are presented as number of patients unless otherwise indicated.

The following technique was used by the authors. The procedure was performed under general anesthesia. The anesthesiologist was typically able to avoid any long-acting muscle relaxants, which may impair sacral nerve stimulation or visualization of their motor response (bellows/toe). The C-arm of the fluoroscope was used to facilitate placement of the quadripolar permanent lead, using a curved stylet. Once it was inserted into the foramen, using the specialized foramen needle, followed by the guide wire and foramen dilator, it was tested by bipolar stimulation for a response (external pulse generator, PW 210, rate 14, amplitude 1–10 volts) in all 4 positions (0, 1, 2, and 3). The dilator sheath was then inserted under fluoroscopic guidance. During lead insertion, in order to confirm the proper positioning, the electrode’s radio-opaque marker on the inner surface of the sacrum at P2, D3, and P3 was adjusted based on the motor response (bellows/toe). We identified the position of electrodes on plain x-ray films in 69 patients during their postoperative follow-up. Electrodes were assessed in relation to the sacral bone in the anteroposterior and lateral views. In the lateral view, we determined the relation of the radio-opaque marker of the electrodes to the inner surface of the sacrum (P3, D3, P2, D2, P1, D1, P0, and D0) (Fig. 2). In the anteroposterior view, the angle between a line through the spinous process shadow and the electrode was measured (0°–30°, 30°–60°, 60°–90°, >90°, or medial) (Fig. 3).
Fig. 2.

Lateral view of the sacral bone with an electrode that was marked as being at P3.

Fig. 3.

Anteroposterior view. The angle is 23° in this example.

RESULTS

Between July 2013 to March 2014, 117 patients received SNM implantation at Toronto Western Hospital by 2 surgeons. Postoperative sacral x-rays were unavailable for 48 patients, so they were excluded. Data from 69 patients were analyzed. Forty-two patients (60.9%) had refractory OAB, 21 (30.4%) had CUR, and 6 (8.7%) had CPP syndrome and lower urinary symptoms. The median age was 55 years (range, 30–87 years). Sixty-one of the patients (88.4%) were female patients, and 8 (11.6%) were male patients. The electrode was implanted in the right S3 (sacral foramen number 3) in 43 patients and in the left S3 in 26 patients (Table 1). The patient’s response and satisfaction were evaluated based on a voiding diary on the last visit to our clinic. The median follow-up was 18 months (range, 6–27 months). Overall, 55 patients (79.7%) were satisfied with the SNM and no further treatment was needed (e.g., an anticholinergic, β3 agonist, Botox injection, or lead revision). Fourteen patients (20.3%) needed further treatment to control their symptoms. Of the 42 patients with refractory OAB, 37 (88.1%) were satisfied with the treatment. Of the 21 patients with CUR, 12 (57.1%) were satisfied with the result and were not using a catheter. All 6 of the patients with pelvic pain with lower urinary tract symptoms (100%) were satisfied (Table 2).
Table 2.

Effects of insertion according to indication, electrode position, and angle

VariableSatisfied with SNMDissatisfied with SNM
Indication for therapy
 Refractory OAB375
 Retention129
 Pelvic pain60
Position
 D000
 P040
 D141
 P152
 D270
 P283
 D3113
 P3165
Angle
 0°–30°204
 30°–60°225
 60°–90°63
 > 90°40
 Medial32

Values are presented as number of patients.

SNM, sacral neuromodulation; OAB, overactive bladder.

Of the patients with an electrode position of 0°–30° in the anteroposterior view, 20 were satisfied and 4 were dissatisfied, and the corresponding values for the electrode positions of >30°–60⁰, >60°–90⁰, >90⁰, and medial were 22 satisfied and 5 dissatisfied, 6 satisfied and 3 dissatisfied, 4 satisfied and 0 dissatisfied, and 3 satisfied and 2 dissatisfied, respectively. No patients had an electrode position of D0 in the lateral view. Of those with an electrode position of P0, 4 were satisfied and 0 were dissatisfied. The corresponding values for the positions of D1, P1, D2, P2, D3, and P3 were 4 satisfied and 1 dissatisfied, 5 satisfied and 2 dissatisfied, 7 satisfied and 0 dissatisfied, 8 satisfied and 3 dissatisfied, 11 satisfied and 3 dissatisfied, and 16 satisfied and 5 dissatisfied, respectively (Table 2). Univariate analysis did not show any correlation between SNM response and electrode position or angle (Table 3, Figs. 4, 5). Dummy regression analysis using response to implantation as the dependent outcome variable did not show any significance for any of the predictors (Table 4).
Table 3.

Univariate analysis, showing no significant correlation between sacral neuromodulation response and electrode position or angle

FactorCorrelation coefficientSignificance
Age-0.0120.925
Sex-0.3800.001
Indication0.0450.715
Position0.0480.694
Angle0.1050.392
Site-0.0200.867
Fig. 4.

The number of satisfied and dissatisfied patients according to the electrode position on lateral sacral x-rays. Despite the trend favoring P3, D3, and P2, the differences were not statistically significant.

Fig. 5.

The number of satisfied and dissatisfied patients according to the electrode position on anteroposterior sacral x-rays. Despite the trend favoring 0°–30° and 30°–60°, the differences were not statistically significant.

Table 4.

Dummy regression analysis using response to implantation as the dependent outcome variable did not show any significance for any of the predictors

PredictorB coefficient (intercept)SESignificanceExp (B)
Age0.2830.2920.3321.327
Sex-2.9600.9570.0020.052
Indication-0.1540.4280.7190.857
Position0.0800.1760.6511.083
Angle0.5660.3010.0601.762
Site-0.1400.7290.8480.870

SE, standard error.

DISCUSSION

Various treatment options are available for OAB, including lifestyle and behavioral changes, such as pelvic floor strengthening, bladder retraining, and management of daily fluid intake [7,9]. The pharmacological treatments used to treat OAB include anticholinergic (antimuscarinic) agents, tricyclic antidepressants, desmopressin, β3-adrenergic receptor agonists, and intravesical Botox (Allergan Inc., Irvine, CA, USA), the latter 2 of which were most recently introduced [10]. For those individuals who are refractory to medications, intravesical botulinum toxin injections and SNM have shown significant positive effects [11]. SNM is safe and effective in OAB patients, with a success rate of 50%–90% [12-16]. SNM improved the quality of life in 86% of patients [12]. The relative simplicity of the technique, as well as the low patient morbidity associated with proper selection, makes SNM an attractive therapeutic alternative to cystectomy and urinary diversion [16]. However, SNM therapy has been hindered by a modestly high rate of complications requiring surgical revision of the neurostimulator package [19]. Siegel et al. [7] summarized the reported efficacy and complications pooled from different studies with identical protocols, devices, efficacy results, and safety profiles. Of the 914 test stimulation procedures performed in 581 patients, 181 adverse events occurred in 166 of procedures (18.2% of the 914 procedures). Most complications were related to lead migration (108 events, 11.8% of procedures). Technical problems and pain represented 2.6% and 2.1% of the adverse events. Among the 219 patients who underwent implantation of the InterStim (Medtronic Inc., Minneapolis, MN, USA) system (lead and generator), pain at the neurostimulator site was the most commonly observed adverse effect at 12 months (15.3%). The main indication for revision was altered stimulation with deterioration of the symptoms (53%). Most of these patients reported a fall just before the altered stimulation [16]. In newer reports, the overall safety profile of InterStim has been found to be excellent, with estimated revision rates of 3%–16%, often due to infection or lack of efficacy [20]. In an elegant study, Jacobs et al. [18] reported certain technical improvements during the implantation of the electrode lead. They showed that the use of a curved stylet instead of a straight stylet increased the battery lifespan because of proper alignment of the electrode closer to the sacral nerve. Using this technique, a lower voltage setting was needed to stimulate the sacral nerve to achieve the proper response. In this study, we aimed to identify whether the radiographic electrode position was correlated with long-term efficacy of SNM therapy. The angle in the anteroposterior view was 0°–60⁰ in 51 of the 69 patients (74%). The position of the electrode on the lateral view at the most proximal radio-opaque mark from the inner surface of the sacral bone was 3D or 3P in 35 of the 69 patients (50.7%). Dummy regression analysis using response to implantation as the dependent outcome variable did not show any significance for any of the predictors. We found that the position of the electrode on follow-up anteroposterior and lateral views of sacral x-rays did not show any relationship with the efficacy of SNM during long-term follow-up. Despite this finding, we still emphasize the important role of good technique and proper placement of the electrodes intraoperatively, which will, in the majority of cases, result in positive outcomes for our patients.
  18 in total

Review 1.  The basis for drug treatment of the overactive bladder.

Authors:  K E Andersson; C Chapple; A Wein
Journal:  World J Urol       Date:  2001-11       Impact factor: 4.226

2.  Economic evaluation of sacral neuromodulation in overactive bladder: A Canadian perspective.

Authors:  Magdy M Hassouna; Hamid Sadri
Journal:  Can Urol Assoc J       Date:  2015 Jul-Aug       Impact factor: 1.862

Review 3.  Sacral nerve stimulation for the treatment of refractory voiding and bowel dysfunction.

Authors:  Karen L Noblett; Lauren A Cadish
Journal:  Am J Obstet Gynecol       Date:  2013-07-27       Impact factor: 8.661

4.  Results of sacral neuromodulation therapy for urinary voiding dysfunction: outcomes of a prospective, worldwide clinical study.

Authors:  Philip E V van Kerrebroeck; Anco C van Voskuilen; John P F A Heesakkers; August A B Lycklama á Nijholt; Steven Siegel; U Jonas; Clare J Fowler; Magnus Fall; Jerzy B Gajewski; Magdy M Hassouna; Francesco Cappellano; Mostafa M Elhilali; Douglas F Milam; Anurag K Das; H E Dijkema; Ubi van den Hombergh
Journal:  J Urol       Date:  2007-09-17       Impact factor: 7.450

5.  Functional evaluation of sacral nerve root integrity. Report of a technique.

Authors:  R A Schmidt; E Senn; E A Tanagho
Journal:  Urology       Date:  1990-05       Impact factor: 2.649

6.  Neuromodulation of sacral nerves for incontinence and voiding dysfunctions. Clinical results and complications.

Authors:  H E Dijkema; E H Weil; P T Mijs; R A Janknegt
Journal:  Eur Urol       Date:  1993       Impact factor: 20.096

7.  Efficacy of sacral nerve stimulation for urinary retention: results 18 months after implantation.

Authors:  U Jonas; C J Fowler; M B Chancellor; M M Elhilali; M Fall; J B Gajewski; V Grünewald; M M Hassouna; U Hombergh; R Janknegt; P E van Kerrebroeck; A A Lylcklama a Nijeholt; S W Siegel; R A Schmidt
Journal:  J Urol       Date:  2001-01       Impact factor: 7.450

Review 8.  Efficacy and adverse events of sacral nerve stimulation for overactive bladder: A systematic review.

Authors:  Nazema Y Siddiqui; Jennifer M Wu; Cindy L Amundsen
Journal:  Neurourol Urodyn       Date:  2010       Impact factor: 2.696

9.  Prevalence of overactive bladder and incontinence in Canada.

Authors:  Jacques Corcos; Erik Schick
Journal:  Can J Urol       Date:  2004-06       Impact factor: 1.344

Review 10.  Pharmacotherapy of overactive bladder in adults: a review of efficacy, tolerability, and quality of life.

Authors:  Jyotsna Jayarajan; Sidney B Radomski
Journal:  Res Rep Urol       Date:  2013-12-06
View more
  3 in total

Review 1.  Abdominal and pelvic radiographs of medical devices and materials- part 2: neurologic and genitourinary devices and materials.

Authors:  Rishi Philip Mathew; Medica Sam; Timothy Alexander; Vimal Patel; Gavin Low
Journal:  Diagn Interv Radiol       Date:  2020-05       Impact factor: 2.630

2.  Predictive Factors in Sacral Neuromodulation: A Systematic Review.

Authors:  Ranjana Jairam; Jamie Drossaerts; Tom Marcelissen; Gommert van Koeveringe; Desiree Vrijens; Philip van Kerrebroeck
Journal:  Urol Int       Date:  2021-05-31       Impact factor: 1.934

3.  Computer-Assisted Lead Placement for Peripheral Nerve Evaluation Test in a Candidate for Sacral Neuromodulation.

Authors:  Gianluca Sampogna; Emanuele Montanari; Michele Spinelli
Journal:  Int Neurourol J       Date:  2020-12-31       Impact factor: 2.835

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.