Wireless Neuromodulation for Chronic Back Pain: Delivery of High-Frequency Dorsal Root Ganglion Stimulation by a Minimally Invasive Technique.

Bart Billet1, Roel Wynendaele1, Niek E Vanquathem2.   

Abstract

OBJECTIVE: To evaluate the analgesic effect of a dorsal root ganglion (DRG) stimulation technology utilizing high-frequency pulse rates to treat intractable chronic back and leg pain.
METHODS: This case study presents the outcomes, with a novel, wireless, minimally invasive miniature neurostimulator system in a case of chronic back pain. The subject was implanted bilaterally with a Freedom 4A quadripolar electrode array at the L2 dorsal root ganglion. Stimulation was applied using 10 kHz pulse rate and 30 μs pulse width. A VAS pain-rating scale, Oswestry Disability Index (ODI), EQ-5D-5L Quality of Life Questionnaire 5 dimensions, and Patients' Global Impression of Change (PGIC) scale were evaluated at 12 weeks and 6 months post implantation.
RESULTS: VAS pain scores for back pain reduced from 91 to 31 mms and 80 to 35 mms for leg pain. Additionally, while stimulation remained paresthesia-free, there were a marked decrease in pain medications and an increase in quality of life. Also, an increase in functionality from crippled to moderate was reported. There were no adverse reactions related to the procedure or device.
CONCLUSION: The minimally invasive, wireless approach to deliver high-frequency, paresthesia-free DRG stimulation for treatment of chronic back and leg pain associated with FBSS was effective and encouraging.

Entities:  

Year:  2017        PMID: 29230247      PMCID: PMC5688258          DOI: 10.1155/2017/4203271

Source DB:  PubMed          Journal:  Case Rep Med


1. Introduction

Spinal cord stimulation (SCS) is a well-established therapy for FBSS and is widely accepted in terms of safety and efficacy [1-3]. Though studies have established the cost-effectiveness of SCS for FBSS, SCS has certain limitations, and only 50% of patients are able to report long-term success [4-6]. Reasons for inferior outcomes include device migration, ineffective stimulation parameters, device failure, and positional changes of the hardware. Positional changes can result in alterations of paresthesia distribution. The relative distance between the electrodes and spinal cord tracts can be different in certain positions (seated or lying down) due to the bulk of the battery or extensive wiring involved in an SCS operating system [2, 7]. These side effects may not be easily seen in the placement and/or displacement of the battery or electrodes. Alternative techniques of stimulation and their targets are recommended to provide a better and more stable relief to intractable pain syndromes. The DRG is a cluster of cells inside the dura that transmits sensory information, including nociceptive signals, to the dorsal columns of the spinal cord. Studies have implicated the role of the DRG in the development and maintenance of chronic pain perception with demonstration of changes in cell membrane and gene expression [8, 9]. In cases not suitable for SCS, stimulation of the DRG may prove to be an important alternative. The bony encasement of the target and its stable anatomical position can mitigate issues, such as lead migration. The anatomical location of the DRG offers a closer proximity to the electrodes compared to the spinal cord and its dorsal columns. Thus, the stimulation parameters are expected to be more tolerable and lower in power consumption [10]. The wireless neurostimulation system presented in this report is aptly suited for overcoming the difficulties of the SCS while profiting from the benefits of DRG stimulation.

2. Materials and Methods

This is a case illustration of a patient with intractable chronic back pain following spine surgery.

2.1. Case Report

This subject had a traffic accident resulting in traumatic disc herniation at L5 and S1. A microdiscectomy was performed at L5 and S1. However, postoperative residual back pain remained, along with neuropathic pain on the right side along L5 dermatome distribution. As a result, an anterior lumbar interbody fusion was performed at L5 and S1, 6 months later. Nevertheless, the subject continued to have invalidating lumbar pain and persisting neuropathic pain along L5 dermatome on the right side. Interventional pain management was offered, as medical management (with tramadol HCl and paracetamol) failed to provide significant relief. Pulsed radiofrequency treatment followed by ablation of the facet joint did not provide relief. At this point, bilateral placement of DRG electrodes at L2 was offered.

2.2. Device Description

Refer to our earlier publication regarding the device [11]. The patient was implanted with two Freedom 4A electrode arrays, each array containing four contacts, at the L2 dorsal root ganglion under fluoroscopic guidance. The stimulator system utilizes an implantable passive electrode contact array, microprocessor receiver, and antenna embedded within the electrode wire that couples to an external transmitting antenna and external pulse generator (EPG) (Figures 1 and 2). The external transmitters (Figure 3) are worn by the patient over a single layer of clothing and are used to transmit power to the stimulator. The EPG is programmed by the clinician for the required stimulation parameters. The system uses 915 MHz radiofrequency energy for the transmission, and the distance between the implant and antenna remains short and the energy was relatively low. Wavelengths and product specifications have been designed to decrease risk related to the wireless transmission of energy [12] and reliably transfer the clinician's desired stimulation parameters with a wide available spectrum (amplitude: 1–24 mA; pulse width: 1–1000 microsec; frequency: 1–10,000 Hz).
Figure 1

Neurostimulator electrode, MRI compatible, for both 1.5 and 3 Tesla (copyright StimRelieve LLC).

Figure 2

Neurostimulator receiver (copyright StimRelieve LLC).

Figure 3

Freedom SCS external device (copyright StimRelieve LLC).

2.3. Surgical Procedure

Under strict aseptic precautions, the skin and subcutaneous tissues were infiltrated with local 1% lidocaine. A small skin incision was made for needle insertion, which was shaped by hand to match the contour to achieve appropriate electrode placement. Insertion of the lead was performed transverse and translaminar through 14 G Tuohy needle at L2 exiting nerve roots, coming from the cephalad end. Biplanar fluoroscopic images were used to monitor electrode positioning (Figures 4 and 5). The device was anchored via a subdermal suture located at the skin entry point. Distal tubing cut at the insertion point was buried subcutaneously, and the skin incision was closed.
Figure 4

AP view of lumbar spine X-ray showing the electrode placement.

Figure 5

X-ray of lumbar spine, lateral view confirming the placement of electrodes.

2.4. Stimulation Protocol

Stimulation parameters were set at pulse widths of 30 microseconds and frequency of 10 kHz with intensity set between 1.5 and 2.5 mA for bilateral stimulation (both devices activated at the same time) with bipolar electrode selection, closest to the dorsal root ganglion. (This is not to be confused with the device communication frequency between the external generator and electrode microprocessor of 915 MHz.) During stimulation sessions, when therapy was needed to alleviate pain, patients wore an external transmitter over a single layer of clothing, positioned over the implant location of the electrode array (Figure 3).

3. Results

A VAS pain-rating scale, Oswestry Disability Index (ODI), EQ-5D-5L Quality of Life Questionnaire, and Patients' Global Impression scale (PGIC) were administered at 3, 5, 8, 12 weeks, and 6 months post implantation. After the procedure, clinical response ensued in the form of 66% improvement for back pain and 56% for leg pain at 6-month follow-up as measured with VAS. There was a steady and progressive improvement of gait and posture. Functionality (ODI) improved with 61% from crippled to moderate disability. The patient reported a definite improvement on the PGIC (6 out 7) and an important increase in quality of life with the EQ-5D-5L (Table 1). Pain medication was reduced to tramadol HCl 37.5 mg twice daily.
Table 1

EQ-5D-5L.

VariableMobilitySelf-careActivityPainAnxietyStateEQ-VAS
Description1 = no problems1 = no problems1 = no problems1 = no pain1 = not anxious5-digit code for EQ-5D-5L999 = missing value
2 = slight problems2 = slight problems2 = slight problems2 = slight pain2 = slightly anxious
3 = moderate problems3 = moderate problems3 = moderate problems3 = moderate pain3 = moderately anxious
4 = severe problems4 = severe problems4 = severe problems4 = severe pain4 = severely anxious
5 = unable to5 = unable to5 = unable to5 = extreme pain5 = extremely anxious
Baseline4444344,44320
6 months2223222,23273

4. Discussion

High-frequency stimulation at the DRG was effective in the treatment of pain associated with FBSS and for pain localized to the back and lower extremities. PROCESS reported that 50–60% of FBSS subjects achieved 50% or better leg pain relief through six months of SCS therapy [4, 13]. DRG stimulation seems to be better in terms of anatomical stability with lead migration of 3% (2 leads out of 67) [12], well below the 13.2% rate of migration for SCS with percutaneous leads placed over the dorsal columns, reported in a literature review of 51 studies [2], and 23% in a prospective study [14]. Strong supporters for DRG stimulation include its anatomical location that mitigates positional or postural influences on the device as well as the patient. There is also positive ability in the target region to induce stimulation coverage for those areas, which SCS would not be able to reach, for example, toes and foot [15, 16]. Additionally, the stimulation parameters require a lower demand in energy consumption. In the present case illustration, we demonstrate the analgesic effect of high-frequency DRG stimulation, applying our minimally invasive placement of stimulation electrodes and wireless neuromodulation. This also supports and expands the number of indications for the treatment of other chronic pain conditions. There is no hardware and thus no complications related to its components. Percutaneous electrode placement devoid of any implanted pulse generator or the long connective wires can be advantageous to both patients and surgeons. They do not only add to comfort and cosmetics but also reduce costs, operating time, and postoperative pain, thus minimizing adverse events while achieving the desired pain control [17]. Larger prospective studies are required to further our knowledge about the wireless neuromodulation technology that has yielded promising results, thus far. We are in the process of evaluating the technology in other related areas of pain management also.
  15 in total

1.  Neurophysiological assessment of spinal cord stimulation in failed back surgery syndrome.

Authors:  Daniel Ciampi de Andrade; Belgacem Bendib; Mohammed Hattou; Yves Keravel; Jean-Paul Nguyen; Jean-Pascal Lefaucheur
Journal:  Pain       Date:  2010-06-29       Impact factor: 6.961

2.  Epidural spinal cord stimulation: calculation of field potentials with special reference to dorsal column nerve fibers.

Authors:  J J Struijk; J Holsheimer; B K van Veen; H B Boom
Journal:  IEEE Trans Biomed Eng       Date:  1991-01       Impact factor: 4.538

Review 3.  Peripheral Neurostimulation with a Microsize Wireless Stimulator.

Authors:  Thomas L Yearwood; Laura Tyler Perryman
Journal:  Prog Neurol Surg       Date:  2015-09-04

4.  Pulsed radiofrequency adjacent to the cervical dorsal root ganglion in chronic cervical radicular pain: a double blind sham controlled randomized clinical trial.

Authors:  Jan Van Zundert; Jacob Patijn; Alfons Kessels; Inge Lamé; Hans van Suijlekom; Maarten van Kleef
Journal:  Pain       Date:  2006-10-18       Impact factor: 6.961

5.  Spinal cord stimulation versus repeated lumbosacral spine surgery for chronic pain: a randomized, controlled trial.

Authors:  Richard B North; David H Kidd; Farrokh Farrokhi; Steven A Piantadosi
Journal:  Neurosurgery       Date:  2005       Impact factor: 4.654

6.  Rates of lead migration and stimulation loss in spinal cord stimulation: a retrospective comparison of laminotomy versus percutaneous implantation.

Authors:  David D Kim; Rakesh Vakharyia; Henry R Kroll; Adam Shuster
Journal:  Pain Physician       Date:  2011 Nov-Dec       Impact factor: 4.965

7.  The effects of spinal cord stimulation in neuropathic pain are sustained: a 24-month follow-up of the prospective randomized controlled multicenter trial of the effectiveness of spinal cord stimulation.

Authors:  Krishna Kumar; Rod S Taylor; Line Jacques; Sam Eldabe; Mario Meglio; Joan Molet; Simon Thomson; Jim O'Callaghan; Elon Eisenberg; Germain Milbouw; Eric Buchser; Gianpaolo Fortini; Jonathan Richardson; Richard B North
Journal:  Neurosurgery       Date:  2008-10       Impact factor: 4.654

8.  Spinal cord stimulation in treatment of chronic benign pain: challenges in treatment planning and present status, a 22-year experience.

Authors:  Krishna Kumar; Gary Hunter; Denny Demeria
Journal:  Neurosurgery       Date:  2006-03       Impact factor: 4.654

Review 9.  Safety and efficacy of spinal cord stimulation for the treatment of chronic pain: a 20-year literature review.

Authors:  Tracy Cameron
Journal:  J Neurosurg       Date:  2004-03       Impact factor: 5.115

10.  Spinal cord stimulation of the dorsal root ganglion for groin pain-a retrospective review.

Authors:  Stefan Schu; Ashish Gulve; Sam ElDabe; Ganesan Baranidharan; Katharina Wolf; Walter Demmel; Dirk Rasche; Manohar Sharma; Daniel Klase; Gunnar Jahnichen; Anders Wahlstedt; Harold Nijhuis; Liong Liem
Journal:  Pain Pract       Date:  2014-04-01       Impact factor: 3.183

View more
  2 in total

Review 1.  Neuromodulation with electrical field stimulation of dorsal root ganglion in various pain syndromes: a systematic review with focus on participant selection.

Authors:  Ivana Vuka; Tihana Marciuš; Svjetlana Došenović; Lejla Ferhatović Hamzić; Katarina Vučić; Damir Sapunar; Livia Puljak
Journal:  J Pain Res       Date:  2019-02-27       Impact factor: 3.133

2.  Somatic inhibition by microscopic magnetic stimulation.

Authors:  Hui Ye; Lauryn Barrett
Journal:  Sci Rep       Date:  2021-06-30       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.