Literature DB >> 29269618

[A Proposal for Evaluating the Positional Accuracy of Add-on Six-degrees-of-freedom Radiotherapy Couch in Couch Rotation for Image-guided Radiotherapy].

Shinji Mawatari1, Yutaro Kiyao1, Yoshifumi Oku1, Masahiko Toyota1, Yasumasa Saigo1.   

Abstract

PURPOSE: In this study, we proposed and evaluated position correction accuracy assessment method with a phantom for IGRT system with add-on six-degrees-of-freedom radiotherapy (6D) couches in couch rotation. METHODS AND MATERIALS: A phantom was used in a self-build phantom. We were scanned with computed tomography (CT) for radiotherapy planning and planned treatment isocenter to fall in line with CT center by treatment planning system. At first, we examined data of CT slice thickness for digitally reconstructed radiograph of QA phantom. Next, we measured uncertainty for IGRT system. We performed position correction accuracy for IGRT system with QA phantom and digital angle meter.
RESULTS: Detection and correction errors for pitch and roll direction were within 0.3 degree in all verifications.
CONCLUSIONS: We proposed a quality control method for position correction accuracy of 6D couch. The method was able to evaluate the accuracy of detection and correction of 6D couch and revealed the deviation of the origin of the couch rotation.

Entities:  

Keywords:  image-guided radiotherapy; position correction accuracy; radiotherapy

Mesh:

Year:  2017        PMID: 29269618     DOI: 10.6009/jjrt.2017_JSRT_73.12.1231

Source DB:  PubMed          Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi        ISSN: 0369-4305


  1 in total

1.  Commissioning and Evaluation of a Third-Party 6 Degrees-of-Freedom Couch Used in Radiotherapy.

Authors:  Maria F Chan; Seng-Boh Lim; Xiang Li; Xiaoli Tang; Peng Zhang; Chengyu Shi
Journal:  Technol Cancer Res Treat       Date:  2019-01-01