Literature DB >> 29272377

Traumatic leaflet injury: comparison of porcine leaflet self-expandable and bovine leaflet balloon-expandable prostheses.

Channing Convelbo1, Hanane El Hafci1, Hervé Petite1, Rachid Zegdi2,3.   

Abstract

OBJECTIVES: The main goal of this study was to compare the occurrence of post-deployment leaflet injury between prostheses made of porcine and bovine pericardium.
METHODS: Two types of prostheses, self-expandable prostheses with porcine pericardial leaflet on one side and balloon-expandable prostheses with bovine pericardial leaflet on the other side, were used. In each group, crimped prostheses were compared with control non-crimped prostheses. Following a 15-min period of crimping, prostheses were deployed, and their leaflets were removed and subjected to analyses. The analyses included determination of global and local hydraulic conductance of the leaflets, leaflet plasma insudation and microscopic analysis. The results were expressed as percentage (standard error of the mean) or median (interquartile range).
RESULTS: A significant increase in the global hydraulic conductance was only observed in the crimped balloon-expandable prostheses: 20.1 (15.5-41.2) ml/h/m2/mmHg vs 12.3 (9.0-15.6) ml/h/m2/mmHg (P = 0.021). Similarly, areas of marked staining (a marker of local hydraulic conductance) were only seen in the bovine pericardium balloon-expandable prostheses. The incidence of leaflet plasmatic insudation was increased in the crimped prostheses compared with the control prostheses. The microscopic study revealed a higher occurrence of traumatic lesions in the crimped prostheses in comparison with the control prostheses: 33.3 ± 21.1% vs 5.5 ± 5.5% (P = 0.039) and 44.4 ± 20.5% vs 5.5 ± 5.5% (P = 0.017) in the bovine and the porcine leaflets, respectively.
CONCLUSIONS: Post-deployment leaflet injury occurred in both types of prostheses. However, alteration of the global and local hydraulic conductance was important in the bovine pericardium balloon-expandable prostheses.

Entities:  

Mesh:

Year:  2018        PMID: 29272377     DOI: 10.1093/ejcts/ezx451

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  5 in total

Review 1.  Principles of TAVR valve design, modelling, and testing.

Authors:  Oren M Rotman; Matteo Bianchi; Ram P Ghosh; Brandon Kovarovic; Danny Bluestein
Journal:  Expert Rev Med Devices       Date:  2018-10-29       Impact factor: 3.166

2.  Perigraft reaction and incorporation of porcine and bovine pericardial patches.

Authors:  Georg Schlachtenberger; Fabian Doerr; Annamaria Brezina; Hruy Menghesha; Matthias B Heldwein; Gerardus Bennink; Michael D Menger; Mohammed Moussavian; Khosro Hekmat; Thorsten Wahlers
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-04-19

3.  In Vitro Durability and Stability Testing of a Novel Polymeric Transcatheter Aortic Valve.

Authors:  Oren M Rotman; Brandon Kovarovic; Matteo Bianchi; Marvin J Slepian; Danny Bluestein
Journal:  ASAIO J       Date:  2020-02       Impact factor: 3.826

4.  A New Detergent for the Effective Decellularization of Bovine and Porcine Pericardia.

Authors:  Martina Todesco; Saima Jalil Imran; Tiago Moderno Fortunato; Deborah Sandrin; Giulia Borile; Filippo Romanato; Martina Casarin; Germana Giuggioli; Fabio Conte; Massimo Marchesan; Gino Gerosa; Andrea Bagno
Journal:  Biomimetics (Basel)       Date:  2022-08-01

5.  Commentary: Leaflet perforation after transcatheter aortic valve implantation in calcified roots: When perfect can be the enemy of good?

Authors:  Aleksander Dokollari; Gianluigi Bisleri
Journal:  JTCVS Tech       Date:  2020-06-01
  5 in total