Compass model-based quality assurance for stereotactic VMAT treatment plans.

Assi Valve1, Jani Keyriläinen2, Jarmo Kulmala2.   

Abstract

PURPOSE: To use Compass as a model-based quality assurance (QA) tool for stereotactic body radiation therapy (SBRT) and stereotactic radiation therapy (SRT) volumetric modulated arc therapy (VMAT) treatment plans calculated with Eclipse treatment planning system (TPS).
MATERIALS AND METHODS: Twenty clinical stereotactic VMAT SBRT and SRT treatment plans were blindly selected for evaluation. Those plans included four different treatment sites: prostate, brain, lung and body. The plans were evaluated against dose-volume histogram (DVH) parameters and 2D and 3D gamma analysis. The dose calculated with Eclipse treatment planning system (TPS) was compared to Compass calculated dose (CCD) and Compass reconstructed dose (CRD).
RESULTS: The maximum differences in mean dose of planning target volume (PTV) were 2.7 ± 1.0% between AAA and Acuros XB calculation algorithm TPS dose, -7.6 ± 3.5% between Eclipse TPS dose and CCD dose and -5.9 ± 3.7% between Eclipse TPS dose and CRD dose for both Eclipse calculation algorithms, respectively. 2D gamma analysis was not able to identify all the cases that 3D gamma analysis specified for further verification.
CONCLUSIONS: Compass is suitable for QA of SBRT and SRT treatment plans. However, the QA process should include wide set of DVH-based dose parameters and 3D gamma analysis should be the preferred method when performing clinical patient QA. The results suggest that the Compass should not be used for smaller field sizes than 3 × 3 cm2 or the beam model should be adjusted separately for both small (FS ≤ 3 cm) and large (FS > 3 cm) field sizes.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Compass system; Model-based QA; Stereotactic treatment; VMAT

Mesh:

Year:  2017        PMID: 29254590     DOI: 10.1016/j.ejmp.2017.11.009

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Validation of new transmission detector transmission factors for online dosimetry: an experimental study.

Authors:  So-Yeon Park; Jong Min Park; Jung-In Kim; Sungyoung Lee; Chang Heon Choi
Journal:  Radiat Oncol       Date:  2018-08-24       Impact factor: 3.481

2.  Implementing and evaluating a high-resolution diode array for patient-specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy.

Authors:  Qianyi Xu; Kiet Huynh; Wei Nie; Mark S Rose; Ashish K Chawla; Kevin S Choe; Samir Kanani; Gregory J Kubicek; Jiajin Fan
Journal:  J Appl Clin Med Phys       Date:  2022-03-12       Impact factor: 2.243

3.  Dosimetric Impact of Acuros XB Dose-to-Water and Dose-to-Medium Reporting Modes on Lung Stereotactic Body Radiation Therapy and Its Dependency on Structure Composition.

Authors:  Carles Muñoz-Montplet; Rafael Fuentes-Raspall; Diego Jurado-Bruggeman; Sebastià Agramunt-Chaler; Albert Onsès-Segarra; Maria Buxó
Journal:  Adv Radiat Oncol       Date:  2021-05-19
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.