| Literature DB >> 29235991 |
Mitsutaka Yamaguchi1, Yuto Nagao, Koki Ando, Seiichi Yamamoto, Makoto Sakai, Raj Kumar Parajuli, Kazuo Arakawa, Naoki Kawachi.
Abstract
A feasibility study on the imaging of monochromatic carbon-ion beams for carbon-ion therapy was performed. The evaluation was based on Monte Carlo simulations and beam-irradiation experiments, using a pinhole x-ray camera, which measured secondary electron bremsstrahlung (SEB). The simulation results indicated that the trajectories of the carbon-ion beams with injection energies of 278, 249 and 218 MeV/u in a water phantom, were clearly imaged by measuring the SEB with energies from 30 to 60 keV, using a pinhole camera. The Bragg-peak positions for these three injection energies were located at the positions where the ratios of the counts of SEB acquisitions to the maximum counts were approximately 0.23, 0.26 and 0.29, respectively. Moreover, we experimentally demonstrated that it was possible to identify the Bragg-peak positons, at the positions where the ratios coincided with the simulation results. However, the estimated Bragg-peak positions for the injection energies of 278 and 249 MeV/u were slightly deeper than the expected positions. In conclusion, for both the simulations and experiments, we found that the 25 mm shifts in the Bragg-peak positions can be observed by this method.Entities:
Mesh:
Year: 2018 PMID: 29235991 DOI: 10.1088/1361-6560/aaa17c
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609