Literature DB >> 29223669

Experimental use of new absorbable tracheal stent.

Luciano F Schopf1, José Carlos Fraga2, Rodrigo Porto3, Luis A Santos4, Douglas R Marques5, Paulo R Sanchez6, Fabíola S Meyer7, Jane M Ulbrich8.   

Abstract

BACKGROUND/
PURPOSE: Silicone and metallic stents are not effective in children with tracheobronchial stenosis or tracheomalacia. Herein, we aimed to evaluate the clinical manifestations and histological reaction of rabbit trachea to the presence of a new poly(lactic-co-glycolic acid) with polyisoprene (PLGA/PI) polymer absorbable stent.
METHODS: Fourteen adult white rabbits (weight, 3.0-3.5kg) were randomly assigned to three groups: Group I (n=6): PLGA/PI spiral stent; Group II (n=6): PLGA/PI fragment; and Group III (n=2): controls. After a longitudinal incision on three cervical tracheal rings, the stents and fragments were inserted into the trachea and fixed onto the lateral wall with nonabsorbable sutures.
RESULTS: The stented group showed significantly more stridor at rest (p=0.0041), agitation (p=0.014), and use of accessory muscles (p=0.0002) and required more emergency endoscopies than the fragment group. Further, it showed significantly more remarkable histological inflammatory damage than the fragment and control groups (p=0.002).
CONCLUSIONS: The new PLGA/PI polymeric stent implanted into the trachea of rabbits caused more clinical manifestations and histologically verified inflammatory reaction than the PLGA/PI polymeric fragment. Future studies should be aimed at reducing the stent-wall thickness.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Absorbable tracheal stent; Bronchoscopy; Experimental tracheal surgery

Mesh:

Substances:

Year:  2017        PMID: 29223669     DOI: 10.1016/j.jpedsurg.2017.11.038

Source DB:  PubMed          Journal:  J Pediatr Surg        ISSN: 0022-3468            Impact factor:   2.545


  4 in total

1.  Preclinical evaluation of a pediatric airway stent for tracheobronchomalacia.

Authors:  Abhijit Mondal; Junhyoung Ha; Vickie Y Jo; Fei-Yi Wu; Aditya K Kaza; Pierre E Dupont
Journal:  J Thorac Cardiovasc Surg       Date:  2020-03-15       Impact factor: 5.209

2.  In-vivo efficacy of biodegradable ultrahigh ductility Mg-Li-Zn alloy tracheal stents for pediatric airway obstruction.

Authors:  Jingyao Wu; Leila J Mady; Abhijit Roy; Ali Mübin Aral; Boeun Lee; Feng Zheng; Toma Catalin; Youngjae Chun; William R Wagner; Ke Yang; Humberto E Trejo Bittar; David Chi; Prashant N Kumta
Journal:  Commun Biol       Date:  2020-12-18

3.  Decellularized Wharton Jelly Implants Do Not Trigger Collagen and Cartilaginous Tissue Production in Tracheal Injury in Rabbits.

Authors:  Katia Martins Foltz; Aloysio Enck Neto; Júlio César Francisco; Rossana Baggio Simeoni; Anna Flávia Ribeiro Dos Santos Miggiolaro; Thatyanne Gradowski do Nascimento; Bassam Felipe Mogharbel; Katherine Athayde Teixeira de Carvalho; José Rocha Faria-Neto; Lúcia de Noronha; Luiz César Guarita-Souza
Journal:  Life (Basel)       Date:  2022-06-23

Review 4.  Scaffold strategies combined with mesenchymal stem cells in vaginal construction: a review.

Authors:  Nicole Andréa Corbellini Henckes; Dalana Faleiro; Laura Chao Chuang; Elizabeth Obino Cirne-Lima
Journal:  Cell Regen       Date:  2021-08-02
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.