| Literature DB >> 2926406 |
I A Gradisar1, M L Hoffmann, M J Askew.
Abstract
The case of a metal-backed tibial component that failed by metal fracture is reported. Fracture occurred through fenestrations in the metal plate surrounding the central metal stem. The fracture followed loss of bony support beneath the medial tibial plateau, which had allowed varus deformity of the replaced joint. A high tibial osteotomy procedure had preceded replacement of the joint.Entities:
Mesh:
Substances:
Year: 1989 PMID: 2926406 DOI: 10.1016/s0883-5403(89)80050-6
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757