Boyuan Nie1, Xueying Chen2, Jing Li2, Dou Wu1, Qiang Liu3.
Abstract
BACKGROUND: The major objective of the present study is to investigate the differences in the load and strain changes in the intertrochanteric region of human cadaveric femora between the loss of medial or lateral wall and after treatment with proximal femoral nail antirotation (PFNA).Entities:
Keywords: Biomechanical testing; Cadaver femur; Femoral intertrochanteric region; Lateral wall; Medial wall; Proximal femoral nail antirotation
Mesh:
Year: 2017 PMID: 29282138 PMCID: PMC5745913 DOI: 10.1186/s13018-017-0673-1
Source DB: PubMed Journal: J Orthop Surg Res ISSN: 1749-799X Impact factor: 2.359
Fig. 1Illustration of the osteotomy range for the medial wall (a) and the lateral wall (b) of intertrochanteric region
Fig. 2Diagram showing the positioning of the uniaxial strain gauges
Fig. 3Illustration of the test setup with a custom-made spherical cap
Comparison of measured parameters of femora in each group (mean ± SD)
| Group | Medial wall | Lateral wall |
|
|---|---|---|---|
| BMD (g/cm2) | 0.506 ± 0.057 | 0.508 ± 0.144 | 0.463 |
| Head diameter (mm) | 47.60 ± 1.92 | 47.43 ± 2.21 | 0.687 |
| Neck length (mm) | 45.32 ± 1.34 | 44.99 ± 1.35 | 0.381 |
| Neck-shaft angle (°) | 136.91 ± 6.24 | 136.68 ± 5.28 | 0.653 |
| Anteversion angle (°) | 12.00 ± 2.28 | 12.83 ± 2.32 | 0.093 |
Results of the mechanical test series at failure point (mean ± SD)
| Medial wall | Lateral wall |
| |
|---|---|---|---|
| Failure load (N) | |||
| Removed | 476.05 ± 138.85 | 1596.78 ± 273.17 | < 0.001 |
| PFNA | 3262.63 ± 742.73 | 3267.66 ± 276.79 | 0.990 |
|
| < 0.001 | < 0.001 | |
| Axial stiffness (N/mm) | |||
| Removed | 225.33 ± 36.31 | 911.43 ± 158.31 | < 0.001 |
| PFNA | 550.71 ± 55.89 | 787.69 ± 84.11 | 0.001 |
|
| < 0.001 | 0.020 | |
| Displacement (mm) | |||
| Removed | 4.37 ± 0.69 | 3.23 ± 0.96 | 0.050 |
| PFNA | 6.51 ± 1.04 | 7.15 ± 1.27 | 0.491 |
|
| < 0.001 | < 0.001 | |
Fig. 4Failure loads for all tested femora in different groups (*p = 0.990)
Fig. 5Axial stiffness for all tested femora in different groups. Statistically significant reduction of the stiffness of PFNA implanted femora compared with the defect modeling specimens in the lateral wall group (*p = 0.002)
Results of the strain value for intertrochanteric region (mean ± SD)
| Load level | Medial wall | Lateral wall | |||
|---|---|---|---|---|---|
| Removed | PFNA | Removed | PFNA | ||
| 350 N | Gauge 1 | − 200.17 ± 67.24 | − 39.67 ± 6.62 | − 529.33 ± 154.42 | − 339.00 ± 47.48 |
| Gauge 2 | − 639.00 ± 93.94 | − 998.50 ± 53.55 | − 31.50 ± 6.98 | 451.67 ± 68.01 | |
| Gauge 3 | 456.00 ± 164.06 | − 61.17 ± 7.36 | − 67.17 ± 18.76 | − 64.17 ± 10.05 | |
| Gauge 4 | − 1464.17 ± 627.05 | − 77.83 ± 9.77 | 330.33 ± 58.86 | − 167.00 ± 34.17 | |
| Gauge 5 | 138.83 ± 29.73 | 562.83 ± 68.92 | − 108.33 ± 15.68 | 451.67 ± 68.01 | |
| 700 N | Gauge 1 | NA | 54.33 ± 9.31 | − 1539.00 ± 298.08 | − 561.83 ± 60.34 |
| Gauge 2 | NA | − 1279.17 ± 82.85 | 169.33 ± 41.31 | − 366.33 ± 47.07 | |
| Gauge 3 | NA | − 106.17 ± 13.96 | 58.67 ± 2.81 | 51.33 ± 9.81 | |
| Gauge 4 | NA | − 126.67 ± 13.78 | 734.50 ± 136.55 | − 249.17 ± 49.37 | |
| Gauge 5 | NA | 947.67 ± 47.83 | − 168.67 ± 33.32 | 781.00 ± 115.90 | |
| 1800 N | Gauge 1 | NA | 140.83 ± 15.75 | NA | − 770.00 ± 98.26 |
| Gauge 2 | NA | − 1618.0 ± 105.73 | NA | − 554.83 ± 91.74 | |
| Gauge 3 | NA | − 137.50 ± 16.72 | NA | 138.50 ± 43.43 | |
| Gauge 4 | NA | − 161.83 ± 19.09 | NA | − 368.00 ± 44.41 | |
| Gauge 5 | NA | 1742.83 ± 143.35 | NA | 1643.83 ± 302.79 | |
| Failure load | Gauge 1 | − 241.61 ± 81.12 | 258.67 ± 77.06 | − 2761.50 ± 422.05 | − 2002.83 ± 373.60 |
| Gauge 2 | − 1048.33 ± 252.15 | − 2973.83 ± 620.60 | 413.83 ± 68.04 | − 1233.00 ± 257.59 | |
| Gauge 3 | 478.17 ± 162.90 | − 250.33 ± 72.50 | 260.67 ± 54.47 | 303.33 ± 82.93 | |
| Gauge 4 | − 1715.26 ± 747.24 | − 299.17 ± 81.17 | 1937.50 ± 571.17 | − 742.17 ± 60.50 | |
| Gauge 5 | 155.07 ± 28.48 | 2155.67 ± 556.98 | − 264.50 ± 45.33 | 2560.17 ± 212.64 | |
Fig. 6Bar chart showing the strain values of all tested femora in different groups at different load level (*showing the part of strain concentration, **p < 0.05, ***p > 0.05)