A double-reprocessing high-level disinfection protocol does not eliminate positive cultures from the elevators of duodenoscopes.

Douglas K Rex1, Marnie Sieber2, Glen A Lehman1, Douglas Webb2, Bryan Schmitt3, Amy Beth Kressel2, Ji Young Bang1, Jeffery Easler1, Lee McHenry1, Ihab El-Hajj1, Evan Fogel1, James Watkins1, Stuart Sherman1.   

Abstract

BACKGROUND AND STUDY AIM: Duodenoscopes have been the source of serious infection, despite correct performance of high-level disinfection (HLD). This study aimed to observe the impact of performing HLD twice on the rate of positive cultures from duodenoscope elevators.
METHODS: We performed double HLD (DHLD; i. e. complete manual cleaning followed by automated reprocessing, with the entire process repeated) and then randomly cultured the elevators of our duodenoscopes on about 30 % of occasions.
RESULTS: DHLD was associated with positive elevator cultures for any microorganism in 9.4 % of cases, with a 0.8 % rate of known pathogens (627 cultures) between May 2015 and February 2016. After February 2016, and in association with changing the precleaning fluid, as well as use of a new FDA-recommended cleaning brush, the rate of positive cultures for any microorganism after DHLD was 4.8 % and 0.2 % for known pathogens (420 cultures). In a third phase, characterized by a change in personnel performing DHLD and retirement of a duodenoscope with a high rate of positive cultures, the rate of positive cultures for any microorganism was 4.9 % (783 cultures) and the rate of positive culture for known pathogens was 0.3 %. To our knowledge, no duodenoscope transmission of infection occurred during the study interval.
CONCLUSIONS: DHLD resulted in a low rate of positive cultures for known pathogens and for organisms of low pathogenic potential, but did not eliminate these, from duodenoscope elevators. Additional improvements in HLD protocols and/or duodenoscope design are needed. © Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 29237202     DOI: 10.1055/s-0043-122378

Source DB:  PubMed          Journal:  Endoscopy        ISSN: 0013-726X            Impact factor:   10.093


  8 in total

Review 1.  Duodenoscope-Associated Infections: Update on an Emerging Problem.

Authors:  M Rubayat Rahman; Abhilash Perisetti; Roxana Coman; Pardeep Bansal; Rajiv Chhabra; Hemant Goyal
Journal:  Dig Dis Sci       Date:  2018-12-19       Impact factor: 3.199

2.  Performance and applicability of a first generation single-use duodenoscope: a single-center cohort study.

Authors:  Diederik Persyn; Hannah Van Malenstein; Emma Vanderschueren; Annette Schuermans; Mira Dreesen; Wouter Meert; Kristof Buysschaert; Schalk Van Der Merwe; Wim Laleman
Journal:  Ann Gastroenterol       Date:  2022-06-02

3.  Rate and impact of duodenoscope contamination: A systematic review and meta-analysis.

Authors:  Sara Larsen; Rasmus Vinther Russell; Lotte Klinten Ockert; Stephen Spanos; Helena Strømstad Travis; Lars Holger Ehlers; Anders Mærkedahl
Journal:  EClinicalMedicine       Date:  2020-07-15

4.  Double high-level disinfection versus liquid chemical sterilization for reprocessing of duodenoscopes used for ERCP: a prospective randomized study.

Authors:  Mark A Gromski; Marnie S Sieber; Stuart Sherman; Douglas K Rex
Journal:  Gastrointest Endosc       Date:  2020-07-31       Impact factor: 9.427

Review 5.  Duodenoscope-associated infection prevention: A call for evidence-based decision making.

Authors:  Cori L Ofstead; Brandy L Buro; Krystina M Hopkins; John E Eiland; Harry P Wetzler; David R Lichtenstein
Journal:  Endosc Int Open       Date:  2020-11-17

6.  Contamination Rates in Duodenoscopes Reprocessed Using Enhanced Surveillance and Reprocessing Techniques: A Systematic Review and Meta-Analysis.

Authors:  Shivanand Bomman; Munish Ashat; Navroop Nagra; Mahendran Jayaraj; Shruti Chandra; Richard A Kozarek; Andrew Ross; Rajesh Krishnamoorthi
Journal:  Clin Endosc       Date:  2022-01-03

Review 7.  A narrative review on current duodenoscope reprocessing techniques and novel developments.

Authors:  Maarten Heuvelmans; Herman F Wunderink; Henny C van der Mei; Jan F Monkelbaan
Journal:  Antimicrob Resist Infect Control       Date:  2021-12-23       Impact factor: 4.887

Review 8.  Single Use (Disposable) Duodenoscope: Recent Development and Future.

Authors:  Kihyun Ryu; Sunguk Jang
Journal:  Clin Endosc       Date:  2021-06-22
  8 in total