Literature DB >> 2925705

Difficulties during removal of fluted femoral intramedullary rods.

A M Star1, R P Whittaker, H M Shuster, J Duda, E Menkowitz.   

Abstract

The difficulties that were encountered during removal of a fluted intramedullary femoral rod from six patients were reviewed. The mechanisms of failure included metal breakage at three different locations along the rod or the rod extractor. Three of the six rods were left in place due to difficulties in removal. The design of the fluted intramedullary rod appears to be the main cause of the problem.

Entities:  

Mesh:

Year:  1989        PMID: 2925705

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  1 in total

Review 1.  [Implant removal after intramedullary osteosyntheses. Literature review, technical details, and tips and tricks].

Authors:  C Krettek; P Mommsen
Journal:  Unfallchirurg       Date:  2012-04       Impact factor: 1.000