Julie Bensimon Etzol1, Marco Valente2, Sandrine Altmeyer3, Caroline Bettencourt1, Samuel Bouvet1, Guillaume Cosler2, François Desangles2, Michel Drouet2, Fabrice Entine4, Francis Hérodin2, Flora Jourquin4, Yannick Lecompte4, Patrick Martigne2, Xavier Michel4, Jérôme Pateux2, Nicolas Ugolin3, Sylvie Chevillard3.
Abstract
DosiKit is a new field-radiation biodosimetry immunoassay for rapid triage of individuals exposed to external total-body irradiation. Here, we report on the validation of this immunoassay in human blood cell extracts 0.5 h after in vitro exposure to 137Cs gamma rays, using γ-H2AX analysis. First, calibration curves were established for five donors at doses ranging from 0 to 10 Gy and dose rates ranging from ∼0.8 to ∼3 Gy/min. The calibration curves, together with a γ-H2AX peptide scale, enabled the definition of inter-experimental correction factors. Using previously calculated correction factors, blind dose estimations were performed at 0.5 h postirradiation, and DosiKit performance was compared against concomitant dicentric chromosome assay (DCA), the current gold standard for external irradiation biodosimetry. A prototype was then assembled and field tested. We show that, despite significant inter-individual variations, DosiKit can estimate total-body irradiation doses from 0.5 to 10 Gy with a strong linear dose-dependent signal and can be used to classify potentially exposed individuals into three dose ranges: below 2 Gy, between 2 and 5 Gy and above 5 Gy. The entire protocol can be performed in 45 min, from sampling to dose estimation, with a new patient triaged every 10 min. While DCA enables precise measurement of doses below 5 Gy, it is a long and difficult method. In contrast, DosiKit is a quick test that can be performed directly in the field by operational staff with minimal training, and is relevant for early field triage and identification of individuals most likely to experience acute radiation syndrome. These findings suggest that DosiKit and DCA are complementary and should be combined for triage in a mass scale event. While the proof-of-concept reported here validates the use of DosiKit at 0.5 h postirradiation, further studies are needed to calibrate and evaluate the performance of the DosiKit assay at longer times after irradiation.Entities:
Mesh:
Year: 2017 PMID: 29215325 DOI: 10.1667/RR14760.1
Source DB: PubMed Journal: Radiat Res ISSN: 0033-7587 Impact factor: 2.841