Method comparison and validation of a prototype device for measurement of ionized calcium concentrations cow-side against a point-of-care instrument and a benchtop blood-gas analyzer reference method.

R C Neves1, T Stokol1, K D Bach1, J A A McArt2.   

Abstract

The objective of this study was to assess an optimized ion-selective electrode Ca-module prototype as a potential cow-side device for ionized Ca (iCa) measurements in bovine blood. A linearity experiment showed no deviation from linearity over a range of iCa concentrations compared with a commercial point-of-care (POC) device commonly used in the field (POCVS; VetScan i-STAT, Abaxis North America, Union City, CA) and a laboratory gold standard benchtop blood-gas analyzer [reference analyzer (RA); ABL-800 FLEX, Radiometer Medical, Copenhagen, Denmark]. Coefficient of variation on 3 samples with high, within-range, and low iCa concentrations ranged from 1.0 to 3.9% for the prototype. A follow-up validation experiment was performed, in which our objectives were to (1) assess the performance of the prototype cow-side against the POCVS (farm gold-standard) using fresh non-anticoagulated whole-blood samples; (2) assess the performance of the prototype and the POCVS against the RA in a diagnostic laboratory using blood collected in a heparin-balanced syringe; and (3) assess the agreement of the prototype and POCVS on-farm (fresh non-anticoagulated whole blood) against the RA on heparin-balanced blood. Finally, sensitivity and specificity of the results obtained by the prototype and the POCVS cow-side compared with the results obtained by the laboratory RA using 3 different iCa cut points for classification of subclinical hypocalcemia were calculated. A total of 101 periparturient Holstein cows from 3 dairy farms in New York State were used for the second experiment. Ionized Ca results from the prototype cow-side were, on average, 0.06 mmol/L higher than the POCVS. With heparin-balanced samples under laboratory conditions, the prototype and POCVS measured an average 0.04 mmol/L higher and lower, respectively, compared with the RA. Results from the prototype and POCVS cow-side were 0.01 mmol/L higher and 0.05 mmol/L lower, respectively, compared with results from the laboratory RA on heparinized blood. Sensitivity and specificity for the prototype and the POCVS under farm conditions at 3 potential subclinical hypocalcemia cut points were 100 and ≥93.5%, respectively. This novel ion-selective electrode Ca-module could become a rapid low-cost tool for assessing iCa cow-side, while qualitatively allowing classification of subclinical hypocalcemia on-farm. The Authors. Published by the Federation of Animal Science Societies and Elsevier Inc. on behalf of the American Dairy Science Association®. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/3.0/).

Entities:  

Keywords:  cow-side; ionized calcium; point-of-care; subclinical hypocalcemia

Mesh:

Substances:

Year:  2017        PMID: 29248221     DOI: 10.3168/jds.2017-13779

Source DB:  PubMed          Journal:  J Dairy Sci        ISSN: 0022-0302            Impact factor:   4.034


  6 in total

1.  Effect of induced hypocalcemia in nonlactating, nonpregnant Holstein cows fed negative DCAD with low, medium, or high concentrations of calcium.

Authors:  Laura A Amundson; Angela D Rowson; Peter M Crump; Austin P Prichard; Adrienne A Cheng; Collin E Wimmler; Marisa Klister; Samantha R Weaver; Scott S Bascom; Dennis E Nuzback; Ken P Zanzalari; Laura L Hernandez
Journal:  J Anim Sci       Date:  2018-12-03       Impact factor: 3.159

2.  Systematic comparisons of various markers for mast cell activation in RBL-2H3 cells.

Authors:  Chuda Rujitharanawong; Sunisa Yoodee; Kanyarat Sueksakit; Paleerath Peerapen; Papapit Tuchinda; Kanokvalai Kulthanan; Visith Thongboonkerd
Journal:  Cell Tissue Res       Date:  2022-09-20       Impact factor: 4.051

3.  Validation of the bovine blood calcium checker as a rapid and simple measuring tool for the ionized calcium concentration in cattle.

Authors:  Kazuyuki Suzuki; Nao Kondo; Kaede Takagi; Akitoyo Nishikawa; Yoshiki Murakami; Marina Otsuka; Kenji Tsukano; Keiko Ikeda; Hisashi Funakura; Ichiro Yasutomi; Satoshi Kawamoto
Journal:  J Vet Med Sci       Date:  2021-03-26       Impact factor: 1.267

4.  In vitro characterization of PrismaLung+: a novel ECCO2R device.

Authors:  Ingeborg Hospach; Jacques Goldstein; Kai Harenski; John G Laffey; Dominique Pouchoulin; Manuela Raible; Stefanie Votteler; Markus Storr
Journal:  Intensive Care Med Exp       Date:  2020-05-13

5.  Comparison of the bovine blood gas parameters produced with three types of portable blood gas analyzers.

Authors:  Younghye Ro; Woojae Choi; Leegon Hong; Eunkyung Kim; Eunhui Choe; Danil Kim
Journal:  J Vet Sci       Date:  2022-07       Impact factor: 1.603

6.  Evaluation of the accuracy of an electronic point-of-care analyzer to quantify blood creatinine concentration in goats.

Authors:  Melanie J Boileau; Leslie Wagner; Jared D Taylor
Journal:  J Vet Intern Med       Date:  2021-02-24       Impact factor: 3.333

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.