Accuracy of oscillometric blood pressure measurement in atrial fibrillation.

Rudolf K Feenstra1, Cor P Allaart2,3, Gijs F N Berkelmans4, Berend E Westerhof5,6, Yvo M Smulders7,3.   

Abstract

OBJECTIVE: The primary aim of this study was to assess the accuracy of automated oscillometry (AO) in outpatients with atrial fibrillation (AF). The secondary aim was to explore whether AO accuracy is influenced by beat-to-beat blood pressure (BP) variability or heart frequency (HF).
METHODS: Systolic blood pressure (SBP) and diastolic blood pressure (DBP) were measured by AO and beat-to-beat BP using a validated Volume Clamp Method (VCM) technique. AO accuracy was analyzed separately in tertiles of beat-to-beat BP variability and HF.
RESULTS: The main study included 58 AF and 38 sinus rhythm (SR) patients in whom the Welch Allyn Spot Vital Signs (WASVS) was used. An auxiliary study in 23 AF patients used the Philips M3002A IntelliVue ×2. For AF and SR patients, respectively, SBP by WASVS deviated by +0.1 (±14.8) mmHg and -7.9 (±15.7) mmHg from VCM. WASVS-DBP was higher than VCM in AF and SR by 6.3 (±9.2) mmHg and 5.0 (±7.7) mmHg, respectively. High beat-to-beat BP variability and high HF decreased WASVS accuracy for both SBP and DBP. SBP and DBP measurements by Philips M3002A IntelliVue ×2 deviated by -6.8 (±13.2) mmHg and 9.4 (±8.1) mmHg, respectively.
CONCLUSION: Overall, AO accuracy in AF is limited; in individual patients, AO inaccuracy may be considerable. AO accuracy is especially reduced in patients showing large beat-to-beat BP variability or high HF.

Entities:  

Mesh:

Year:  2018        PMID: 29256921     DOI: 10.1097/MBP.0000000000000305

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  2 in total

1.  Orthostatic hypotension among elderly patients in Italian internal medicine wards: an observational study.

Authors:  Luca Pasina; Monica Casati; Laura Cortesi; Mauro Tettamanti; Ramona Pellegrini; Ivan Oppedisano; Natale Dugnani; Androula Marinou; Gian Galeazzo Riario Sforza; Antonio Brucato
Journal:  Intern Emerg Med       Date:  2019-08-19       Impact factor: 3.397

2.  Insights into postoperative respiration by using continuous wireless monitoring of respiratory rate on the postoperative ward: a cohort study.

Authors:  Linda M Posthuma; Maarten J Visscher; Philipp B Lirk; Els J M Nieveen van Dijkum; Markus W Hollmann; Benedikt Preckel
Journal:  J Clin Monit Comput       Date:  2019-11-13       Impact factor: 2.502

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.