Laboratory Evaluation of the Influence of Decentration and Pupil Size on the Optical Performance of a Monofocal, Bifocal, and Trifocal Intraocular Lens.

Tamer Tandogan, Hyeck Soo Son, Chul Young Choi, Michael C Knorz, Gerd U Auffarth, Ramin Khoramnia.   

Abstract

PURPOSE: To assess the influence of decentration and aperture size on the optical quality of different intraocular lenses (IOLs) of the same material, body design, and refractive power using standardized optical bench testing.
METHODS: Using an optical bench set-up, an aspheric monofocal (CT ASPHINA 409M; Carl Zeiss Meditec, Jena, Germany), an aspheric diffractive bifocal (AT LISA 809M; Carl Zeiss Meditec), and an aspheric diffractive trifocal (AT LISA 839M; Carl Zeiss Meditec) intraocular lens (IOL) were evaluated, each with the same distance power, body design, and material. Modulation transfer function (MTF) values were measured at spatial frequencies of 50 lp/mm and aperture sizes of 3 and 4.5 mm. Each IOL was measured while centered, then decentered by 0.25, 0.5, 0.75, and 1 mm.
RESULTS: MTF values for the monofocal IOL at far focus with 3- and 4.5-mm aperture size were 0.80/0.80 with maximum reduction to 0.77/0.73 for 1-mm decentration, respectively. Centered IOL MTFs of the bifocal and trifocal IOLs were lower for the far focus at 0.46/0.41 and 0.39/0.26, with reduction at 1-mm decentration to 0.35/0.25 and 0.25/0.18, respectively. Values for near focus of the bifocal and trifocal IOLs reduced from 0.27/0.31 and 0.19/0.18 to 0.2/0.21 and 0.12/0.13, respectively. The trifocal intermediate focus MTF reduced from 0.15/0.10 to 0.12/0.08. MTF values of all three lenses decreased significantly under all conditions with decentration of 0.5 to 0.75 mm.
CONCLUSIONS: Monofocal lenses were least negatively affected by decentration, with mean optical quality reduction of less than 10% for 1-mm decentration at physiological pupil sizes. For diffractive bifocal and trifocal lenses, optical quality at all distances was significantly reduced if decentration exceeded 0.75 mm, with intermediate focus showing the least reduction. [J Refract Surg. 2017;33(12):808-812.]. Copyright 2017, SLACK Incorporated.

Mesh:

Year:  2017        PMID: 29227508     DOI: 10.3928/1081597X-20171004-02

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  10 in total

1.  Reversibility of the duet procedure: Bilateral exchange of a supplementary trifocal sulcus-fixated intraocular lens for correction of a postoperative refractive error.

Authors:  Isabella Diana Baur; Gerd Uwe Auffarth; Timur Mert Yildirim; Christian Steffen Mayer; Ramin Khoramnia
Journal:  Am J Ophthalmol Case Rep       Date:  2020-10-10

Review 2.  Tilt and decentration with various intraocular lenses: A narrative review.

Authors:  Xiao-Yong Chen; Yu-Chen Wang; Tian-Yao Zhao; Zi-Zhen Wang; Wei Wang
Journal:  World J Clin Cases       Date:  2022-04-26       Impact factor: 1.534

3.  Bilateral trifocal IOL implantation in a pediatric case of cataract following steroid-therapy for acute lymphoblastic leukemia.

Authors:  Timur M Yildirim; Gerd U Auffarth; Hyeck-Soo Son; Christian Huber; Flemming Beisse; Ramin Khoramnia
Journal:  Am J Ophthalmol Case Rep       Date:  2018-11-20

4.  Unilateral implantation of a new non-diffractive extended range-of-vision IOL in a young patient with Curschmann-Steinert myotonic dystrophy.

Authors:  Isabella Baur; Gerd U Auffarth; Grzegorz Łabuz; Ramin Khoramnia
Journal:  Am J Ophthalmol Case Rep       Date:  2021-04-17

5.  [Unilateral implantation of a new extended range of vision intraocular lens in a young patient].

Authors:  Isabella D Baur; Gerd U Auffarth; Grzegorz Łabuz; Christian S Mayer; Ramin Khoramnia
Journal:  Ophthalmologe       Date:  2020-12-10       Impact factor: 1.059

Review 6.  Refractive Outcomes after Cataract Surgery.

Authors:  Ramin Khoramnia; Gerd Auffarth; Grzegorz Łabuz; George Pettit; Rajaraman Suryakumar
Journal:  Diagnostics (Basel)       Date:  2022-01-19

7.  Monofocal intraocular lens with enhanced intermediate function as substitute for multifocal intraocular lens in positive dysphotopsia.

Authors:  Tadas Naujokaitis; Ling Zhao; Debora Scharf; Ramin Khoramnia; Gerd U Auffarth
Journal:  Am J Ophthalmol Case Rep       Date:  2022-04-10

8.  Decentration and tilt of plate-haptic multifocal intraocular lenses in myopic eyes.

Authors:  Jiaqi Meng; Wenwen He; Xianfang Rong; Ao Miao; Yi Lu; Xiangjia Zhu
Journal:  Eye Vis (Lond)       Date:  2020-04-07

9.  Ray propagation imaging and optical quality evaluation of different intraocular lens models.

Authors:  Hyeck Soo Son; Grzegorz Labuz; Ramin Khoramnia; Patrick Merz; Timur M Yildirim; Gerd U Auffarth
Journal:  PLoS One       Date:  2020-02-04       Impact factor: 3.240

10.  Optical Bench Analysis of 2 Depth of Focus Intraocular Lenses.

Authors:  Andreas F Borkenstein; Eva-Maria Borkenstein; Holger Luedtke; Ruediger Schmid
Journal:  Biomed Hub       Date:  2021-09-27
  10 in total