Perfusion index in healthy newborns during critical congenital heart disease screening at 24 hours: retrospective observational study from the USA.

Priya Jegatheesan1, Matthew Nudelman1, Keshav Goel1, Dongli Song1, Balaji Govindaswami1.   

Abstract

OBJECTIVE: To describe the distribution of perfusion index (PI) in asymptomatic newborns at 24 hours of life when screening for critical congenital heart disease (CCHD) using an automated data selection method.
DESIGN: This is a retrospective observational study.
SETTING: Newborn nursery in a California public hospital with ~3500 deliveries annually.
METHODS: We developed an automated programme to select the PI values from CCHD screens. Included were term and late preterm infants who were screened for CCHD from November 2013 to January 2014 and from May 2015 to July 2015. PI measurements were downloaded every 2 s from the pulse oximeter and median PI were calculated for each oxygen saturation screen in our cohort.
RESULTS: We included data from 2768 oxygen saturation screens. Each screen had a median of 29 data points (IQR 17 to 49). The median PI in our study cohort was 1.8 (95% CI 1.8 to 1.9) with IQR 1.2 to 2.7. The median preductal PI was significantly higher than the median postductal (1.9 vs 1.8, p=0.03) although this difference may not be clinically significant.
CONCLUSION: Using an automated data selection method, the median PI in asymptomatic newborns at 24 hours of life is 1.8 with a narrow IQR of 1.2 to 2.7. This automated data selection method may improve accuracy and precision compared with manual data collection method. Further studies are needed to establish external validity of this automated data selection method and its clinical application for CCHD screening. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2017. All rights reserved. No commercial use is permitted unless otherwise expressly granted.

Entities:  

Keywords:  asymptomatic newborns; critical congenital heart disease screening; noninvasive monitoring; oxygen saturation screening; perfusion index nomogram; perfusion index screening

Mesh:

Year:  2017        PMID: 29273653      PMCID: PMC5778276          DOI: 10.1136/bmjopen-2017-017580

Source DB:  PubMed          Journal:  BMJ Open        ISSN: 2044-6055            Impact factor:   2.692


This study describes a novel automated data selection method to calculate the median perfusion index (PI) from data downloaded directly from pulse oximeter. The automated data selection method used in this study potentially minimises data collection errors and bias. This study describes the distribution of PI in asymptomatic newborns at 24 hours during critical congenital heart disease screen. In this study, the PI data obtained from pulse oximeter were deidentified and therefore not directly linked to the clinical outcome of the infants. This was a single-centre study from a public hospital that serves a primarily Hispanic population.

Introduction

Critical congenital heart disease (CCHD) refers to a group of heart defects that require surgery or catheter-based intervention within the first year of life. Universal oxygen saturation screening of all newborns at 24 hours of life with pulse oximetry has been recommended and is now standard of care in many centres to improve recognition of CCHD in asymptomatic newborns.1–6 Despite this current screening, we are still unable to detect all CCHD cases prior to newborn hospital discharge, especially in lesions with left ventricular outflow obstruction.4 7–9 Perfusion index (PI) reflects the ratio of the pulsatile to non-pulsatile signal components of blood flow in the peripheral tissue. It is derived from the photoelectric plethysmographic signal of a pulse oximeter. PI thus non-invasively provides continuous real-time perfusion status of the selected monitoring site.10 11 In recent years, PI has been suggested to have use as an adjunct to pulse oximetry screening to detect non-cyanotic CCHD cases.4 7 12–14 Multiple studies have described PI values in the first few days of life in term and preterm infants.12 15–19 A lower PI in neonates has been correlated with lower superior vena cava flow20 and shown to be a predictor of illness severity and subclinical chorioamnionitis.21 22 Despite PI being a simple, non-invasive objective measurement with potential clinical applications, it has been underused. This may in part be due to the inherent variability in PI that is seen during a brief monitoring period.19 Values with variability that need to be manually collected are difficult to standardise. Using automated data selection methods may assist with standardised data collection. The objective of our study is to describe the distribution of PI in asymptomatic newborns at 24 hours of life when they undergo CCHD screening using pulse oximetry with an automated data selection method.

Methods

Design

This is a retrospective observational study.

Subjects and setting

Data were collected directly from pulse oximeters used for CCHD screen in asymptomatic term and late preterm infants in the newborn nursery at a public hospital with ∼3500 deliveries annually. Data collection was Institutional Review Board approved.

CCHD screen

All infants in the newborn nursery underwent CCHD screen at 24 hours of age using two motion-tolerant pulse oximeter monitors (Rad 7, Masimo Corporation, Irvine, California, USA) with an 8 s averaging time to measure both preductal and postductal oxygen saturation (SpO2) simultaneously. Reusable probes with disposable wraps were placed on the right palm or wrist for preductal and on either foot for postductal saturation. Nursing staff continued the pulse oximeter monitoring until they obtained stable tracings of the SpO2 and heart rate (HR) for the CCHD screen. While the SpO2 values were recorded by clinical nurses manually when both preductal and postductal pulse oximetry showed stable waveforms and the HRs were correlating, the machine recorded the values continuously. The screen was repeated if either the preductal and postductal saturations was 90% to 94%, or there was a>3% difference between the two values; SpO2 <90% was considered a failed screen and required immediate Neonatal Intensive Care Unit (NICU) evaluation. Median age when CCHD screen occurred in the nursery was 26.1 hours of life (IQR 24.8–28.3 hours). These pulse oximeters were used for CCHD screening and for repeat measurements done based on the screening protocol or nurse’s clinical judgement in newborn nursery.

Data collection

Convenience sample was used based on the amount of data stored in the pulse oximeters. The values of PI measurements, SpO2, HR, signal quality indicators, date and time of the screen were recorded every 2 s. Deidentified data were downloaded using Trendcom software (Masimo Corporation, Irvine, California, USA). The screening data from the two pulse oximeters with identical time periods were used for analysis. Since the duration of recording for each screen varies, the total number of preductal and postductal screens recorded in the same time period were slightly different. The first screening dataset had recordings from November 2013 to January 2014 including both preductal and postductal data because the pulse oximeters were not selectively labelled as preductal or postductal. Pulse oximeters were subsequently labelled for the second dataset. The second dataset had non-paired preductal and postductal data recordings from May 2015 to July 2015.

Data selection

As a pilot study, we manually selected a subset of the PI data from the first dataset (222 screens; 7426 data values). For each individual screening data, the values from the point at which the HR and SpO2 had stabilised were included. Values with ‘low SpO2 signal quality’ (as preprogrammed in the pulse oximeter by the manufacturer) were eliminated. Subsequently, we developed an automated Excel program for data selection. The Excel program eliminated data rows without any numerical values recorded and then assigned unique dataset identifications based on the time intervals separating individual screens. Data points with a HR or a SpO2 equal to zero were excluded. Data points with low SpO2 signal quality were eliminated (figure 1). The Excel program was validated in the pilot study subset and then used to select the data points in the complete dataset for this study.6 7
Figure 1

Data collection and automated data selection method.

Data collection and automated data selection method.

Data analysis

Median PI was calculated for each individual screening data. The percentile distributions of the median PI values and their 95% CIs were calculated for the unlabelled first study dataset and for the preductal and postductal PIs values in the second dataset separately. The median preductal and postductal PI in the second dataset were compared using the Mann-Whitney-Wilcoxon rank sum test. Statistical data analysis software STATA V.14.2 (StataCorp LP, College Station, Texas USA) was used for analysis.

Results

The study dataset included data from 2768 screens. The median recording time per individual screen was 72 s (IQR 42–126 s). The data selection process is outlined in (figure 1). Based on the automated program, 3.4% of data points were eliminated for HR of zero, 0.3% for SpO2 of zero and 17.8% for low-signal quality. Of the 2768 screening data, only three screens (0.1%) were eliminated based on the automated data selection process. Each screen had a median of 29 data points (IQR 17 to 49) after the data selection process. The median PI value in our whole study cohort was 1.8 (95% CI 1.8 to 1.9) with IQR 1.2 to 2.7. The percentile distribution of the median PIs in the first and second dataset are shown in (table 1). The distribution of the median PI for the unlabelled first dataset is shown in (figure 2). The second study dataset included 1352 screens, of which 682 and 670 were preductal and postductal measurements, respectively. The median preductal PI was significantly higher than the median postductal (1.9 vs 1.8, p=0.03) although this difference may not be clinically significant. The distribution of the preductal and postductal median PIs are shown in (figure 3).
Table 1

Distribution of median PI

Unlabelled PI (n=1416)Preductal PI (n=682)Postductal PI (n=670)
Median (95% CI)Median (95% CI)Median (95% CI)
1st %ile0.50 (0.40 to 0.60)0.50 (0.37 to 0.54)0.40 (0.40 to 0.50)
5th %ile0.70 (0.65 to 0.70)0.70 (0.70 to 0.80)0.60 (0.60 to 0.70)
25th %ile1.20 (1.20 to 1.30)1.30 (1.20 to 1.30)1.10 (1.10 to 1.20)
50th %ile1.80 (1.80 to 1.90)1.90 (1.80 to 2.00)1.80 (1.70 to 1.90)
75th %ile2.80 (2.65 to 2.91)2.70 (2.55 to 2.90)2.60 (2.50 to 2.80)
95th %ile5.90 (5.46 to 6.93)5.67 (4.80 to 7.15)4.67 (4.20 to 5.33)
99th %ile17.30 (12.4 to 20.00)17.50 (10.11 to 20.0)8.16 (6.25 to 14.24)

PI, perfusion index.

Figure 2

(A) Histogram of the median perfusion index. (B) Distribution of median perfusion index of the unlabeled dataset. The boxes represent the IQR from 25th–75th percentile (IQR). The marker within the box is the median and the ‘whiskers’ reach 1.5 times IQR.

Figure 3

Distribution of preductal and postductal median perfusion index. The boxes represent the IQR from 25th–75th percentile. The marker within the box is the median and the ‘whiskers’ reach 1.5 times IQR.

Distribution of median PI PI, perfusion index. (A) Histogram of the median perfusion index. (B) Distribution of median perfusion index of the unlabeled dataset. The boxes represent the IQR from 25th–75th percentile (IQR). The marker within the box is the median and the ‘whiskers’ reach 1.5 times IQR. Distribution of preductal and postductal median perfusion index. The boxes represent the IQR from 25th–75th percentile. The marker within the box is the median and the ‘whiskers’ reach 1.5 times IQR.

Discussion

Prior to identifying cut-off values for PI to improve the detection of lesions with left ventricular outflow tract obstruction during CCHD screening, it is necessary to describe the normal PI distribution using a reliable automated method in asymptomatic infants at the time of screening. Although multiple studies have described PI in preterm and term newborns with various newborn conditions,7 12 14–24 few have described the PI distribution in asymptomatic term and late preterm infants, at 24 hours of life when CCHD screening using pulse oximetry is performed. Granelli et al described the distribution of preductal and postductal PI in 10 000 normal healthy newborns prior to discharge, with a time of measurement ranging from 1 to 120 hours.12 Despite the difference in the age of measurement, the PI distribution in our cohort measured at 24 hours, median PI of 1.8 and IQR 1.2 to 2.7, is almost identical to the values reported by Granelli et al, median PI of 1.7 and IQR 1.18 to 2.5. These two studies show that PI distribution is relatively consistent and stable in asymptomatic newborns during the CCHD screening time period in the first few days of life. A PI value of <0.44 has been shown to be indicative of low superior vena cava flow of <40 mL/kg/min in very low birthweight infants.20 In a case–control study, Granelli et al suggested using a PI value of 0.7 as an adjunct to oxygen saturation screening to evaluate whether a newborn has left heart obstructive lesions.12 In our study, the first percentile PI value was 0.5 (95% CI 0.40 to 0.60) and the fifth percentile PI value was 0.7 (95% CI 0.65 to 0.70). A low PI value of <0.5 or<0.7 can be considered as a screening marker of hypoperfusion in the newborn, however, the sensitivity and specificity of such values to detect CCHD or other conditions is yet to be determined. A more recent study by Schena et al, in 42 000 newborns, using a combined pulse oximeter and PI screen to detect CCHD, chose PI <0.9 as abnormal for their CCHD screen done between 48 and 72 hours of life.14 The one case of coarctation of aorta that was identified based on PI alone had a PI value of 0.3. This study further showed that adding PI to pulse oximeter screens still missed two cases of left-sided obstructive CCHD. Hence, the clinical use of lower limit of PI in screening for CCHD in asymptomatic newborns needs further study. Our study demonstrates that at 24 hours postductal PI is lower than the preductal PI by 0.1. Others have reported that postductal PI is higher than preductal PI.12 Granelli et al showed that the postductal PI is 0.02 higher than preductal PI. Such small differences in PI are not clinically significant in healthy newborns. However, this difference was much greater in Hakan’s study of 196 healthy term infants that described median PI value and IQR for the paired preductal and postductal as 1.35 (1.02–1.91) and 0.88 (0.62–1.22) in the first day of life, and the difference was seen up to 3 days of life.18 The reason for the discrepancy in the preductal and postductal PI differences among these three studies is unclear. There are differences in methodology between our study and the other two studies. We screened newborns at 24 hours, Granelli et al screened between 1 and 120 hours and Hakan et al screened at ‘first day’ but the hour of life is unspecified. While our study compared the overall distribution of the preductal and postductal PI, the other two studies used paired preductal and postductal data samples. We should consider collecting and evaluating the preductal and postductal PIs in a larger patient population to further improve CCHD screening methodology. It is common to have PI values fluctuate as they reflect a brief momentary measurement of one’s peripheral perfusion at the site of the pulse oximeter. Peripheral perfusion is dynamically affected by a variety of factors. For this reason, a single transient measurement has limited value because it may not accurately reflect one’s true perfusion status. Instead, the median PI calculated from monitoring PI over a period of time after the pulse oximeter tracing stabilises may provide a more accurate depiction of true perfusion status.13 Our study recorded PI for a median of 72 s. As an additional step to ensure measurement accuracy, we used an automated (Excel-based) data selection process to capture only artefact-free PI values with correspondingly stable SpO2 and HR. Even though almost 20% (3%–4% due to zero HR/saturation; 18% due to low signal quality) of the data points were eliminated by this method, we only lost 0.1% of the individual screening data. Other studies have used a similar methodology of selecting the median PI of artefact-free readings instead of using a single PI value.15 18 19 Using electronic data output and data selection methods to standardise practice and eliminate human transcription error is essential in evaluating the efficacy of different PI values and their clinical significance. The data selection methods used in this and other studies will help to develop algorithms for automated PI data selection and output for clinical use. A limitation of our study is that it is a retrospective single-centre study and the data were deidentified. We do not have paired samples to evaluate the individual preductal and postductal differences. We are unable to link PI data directly to clinical outcome of these infants. However, we have a prospectively maintained NICU database which allowed us to confirm that during the study period, there were no CCHD cases or early onset bacterial sepsis or pneumonia identified in the newborn nursery prior to discharge. The majority of infants born in our hospital continue to obtain care within our hospital system. Our hospital database provided sufficient follow-up information to determine whether any infant discharged during the study period was later diagnosed with CCHD. An exception would be if an infant was later diagnosed with CCHD at a different hospital. The generalisability of our study is limited to the availability of the software to download data from the pulse oximeters. Once the data are downloaded, our automated data selection process can be reproduced easily. It is now standard of care to screen most newborns in the USA at 24 hours of life for CCHD. Centres using pulse oximeters that have numerical PI capability should be able to replicate our results. Generalisability for clinical use requires development of an automated algorithm in the pulse oximeters to provide the median PI value as output.

Conclusion

The median PI calculated using an automated data selection method in asymptomatic newborns at 24 hours of life is 1.8, with a narrow IQR of 1.2 to 2.7. This distribution of PI is consistent with the previously published Swedish nomogram (median 1.7 and IQR 1.2, 2.5) described at 1–120 hours of life. This suggests that the PI values in asymptomatic newborns have a narrow range of distribution during the time of CCHD screening. Further studies are needed to establish external validity of this automated data selection method and its clinical application for CCHD screening.
  23 in total

Review 1.  Noninvasive monitoring of peripheral perfusion.

Authors:  Alexandre Lima; Jan Bakker
Journal:  Intensive Care Med       Date:  2005-09-17       Impact factor: 17.440

Review 2.  Pulse oximetry screening for critical congenital heart defects in asymptomatic newborn babies: a systematic review and meta-analysis.

Authors:  Shakila Thangaratinam; Kiritrea Brown; Javier Zamora; Khalid S Khan; Andrew K Ewer
Journal:  Lancet       Date:  2012-05-02       Impact factor: 79.321

3.  Strategies for implementing screening for critical congenital heart disease.

Authors:  Alex R Kemper; William T Mahle; Gerard R Martin; W Carl Cooley; Praveen Kumar; W Robert Morrow; Kellie Kelm; Gail D Pearson; Jill Glidewell; Scott D Grosse; R Rodney Howell
Journal:  Pediatrics       Date:  2011-10-10       Impact factor: 7.124

Review 4.  Perfusion index in newborn infants: a noninvasive tool for neonatal monitoring.

Authors:  Carolina Z Piasek; Frank Van Bel; Augusto Sola
Journal:  Acta Paediatr       Date:  2014-02-23       Impact factor: 2.299

5.  Estimated number of infants detected and missed by critical congenital heart defect screening.

Authors:  Elizabeth C Ailes; Suzanne M Gilboa; Margaret A Honein; Matthew E Oster
Journal:  Pediatrics       Date:  2015-05-11       Impact factor: 7.124

6.  Perfusion index variations in clinically and hemodynamically stable preterm newborns in the first week of life.

Authors:  Francesco Cresi; Emanuela Pelle; Roberto Calabrese; Luciana Costa; Daniela Farinasso; Leandra Silvestro
Journal:  Ital J Pediatr       Date:  2010-01-18       Impact factor: 2.638

7.  Role of pulse oximetry in examining newborns for congenital heart disease: a scientific statement from the AHA and AAP.

Authors:  William T Mahle; Jane W Newburger; G Paul Matherne; Frank C Smith; Tracey R Hoke; Robert Koppel; Samuel S Gidding; Robert H Beekman; Scott D Grosse
Journal:  Pediatrics       Date:  2009-07-06       Impact factor: 7.124

8.  Early dynamic changes in pulse oximetry signals in preterm newborns with histologic chorioamnionitis.

Authors:  Claudio De Felice; Mitchell R Goldstein; Stefano Parrini; Alberto Verrotti; Mario Criscuolo; Giuseppe Latini
Journal:  Pediatr Crit Care Med       Date:  2006-03       Impact factor: 3.624

9.  The perfusion index derived from a pulse oximeter for predicting low superior vena cava flow in very low birth weight infants.

Authors:  S Takahashi; S Kakiuchi; Y Nanba; K Tsukamoto; T Nakamura; Y Ito
Journal:  J Perinatol       Date:  2009-11-12       Impact factor: 2.521

10.  The perfusion index of healthy term infants during transition at birth.

Authors:  Jacco K Kroese; Jeroen J van Vonderen; Ilona C Narayen; Frans J Walther; Stuart Hooper; Arjan B te Pas
Journal:  Eur J Pediatr       Date:  2015-10-24       Impact factor: 3.183

View more
  3 in total

1.  Oxygen saturation and perfusion index screening in neonates at high altitudes: can PDA be predicted?

Authors:  Kadir Şerafettin Tekgündüz; Merve Bilen; Mustafa Kara; Fuat Laloğlu; Naci Ceviz
Journal:  Eur J Pediatr       Date:  2020-06-05       Impact factor: 3.183

2.  Peripheral Saturation and Perfusion Index on the First Day of Life Play a Role in Early Discharge of Healthy Term Newborns.

Authors:  Serafina Perrone; Maurizio Giordano; Giuseppe De Bernardo; Mara Corradi; Giulia Cecconi; Ilenia Fontanarosa; Elisa Laschi; Giuseppe Buonocore; Susanna Esposito
Journal:  Int J Clin Pract       Date:  2022-05-29       Impact factor: 3.149

3.  A novel system to collect dual pulse oximetry data for critical congenital heart disease screening research.

Authors:  Kavish Doshi; Gregory B Rehm; Pranjali Vadlaputi; Zhengfeng Lai; Satyan Lakshminrusimha; Chen-Nee Chuah; Heather M Siefkes
Journal:  J Clin Transl Sci       Date:  2020-10-19
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.