Literature DB >> 2926406

Fracture of a fenestrated metal backing of a tibial knee component. A case report.

I A Gradisar1, M L Hoffmann, M J Askew.   

Abstract

The case of a metal-backed tibial component that failed by metal fracture is reported. Fracture occurred through fenestrations in the metal plate surrounding the central metal stem. The fracture followed loss of bony support beneath the medial tibial plateau, which had allowed varus deformity of the replaced joint. A high tibial osteotomy procedure had preceded replacement of the joint.

Entities:  

Mesh:

Substances:

Year:  1989        PMID: 2926406     DOI: 10.1016/s0883-5403(89)80050-6

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  4 in total

1.  Early Tibial Component Fractures in a Cementless, 3D-Printed, Titanium Implant.

Authors:  Alan D Lam; Gavan P Duffy
Journal:  Arthroplast Today       Date:  2022-10-12

2.  SCREW MIGRATION IN TOTAL KNEE ARTHROPLASTY: CLINICAL REPORT.

Authors:  Fernando Fonseca; José Tomé; Manuel Barreto
Journal:  Rev Bras Ortop       Date:  2015-11-16

3.  Fracture of the tibial tray following PCA knee replacement. A report of two cases.

Authors:  M Maruyama; K Terayama; H Sunohara; T Adachi; S Suzuki; T Fukuzawa
Journal:  Arch Orthop Trauma Surg       Date:  1994       Impact factor: 3.067

4.  Tibial tray fracture in a modern prosthesis with retrieval analysis.

Authors:  John J Callaghan; David E DeMik; Nicholas A Bedard; Andrew N Odland; William M Kane; Steven M Kurtz
Journal:  Arthroplast Today       Date:  2018-02-10
  4 in total