The Clinical Utility of Noninvasive Pulse Co-oximetry Hemoglobin Measurements in Dark-Skinned Critically Ill Patients.

Susan M Murphy1, Shahed Omar.   

Abstract

BACKGROUND: The primary objective of this study was to assess the clinical usefulness of a point-of-care device which measures hemoglobin noninvasively (SpHb) in a group of critically ill participants with dark skin pigmentation.
METHODS: One hundred forty-six adult and pediatric participants from a multidisciplinary intensive care unit had intermittent readings of noninvasive hemoglobin measurements performed at a minimum of 4 hourly intervals. A total of 371 readings were analyzed. Concurrent blood samples were taken to assess hemoglobin levels using point-of-care blood gas analyzer, as well as sent to a central laboratory where hemoglobin was measured using the sodium lauryl sulfate method. Bland-Altman plots were constructed to assess the agreement between results from the 2 point-of-care devices with the reference standard (laboratory hemoglobin).
RESULTS: SpHb exhibited significant bias when compared to laboratory hemoglobin, while blood gas hemoglobin did not. Mean bias for SpHb was +1.64 with limits of agreement of -1.03 to 4.31 compared to blood gas hemoglobin which showed a bias of 0.26 and limits of agreement of -0.84 to 1.37. The magnitude of the bias for SpHb increased with increasing mean hemoglobin levels. Of all the additional study variables assessed for effect on the bias, only Acute Physiology and Chronic Health Evaluation II score in adult patients (P < .0001) and mean arterial blood pressure (P = .001) had an effect. Skin pigmentation did not have any effect on the magnitude of bias.
CONCLUSIONS: Noninvasive Hemoglobin measurement is a promising tool in dark-skinned critically ill patients with low hemoglobin levels, but requires further refinements for it to have clinical usefulness.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29239951     DOI: 10.1213/ANE.0000000000002721

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  2 in total

1.  Accuracy of Continuous and Noninvasive Hemoglobin Monitoring in the Presence of CO2 Insufflation: An Observational Pilot Study.

Authors:  Hye Sun Lee; Ji Won Yoo; Ha Yeon Kim; Na Young Kim; Ji Eun Kim
Journal:  Med Sci Monit       Date:  2021-10-27

2.  Impact of Increased End-Tidal Carbon Dioxide on Continuous Noninvasive Hemoglobin Monitoring during Laparoscopic Gastrectomy: A Randomized Controlled Study.

Authors:  Ha-Yeon Kim; Jong-Bum Choi; Sang-Uk Han; Hye-Sun Lee; Kyuhyeok Lee; Ji-Eun Kim
Journal:  J Pers Med       Date:  2022-01-26
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.