Literature DB >> 29232764

Brain biopsy performed with the RONNA G3 system: a case study on using a novel robotic navigation device for stereotactic neurosurgery.

Domagoj Dlaka1, Marko Švaco1, Darko Chudy1, Bojan Jerbić1, Bojan Šekoranja1, Filip Šuligoj1, Josip Vidaković1, Fadi Almahariq1, Dominik Romić1.   

Abstract

BACKGROUND: Robotic neuronavigation is becoming an important tool for neurosurgeons. We present a case study of a frameless stereotactic biopsy guided by the RONNA G3 robotic neuronavigation system.
METHODS: A 45 year-old patient with a history of vertigo, nausea and vomiting was diagnosed with multiple periventricular lesions. Neurological status was unremarkable. A frameless robotic biopsy of a brain lesion was performed.
RESULTS: Three tissue samples were obtained. There were no intraoperative or postoperative complications. Histological analysis showed a B-cell lymphoma. After merging the preoperative CT scan with the postoperative MRI and CT scans, the measured error between the planned and the postoperatively measured entry point was 2.24 mm and the measured error between the planned and postoperatively measured target point was 2.33 mm.
CONCLUSIONS: The RONNA G3 robotic system was used to navigate a Sedan brain biopsy needle to take tissue samples and could be a safe and precise tool for brain biopsy.
Copyright © 2017 John Wiley & Sons, Ltd.

Entities:  

Keywords:  RONNA G3; brain biopsy; computer-assisted surgery; frameless stereotaxy; robotic neurosurgery

Mesh:

Year:  2017        PMID: 29232764     DOI: 10.1002/rcs.1884

Source DB:  PubMed          Journal:  Int J Med Robot        ISSN: 1478-5951            Impact factor:   2.547


  4 in total

1.  Frameless stereotactic brain biopsy and external ventricular drainage placement using the RONNA G4 system.

Authors:  Marina Raguž; Domagoj Dlaka; Darko Orešković; Anđelo Kaštelančić; Darko Chudy; Bojan Jerbić; Bojan Šekoranja; Filip Šuligoj; Marko Švaco
Journal:  J Surg Case Rep       Date:  2022-05-31

2.  Robot-assisted stereotactic brainstem biopsy in children: prospective cohort study.

Authors:  William Dawes; Hani J Marcus; Martin Tisdall; Kristian Aquilina
Journal:  J Robot Surg       Date:  2018-12-06

3.  Robot-assisted stereotactic brain biopsy: systematic review and bibliometric analysis.

Authors:  Hani J Marcus; Vejay N Vakharia; Sebastien Ourselin; John Duncan; Martin Tisdall; Kristian Aquilina
Journal:  Childs Nerv Syst       Date:  2018-05-10       Impact factor: 1.475

4.  Stereotactic Neuro-Navigation Phantom Designs: A Systematic Review.

Authors:  Marko Švaco; Ivan Stiperski; Domagoj Dlaka; Filip Šuligoj; Bojan Jerbić; Darko Chudy; Marina Raguž
Journal:  Front Neurorobot       Date:  2020-10-23       Impact factor: 2.650