Porous architected biomaterial for a tibial-knee implant with minimum bone resorption and bone-implant interface micromotion.

Amirmohammad Rahimizadeh1, Zahra Nourmohammadi1, Sajad Arabnejad2, Michael Tanzer3, Damiano Pasini4.   

Abstract

This investigation presents the numerical development of a fully porous tibial knee implant that is suggested to alleviate the clinical problems associated with current prostheses that are fully solid. A scheme combining multiscale mechanics and topology optimization is proposed to handle the homogenized analysis and property tailoring of the porous architecture with the aim of reducing the stiffness mismatch between the implant and surrounding bone. The outcome of applying this scheme is a graded lattice microarchitecture that can potentially offer the implant an improved degree of load bearing capacity while reducing concurrently bone resorption and interface micromotion. Asymptotic Homogenization theory is used to characterize the mechanics of its building block, a tetrahedron based unit cell, and the Soderberg fatigue criterion to represent the implant fatigue resistance under multiaxial physiological loadings. The numerical results suggest that the overall amount of bone resorption around the graded porous tibial stem is 26% lower than that around a conventional, commercially available, fully dense titanium implant of identical shape and size. In addition, an improved interface micromotion is observed along the tibial stem, with values at the tip of the stem as low as 17µm during gait cycle and 22µm for deep bend compared to a fully dense implant. This decrease in micromotion compared to that of an identical solid implant made of titanium can reasonably be expected to alleviate post-operative end of stem pain suffered by some patients undergoing surgery at the present time.
Copyright © 2017 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Bone resorption; End-of-stem pain; Multiscale mechanics; Porous biomaterial; Tibial knee implant; Topology optimization

Mesh:

Substances:

Year:  2017        PMID: 29247979     DOI: 10.1016/j.jmbbm.2017.11.041

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  2 in total

1.  Novel Design of the Compound Sleeve and Stem Prosthesis for Treatment of Proximal Femur Bone Defects Based on Topology Optimization.

Authors:  Haowen Xue; Haotian Bai; Rongqi Zhou; Jincheng Wang; Bin Zhou; Xiaonan Wang; Wenbin Luo; Xin Zhao
Journal:  Front Bioeng Biotechnol       Date:  2022-06-24

2.  Optimal design and biomechanical analysis of sandwich composite metal locking screws for far cortical locking constructs.

Authors:  Yuping Deng; Dongliang Zhao; Yang Yang; Hanbin Ouyang; Chujiang Xu; Liang Xiong; Yanbin Li; Wenchang Tan; Gang Huang; Wenhua Huang
Journal:  Front Bioeng Biotechnol       Date:  2022-09-27
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.